Abstract
A hyperthermia system which has been developed for clinical use is described. Different direct-contact applicators working at the ISM microwave frequency of 915 MHz are used for inducing heat in superficial tumours to a depth of about 3 cm. The system is a further development of an earlier prototype which worked at 2450 MHz. The microwave generator is controlled by a microcomputer governed by temperature measurements with thermistor probes inserted into the tissue. To minimise the influence of the interaction between the microwaves and the temperature probes, a pulsed irradiation technique is used. Up to July 1984, 80 superficial tumours had been treated with radiotherapy combined with local hyperthermia applied using this system.
Please use this url to cite or link to this publication:
author
and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Physics in Medicine and Biology
volume
30
issue
4
pages
283 - 292
publisher
IOP Publishing
external identifiers
  • pmid:4001156
  • scopus:36149038767
ISSN
1361-6560
DOI
10.1088/0031-9155/30/4/001
language
English
LU publication?
yes
id
6b290525-bbb0-418c-99e6-cd2ec589bdb3 (old id 1103386)
date added to LUP
2016-04-01 12:09:36
date last changed
2021-12-27 04:04:29
@article{6b290525-bbb0-418c-99e6-cd2ec589bdb3,
  abstract     = {{A hyperthermia system which has been developed for clinical use is described. Different direct-contact applicators working at the ISM microwave frequency of 915 MHz are used for inducing heat in superficial tumours to a depth of about 3 cm. The system is a further development of an earlier prototype which worked at 2450 MHz. The microwave generator is controlled by a microcomputer governed by temperature measurements with thermistor probes inserted into the tissue. To minimise the influence of the interaction between the microwaves and the temperature probes, a pulsed irradiation technique is used. Up to July 1984, 80 superficial tumours had been treated with radiotherapy combined with local hyperthermia applied using this system.}},
  author       = {{Nilsson, Per and Persson, B}},
  issn         = {{1361-6560}},
  language     = {{eng}},
  number       = {{4}},
  pages        = {{283--292}},
  publisher    = {{IOP Publishing}},
  series       = {{Physics in Medicine and Biology}},
  title        = {{Computer controlled microwave system for clinical hyperthermia}},
  url          = {{http://dx.doi.org/10.1088/0031-9155/30/4/001}},
  doi          = {{10.1088/0031-9155/30/4/001}},
  volume       = {{30}},
  year         = {{1985}},
}