Low-dose computed tomography of the paranasal sinuses: radiation doses and reliability analysis.

Abul-Kasim, Kasim LU ; Strömbeck, Anita and Sahlstrand Johnson, Pernilla LU (2011) In American Journal of Otolaryngology 32. p.47-51
Abstract
PURPOSE: The study aimed to (1) optimize the radiation doses of computed tomography (CT) of paranasal sinuses, (2) compare the radiation doses of different CT protocols with that of plain radiography, and (3) evaluate the reliability of low-dose CT in the detection of pathology and characterization of the detected pathology. MATERIALS AND METHODS: A head phantom was examined with different scan parameters to define a cutoff value to which the radiation dose can be reduced without negative impact on image quality. Kruskal-Wallis test and Wilcoxon W test were performed to compare the effective doses of the plain radiography in 30 patients with that of 3 different CT protocols in a total of 90 patients. The interobserver and intraobserver... (More)
PURPOSE: The study aimed to (1) optimize the radiation doses of computed tomography (CT) of paranasal sinuses, (2) compare the radiation doses of different CT protocols with that of plain radiography, and (3) evaluate the reliability of low-dose CT in the detection of pathology and characterization of the detected pathology. MATERIALS AND METHODS: A head phantom was examined with different scan parameters to define a cutoff value to which the radiation dose can be reduced without negative impact on image quality. Kruskal-Wallis test and Wilcoxon W test were performed to compare the effective doses of the plain radiography in 30 patients with that of 3 different CT protocols in a total of 90 patients. The interobserver and intraobserver agreement in the detection of pathologic findings and in characterization of the pathology was estimated by calculating kappa value. RESULTS: The effective doses of plain radiography and low-dose CT were 0.098 and 0.045 mSv, respectively (P < .001). The effective dose of standard CT of sinuses (0.371 mSv) was 3.8 times higher than that of plain radiography and 8.2 times higher than that of low-dose CT (P < .001). The interobserver and intraobserver agreement on CT with regard to detection of pathology and pathology characterization was almost perfect (kappa values 0.81-1) compared to fair (kappa values 0.38-0.39) in plain radiography. CONCLUSIONS: The here proposed low-dose CT means significant dose reduction and is a reliable method in the investigation of the paranasal sinuses. (Less)
Please use this url to cite or link to this publication:
author
; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
American Journal of Otolaryngology
volume
32
pages
47 - 51
publisher
W.B. Saunders
external identifiers
  • wos:000285370200009
  • pmid:20015806
  • scopus:78649738328
  • pmid:20015806
ISSN
1532-818X
DOI
10.1016/j.amjoto.2009.08.004
language
English
LU publication?
yes
id
ad8ffa66-1398-437d-81e6-a169a140a44e (old id 1523555)
alternative location
http://www.ncbi.nlm.nih.gov/pubmed/20015806?dopt=Abstract
date added to LUP
2016-04-04 08:43:52
date last changed
2022-01-29 03:53:44
@article{ad8ffa66-1398-437d-81e6-a169a140a44e,
  abstract     = {{PURPOSE: The study aimed to (1) optimize the radiation doses of computed tomography (CT) of paranasal sinuses, (2) compare the radiation doses of different CT protocols with that of plain radiography, and (3) evaluate the reliability of low-dose CT in the detection of pathology and characterization of the detected pathology. MATERIALS AND METHODS: A head phantom was examined with different scan parameters to define a cutoff value to which the radiation dose can be reduced without negative impact on image quality. Kruskal-Wallis test and Wilcoxon W test were performed to compare the effective doses of the plain radiography in 30 patients with that of 3 different CT protocols in a total of 90 patients. The interobserver and intraobserver agreement in the detection of pathologic findings and in characterization of the pathology was estimated by calculating kappa value. RESULTS: The effective doses of plain radiography and low-dose CT were 0.098 and 0.045 mSv, respectively (P &lt; .001). The effective dose of standard CT of sinuses (0.371 mSv) was 3.8 times higher than that of plain radiography and 8.2 times higher than that of low-dose CT (P &lt; .001). The interobserver and intraobserver agreement on CT with regard to detection of pathology and pathology characterization was almost perfect (kappa values 0.81-1) compared to fair (kappa values 0.38-0.39) in plain radiography. CONCLUSIONS: The here proposed low-dose CT means significant dose reduction and is a reliable method in the investigation of the paranasal sinuses.}},
  author       = {{Abul-Kasim, Kasim and Strömbeck, Anita and Sahlstrand Johnson, Pernilla}},
  issn         = {{1532-818X}},
  language     = {{eng}},
  pages        = {{47--51}},
  publisher    = {{W.B. Saunders}},
  series       = {{American Journal of Otolaryngology}},
  title        = {{Low-dose computed tomography of the paranasal sinuses: radiation doses and reliability analysis.}},
  url          = {{http://dx.doi.org/10.1016/j.amjoto.2009.08.004}},
  doi          = {{10.1016/j.amjoto.2009.08.004}},
  volume       = {{32}},
  year         = {{2011}},
}