The Artelon CMC spacer compared with tendon interposition arthroplasty

Nilsson, Anders ; Wiig, Monica ; Alnehill, Hakan ; Berggren, Magnus ; Bjornum, Sten ; Geijer, Mats ; Kopylov, Philippe LU and Sollerman, Christer (2010) In Acta Orthopaedica 81(2). p.239-246
Abstract
Background and purpose The Artelon CMC spacer is designed for surgical treatment of osteoarthritis (OA) in the carpometacarpal joint of the thumb (CMC-I). Good results using this degradable device were previously presented in a pilot study. We now present results from a larger randomized, controlled, multicenter study. Patients and methods 109 patients (94 females) with a mean age of 60 (42-83) years, suffering from painful CMC OA, were included in the study at 7 centers in Sweden. The patients were randomized to Artelon CMC spacer (test, n = 72) or tendon arthroplasty (control, n = 37) at a ratio of 2:1. Perceived pain was recorded on a visual analog scale (VAS) before treatment and after 3, 6, and 12 months, when measuring maximal tripod... (More)
Background and purpose The Artelon CMC spacer is designed for surgical treatment of osteoarthritis (OA) in the carpometacarpal joint of the thumb (CMC-I). Good results using this degradable device were previously presented in a pilot study. We now present results from a larger randomized, controlled, multicenter study. Patients and methods 109 patients (94 females) with a mean age of 60 (42-83) years, suffering from painful CMC OA, were included in the study at 7 centers in Sweden. The patients were randomized to Artelon CMC spacer (test, n = 72) or tendon arthroplasty (control, n = 37) at a ratio of 2:1. Perceived pain was recorded on a visual analog scale (VAS) before treatment and after 3, 6, and 12 months, when measuring maximal tripod pinch strength (primary outcome measure). In addition, range of motion, radiographic findings, and functional testing were recorded pre- and postoperatively. Results Swelling and pain were more common in the test group and 6 implants were removed because of such symptoms. 5 of these patients did not receive antibiotics preoperatively according to the study protocol. In a per-protocol analysis, i.e. patients without signs of concomitant OA in the scaphoid-trapezium-trapezoid (STT) joint and those in the test group who received antibiotics, the mean difference in tripod pinch strength increase, adjusted for baseline, was 1.4 kg in favor of the test group (not statistically significant). Statistically significant pain relief was achieved in both groups, with perceived pain gradually decreasing during the follow-up period. In the intention-to-treat analysis but not in the per-protocol analysis, significantly better pain relief (VAS) was obtained in the control group. Patient-perceived disability evaluated by the DASH questionnaire improved in both groups. Interpretation The Artelon CMC spacer did not show superior results compared to tendon interposition arthroplasty. Proper use of preoperative antibiotics and a thorough patient selection appear to be important for the results. (Less)
Please use this url to cite or link to this publication:
author
; ; ; ; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Acta Orthopaedica
volume
81
issue
2
pages
239 - 246
publisher
Taylor & Francis
external identifiers
  • wos:000276361700015
  • scopus:77950571529
  • pmid:20180717
ISSN
1745-3682
DOI
10.3109/17453671003635835
language
English
LU publication?
yes
id
92e79194-1955-48e6-8ae9-636c6625719b (old id 1604000)
date added to LUP
2016-04-01 13:32:06
date last changed
2022-01-27 19:41:59
@article{92e79194-1955-48e6-8ae9-636c6625719b,
  abstract     = {{Background and purpose The Artelon CMC spacer is designed for surgical treatment of osteoarthritis (OA) in the carpometacarpal joint of the thumb (CMC-I). Good results using this degradable device were previously presented in a pilot study. We now present results from a larger randomized, controlled, multicenter study. Patients and methods 109 patients (94 females) with a mean age of 60 (42-83) years, suffering from painful CMC OA, were included in the study at 7 centers in Sweden. The patients were randomized to Artelon CMC spacer (test, n = 72) or tendon arthroplasty (control, n = 37) at a ratio of 2:1. Perceived pain was recorded on a visual analog scale (VAS) before treatment and after 3, 6, and 12 months, when measuring maximal tripod pinch strength (primary outcome measure). In addition, range of motion, radiographic findings, and functional testing were recorded pre- and postoperatively. Results Swelling and pain were more common in the test group and 6 implants were removed because of such symptoms. 5 of these patients did not receive antibiotics preoperatively according to the study protocol. In a per-protocol analysis, i.e. patients without signs of concomitant OA in the scaphoid-trapezium-trapezoid (STT) joint and those in the test group who received antibiotics, the mean difference in tripod pinch strength increase, adjusted for baseline, was 1.4 kg in favor of the test group (not statistically significant). Statistically significant pain relief was achieved in both groups, with perceived pain gradually decreasing during the follow-up period. In the intention-to-treat analysis but not in the per-protocol analysis, significantly better pain relief (VAS) was obtained in the control group. Patient-perceived disability evaluated by the DASH questionnaire improved in both groups. Interpretation The Artelon CMC spacer did not show superior results compared to tendon interposition arthroplasty. Proper use of preoperative antibiotics and a thorough patient selection appear to be important for the results.}},
  author       = {{Nilsson, Anders and Wiig, Monica and Alnehill, Hakan and Berggren, Magnus and Bjornum, Sten and Geijer, Mats and Kopylov, Philippe and Sollerman, Christer}},
  issn         = {{1745-3682}},
  language     = {{eng}},
  number       = {{2}},
  pages        = {{239--246}},
  publisher    = {{Taylor & Francis}},
  series       = {{Acta Orthopaedica}},
  title        = {{The Artelon CMC spacer compared with tendon interposition arthroplasty}},
  url          = {{http://dx.doi.org/10.3109/17453671003635835}},
  doi          = {{10.3109/17453671003635835}},
  volume       = {{81}},
  year         = {{2010}},
}