Skip to main content

Lund University Publications

LUND UNIVERSITY LIBRARIES

Evaluation of aerodynamic particle sizer and electrical low-pressure impactor for unimodal and bimodal mass-weighted size distributions

Pagels, Joakim LU ; Gudmundsson, Anders LU ; Gustavsson, Eva ; Asking, Lars and Bohgard, Mats LU (2005) In Aerosol Science and Technology 39(9). p.871-887
Abstract
The objective of this study was to investigate the feasibility of the Aerodynamic Particle Sizer (APS) and the Electrical Low-Pressure Impactor (ELPI) to study mass weighted particle size distributions. Unimodal and bimodal liquid test aerosols were produced to a small chamber. Simultaneous measurements were performed with an APS 3320, an APS 3321, an ELPI and a Dekati Low-Pressure Impactor (DLPI) analyzed gravimetrically. ELPI current and mass responses were simulated for lognormal size distributions using a parameterization of the impactor kernel functions. In experiments with a single coarse mode, the mass ratio to the DLPI was between 0.75 and 1.15 for both APS models up to 5 mu m and for the ELPI up to 3 mu m. For larger sizes the... (More)
The objective of this study was to investigate the feasibility of the Aerodynamic Particle Sizer (APS) and the Electrical Low-Pressure Impactor (ELPI) to study mass weighted particle size distributions. Unimodal and bimodal liquid test aerosols were produced to a small chamber. Simultaneous measurements were performed with an APS 3320, an APS 3321, an ELPI and a Dekati Low-Pressure Impactor (DLPI) analyzed gravimetrically. ELPI current and mass responses were simulated for lognormal size distributions using a parameterization of the impactor kernel functions. In experiments with a single coarse mode, the mass ratio to the DLPI was between 0.75 and 1.15 for both APS models up to 5 mu m and for the ELPI up to 3 mu m. For larger sizes the ELPI and APS 3320 overestimated and the APS 3321 underestimated the concentration. In experiments with a single fine mode, submicrometer ELPI and DLPI results were in good agreement. However, in contrast to the DLPI all three spectrometers showed a significant mass fraction above 1 mu m. In experiments with a bimodal size distribution, the mass ratios were altered compared to single coarse mode experiments. Simulations showed that uncertainties in ELPI measurements of larger particles occur when concentrations of small particles are high. Several mechanisms that may bias ELPI and APS measurements are described. With knowledge of these, ELPI and APS 3321 can, under many circumstances give accurate time-resolved mass size distributions for particles smaller than 3 and 5 mu m, respectively. (Less)
Please use this url to cite or link to this publication:
author
; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Aerosol Science and Technology
volume
39
issue
9
pages
871 - 887
publisher
Taylor & Francis
external identifiers
  • wos:000233287000006
  • scopus:27744548052
ISSN
1521-7388
DOI
10.1080/02786820500295677
language
English
LU publication?
yes
id
b2416690-4ca4-40c8-8eff-ac20674060f6 (old id 213568)
date added to LUP
2016-04-01 11:53:50
date last changed
2022-03-28 17:17:23
@article{b2416690-4ca4-40c8-8eff-ac20674060f6,
  abstract     = {{The objective of this study was to investigate the feasibility of the Aerodynamic Particle Sizer (APS) and the Electrical Low-Pressure Impactor (ELPI) to study mass weighted particle size distributions. Unimodal and bimodal liquid test aerosols were produced to a small chamber. Simultaneous measurements were performed with an APS 3320, an APS 3321, an ELPI and a Dekati Low-Pressure Impactor (DLPI) analyzed gravimetrically. ELPI current and mass responses were simulated for lognormal size distributions using a parameterization of the impactor kernel functions. In experiments with a single coarse mode, the mass ratio to the DLPI was between 0.75 and 1.15 for both APS models up to 5 mu m and for the ELPI up to 3 mu m. For larger sizes the ELPI and APS 3320 overestimated and the APS 3321 underestimated the concentration. In experiments with a single fine mode, submicrometer ELPI and DLPI results were in good agreement. However, in contrast to the DLPI all three spectrometers showed a significant mass fraction above 1 mu m. In experiments with a bimodal size distribution, the mass ratios were altered compared to single coarse mode experiments. Simulations showed that uncertainties in ELPI measurements of larger particles occur when concentrations of small particles are high. Several mechanisms that may bias ELPI and APS measurements are described. With knowledge of these, ELPI and APS 3321 can, under many circumstances give accurate time-resolved mass size distributions for particles smaller than 3 and 5 mu m, respectively.}},
  author       = {{Pagels, Joakim and Gudmundsson, Anders and Gustavsson, Eva and Asking, Lars and Bohgard, Mats}},
  issn         = {{1521-7388}},
  language     = {{eng}},
  number       = {{9}},
  pages        = {{871--887}},
  publisher    = {{Taylor & Francis}},
  series       = {{Aerosol Science and Technology}},
  title        = {{Evaluation of aerodynamic particle sizer and electrical low-pressure impactor for unimodal and bimodal mass-weighted size distributions}},
  url          = {{http://dx.doi.org/10.1080/02786820500295677}},
  doi          = {{10.1080/02786820500295677}},
  volume       = {{39}},
  year         = {{2005}},
}