Анализ уровней облучения взрослых пациентов при проведении наиболее распространенных рентгенографических исследований в Российской Федерации в 2009–2014 гг.

Vodovatov, Aleksandr LU ; Golikov, Vladislav ; Kalnitsky, Sergey ; Shatskiy, Iliya and Chipiga, Larisa (2017) In Radiatsionnaya Gygiena 10(3). p.66-75
Abstract
Diagnostic reference levels are the main and the most effective tools of optimization of the radiation protection of patients from medical exposure. Diagnostic reference levels should be established based on the
results of dedicated dose surveys, allowing evaluating typical patient dose distributions in a selected dose quantity for the selected X-ray examinations. The aim of the current study was to assess the distributions of typical effective doses in representative Russian regions. Materials and methods: Typical patient effective doses for the 13 most common radiographic X-ray examinations were collected in 203 X-ray rooms in 101 hospitals in six regions of Russian Federation in 2009–2014. A differentiated approach was used for... (More)
Diagnostic reference levels are the main and the most effective tools of optimization of the radiation protection of patients from medical exposure. Diagnostic reference levels should be established based on the
results of dedicated dose surveys, allowing evaluating typical patient dose distributions in a selected dose quantity for the selected X-ray examinations. The aim of the current study was to assess the distributions of typical effective doses in representative Russian regions. Materials and methods: Typical patient effective doses for the 13 most common radiographic X-ray examinations were collected in 203 X-ray rooms in 101 hospitals in six regions of Russian Federation in 2009–2014. A differentiated approach was used for the
estimation of the typical effective doses depending on the image acquisition technology. Effective doses were estimated using «EDEREX» (Russia) computational software. Results and discussion: Results of the dose
data analysis indicate the lack of significant differences between the distributions of the typical effective doses between the selected regions, allowing merging the regional samples and further evaluating the pooled (joint) sample. A significant ratio of maximum to minimum (up to two orders of magnitude) due to a presence
of X-ray units with abnormally high and low typical effective doses was observed for all 13 selected X-ray examinations. Abnormally high typical effective doses can be explained by performing the examinations using high values of tube current-time product (150–600 mAs) on a maximum field size (up 40×40 cm). Removal of the typical effective doses below 5%-percentile and above 95%-percentile of typical effective dose distributions for all examinations would result in a reduction of a mean effective dose by up to 30% and reduction of a
75%-percentile of the distributions by up to 15%. No significant differences between the distributions of TED for analogue and digital X-ray units were observed for the pooled sample for selected examinations except for
the examination of the chest in posterior-anterior projection. Conclusions: These results should be considered in the process of establishing and implementing DRLs as well as in the cost-benefit analysis of the optimization in radiography (Less)
Please use this url to cite or link to this publication:
author
; ; ; and
alternative title
Evaluation of levels of exposure of adult patients from common radiographic examinations in the Russian Federation in 2009–2014
publishing date
type
Contribution to journal
publication status
published
subject
keywords
optimization, radiography, Effective dose
in
Radiatsionnaya Gygiena
volume
10
issue
3
pages
10 pages
publisher
Saint-Petersburg Research Institute of Radiation Hygiene after Professor P.V. Ramzaev
external identifiers
  • scopus:85035224379
ISSN
1998-426X
DOI
10.21514/1998-426X-2017-10-3-66-75
language
Russian
LU publication?
no
id
11152243-f7bd-4222-a26f-a0b7fd33d61f
date added to LUP
2018-01-04 12:12:58
date last changed
2022-04-25 04:50:58
@article{11152243-f7bd-4222-a26f-a0b7fd33d61f,
  abstract     = {{Diagnostic reference levels are the main and the most effective tools of optimization of the radiation protection of patients from medical exposure. Diagnostic reference levels should be established based on the<br/>results of dedicated dose surveys, allowing evaluating typical patient dose distributions in a selected dose quantity for the selected X-ray examinations. The aim of the current study was to assess the distributions of typical effective doses in representative Russian regions. Materials and methods: Typical patient effective doses for the 13 most common radiographic X-ray examinations were collected in 203 X-ray rooms in 101 hospitals in six regions of Russian Federation in 2009–2014. A differentiated approach was used for the<br/>estimation of the typical effective doses depending on the image acquisition technology. Effective doses were estimated using «EDEREX» (Russia) computational software. Results and discussion: Results of the dose<br/>data analysis indicate the lack of significant differences between the distributions of the typical effective doses between the selected regions, allowing merging the regional samples and further evaluating the pooled (joint) sample. A significant ratio of maximum to minimum (up to two orders of magnitude) due to a presence<br/>of X-ray units with abnormally high and low typical effective doses was observed for all 13 selected X-ray examinations. Abnormally high typical effective doses can be explained by performing the examinations using high values of tube current-time product (150–600 mAs) on a maximum field size (up 40×40 cm). Removal of the typical effective doses below 5%-percentile and above 95%-percentile of typical effective dose distributions for all examinations would result in a reduction of a mean effective dose by up to 30% and reduction of a<br/>75%-percentile of the distributions by up to 15%. No significant differences between the distributions of TED for analogue and digital X-ray units were observed for the pooled sample for selected examinations except for<br/>the examination of the chest in posterior-anterior projection. Conclusions: These results should be considered in the process of establishing and implementing DRLs as well as in the cost-benefit analysis of the optimization in radiography}},
  author       = {{Vodovatov, Aleksandr and Golikov, Vladislav and Kalnitsky, Sergey and Shatskiy, Iliya and Chipiga, Larisa}},
  issn         = {{1998-426X}},
  keywords     = {{optimization; radiography; Effective dose}},
  language     = {{rus}},
  number       = {{3}},
  pages        = {{66--75}},
  publisher    = {{Saint-Petersburg Research Institute of Radiation Hygiene after Professor P.V. Ramzaev}},
  series       = {{Radiatsionnaya Gygiena}},
  title        = {{Анализ уровней облучения взрослых пациентов при проведении наиболее распространенных рентгенографических исследований в Российской Федерации в 2009–2014 гг.}},
  url          = {{https://lup.lub.lu.se/search/files/36213742/518_1109_1_SM.pdf}},
  doi          = {{10.21514/1998-426X-2017-10-3-66-75}},
  volume       = {{10}},
  year         = {{2017}},
}