Verification of dynamic radiotherapy: the potential for 3D dosimetry under respiratory-like motion using polymer gel.

Ceberg, Sofie LU ; Karlsson, Anna LU ; Gustafsson, Helen LU ; Wittgren, Lena LU and Bäck, Sven LU (2008) In Physics in Medicine and Biology 53(20). p.387-396
Abstract
Following the implementation of advanced treatment procedures in radiotherapy, there is a need for dynamic dose verification in 3D. Gel dosimetry could potentially be used for such measurements. However, recently published data show that certain types of gels have a dose rate and fractionation dependence. The aim of this study was to investigate the feasibility of using a polymer gel dosimeter for dose verification of dynamic radiotherapy. To investigate the influence of dose rate dependence during respiratory-like motion in and out of the beam, a respiration robot together with two types of gel systems (normoxic methacrylic acid gel (nMAG) and normoxic polyacrylamide gel (nPAG)) were used. Reference measurements were obtained using a... (More)
Following the implementation of advanced treatment procedures in radiotherapy, there is a need for dynamic dose verification in 3D. Gel dosimetry could potentially be used for such measurements. However, recently published data show that certain types of gels have a dose rate and fractionation dependence. The aim of this study was to investigate the feasibility of using a polymer gel dosimeter for dose verification of dynamic radiotherapy. To investigate the influence of dose rate dependence during respiratory-like motion in and out of the beam, a respiration robot together with two types of gel systems (normoxic methacrylic acid gel (nMAG) and normoxic polyacrylamide gel (nPAG)) were used. Reference measurements were obtained using a linear diode array (LDA). Expected results, if there was no influence of the dose rate variation, were calculated by convolving the static irradiated gel data with the motion function controlling the robot. To investigate the fractionation dependence, the gels were irradiated using gated and ungated deliveries. Magnetic resonance imaging was used to evaluate the absorbed dose response of the gel. The measured gel data coincided well with the LDA data. Also, the calculated data agreed well with the measured dynamic gel data, i.e. no dose rate dependence due to motion was observed. The difference in the R2 response for the gels receiving ungated and gated, i.e. fractionated, deliveries was less than 1% for the nPAG and 4% for the nMAG, for absorbed doses up to 2 Gy. The maximum difference was 1.2% for the nPAG and 9% for the nMAG, which occurred at the highest given dose (4 Gy). The investigated gels were found to be feasible detectors for dose measurements under respiratory-like motion. For dose verification of dynamic RT involving gated delivery, e.g. breathing-adapted radiotherapy, relative absorbed dose evaluation should be used in order to minimize the effects of fractionated irradiation. (Less)
Please use this url to cite or link to this publication:
author
; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Physics in Medicine and Biology
volume
53
issue
20
pages
387 - 396
publisher
IOP Publishing
external identifiers
  • wos:000259736400025
  • pmid:18827319
  • scopus:58149288142
ISSN
1361-6560
DOI
10.1088/0031-9155/53/20/N02
language
English
LU publication?
yes
additional info
The information about affiliations in this record was updated in December 2015. The record was previously connected to the following departments: Department of Radiology and Physiology, Malmö (013037000), Medical Radiation Physics, Malmö (013243210), Emergency medicine/Medicine/Surgery (013240200)
id
4063e6f3-c6e1-4bfa-a622-8d1095d85307 (old id 1262660)
alternative location
http://www.ncbi.nlm.nih.gov/pubmed/18827319?dopt=Abstract
date added to LUP
2016-04-04 08:23:33
date last changed
2022-02-05 23:49:16
@article{4063e6f3-c6e1-4bfa-a622-8d1095d85307,
  abstract     = {{Following the implementation of advanced treatment procedures in radiotherapy, there is a need for dynamic dose verification in 3D. Gel dosimetry could potentially be used for such measurements. However, recently published data show that certain types of gels have a dose rate and fractionation dependence. The aim of this study was to investigate the feasibility of using a polymer gel dosimeter for dose verification of dynamic radiotherapy. To investigate the influence of dose rate dependence during respiratory-like motion in and out of the beam, a respiration robot together with two types of gel systems (normoxic methacrylic acid gel (nMAG) and normoxic polyacrylamide gel (nPAG)) were used. Reference measurements were obtained using a linear diode array (LDA). Expected results, if there was no influence of the dose rate variation, were calculated by convolving the static irradiated gel data with the motion function controlling the robot. To investigate the fractionation dependence, the gels were irradiated using gated and ungated deliveries. Magnetic resonance imaging was used to evaluate the absorbed dose response of the gel. The measured gel data coincided well with the LDA data. Also, the calculated data agreed well with the measured dynamic gel data, i.e. no dose rate dependence due to motion was observed. The difference in the R2 response for the gels receiving ungated and gated, i.e. fractionated, deliveries was less than 1% for the nPAG and 4% for the nMAG, for absorbed doses up to 2 Gy. The maximum difference was 1.2% for the nPAG and 9% for the nMAG, which occurred at the highest given dose (4 Gy). The investigated gels were found to be feasible detectors for dose measurements under respiratory-like motion. For dose verification of dynamic RT involving gated delivery, e.g. breathing-adapted radiotherapy, relative absorbed dose evaluation should be used in order to minimize the effects of fractionated irradiation.}},
  author       = {{Ceberg, Sofie and Karlsson, Anna and Gustafsson, Helen and Wittgren, Lena and Bäck, Sven}},
  issn         = {{1361-6560}},
  language     = {{eng}},
  number       = {{20}},
  pages        = {{387--396}},
  publisher    = {{IOP Publishing}},
  series       = {{Physics in Medicine and Biology}},
  title        = {{Verification of dynamic radiotherapy: the potential for 3D dosimetry under respiratory-like motion using polymer gel.}},
  url          = {{http://dx.doi.org/10.1088/0031-9155/53/20/N02}},
  doi          = {{10.1088/0031-9155/53/20/N02}},
  volume       = {{53}},
  year         = {{2008}},
}