Fibrinogen measurement in cardiac surgery with cardiopulmonary bypass: Analysis of repeatability and agreement of Clauss method within and between six different laboratories

Solomon, Cristina ; Baryshnikova, Ekaterina ; Tripodi, Armando ; Schlimp, Christoph J. ; Schoechl, Herbert ; Cadamuro, Janne ; Winstedt, Dag LU ; Asmis, Lars and Ranucci, Marco (2014) In Thrombosis and Haemostasis 112(1). p.109-117
Abstract
Plasma fibrinogen concentration is important for coagulopathy assessment, and is most commonly measured using the Clauss method. Several factors, including device type and reagent, have been shown to affect results, The study objective was to evaluate performance and repeatability of the Clauss method and to assess differences between measurements performed during and after cardiopulmonary bypass (CPB), by testing plasma samples from patients undergoing cardiac surgery with CPB. Samples were collected from 30 patients before surgery, approximately 20 minutes before weaning from CPB, and 5 minutes after CPB and protamine. Fibrinogen concentration was determined using the Clauss method at six quality-controlled specialised laboratories,... (More)
Plasma fibrinogen concentration is important for coagulopathy assessment, and is most commonly measured using the Clauss method. Several factors, including device type and reagent, have been shown to affect results, The study objective was to evaluate performance and repeatability of the Clauss method and to assess differences between measurements performed during and after cardiopulmonary bypass (CPB), by testing plasma samples from patients undergoing cardiac surgery with CPB. Samples were collected from 30 patients before surgery, approximately 20 minutes before weaning from CPB, and 5 minutes after CPB and protamine. Fibrinogen concentration was determined using the Clauss method at six quality-controlled specialised laboratories, according to accredited standard operating procedures. Regarding within-centre agreement for Clauss measurement, mean differences between duplicate measurements were between 0.00 g/l and 0.15 g/l, with intervals for 95% limits of agreement for mean Bland-Altman differences up to 1.3 g/l. Regarding between-centre agreement, some mean differences between pairs of centres were above 0.5 g/l. Differences of up to similar to 2 g/l were observed with individual samples. Increased variability was observed between centres, with inter-class correlation values below 0.5 suggesting only fair agreement. There were no significant differences in fibrinogen concentration before weaning from CPB and after CPB for most centres and methods. In conclusion, considerable differences exist between Clauss-based plasma fibrinogen measured using different detection methods. Nevertheless, the similarity between measurements shortly before weaning from CPB and after CPB within centres suggests that on-pump measurements could provide an early estimation of fibrinogen deficit after CPB and thus guidance for haemostatic therapy. (Less)
Please use this url to cite or link to this publication:
author
; ; ; ; ; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
Blood coagulation tests, clinical laboratory techniques, fibrinogen, nephelometry and turbidimetry, patient care management
in
Thrombosis and Haemostasis
volume
112
issue
1
pages
109 - 117
publisher
Schattauer GmbH
external identifiers
  • wos:000338820300016
  • scopus:84903769307
  • pmid:24633448
ISSN
0340-6245
DOI
10.1160/TH13-12-0997
project
Koagulation vid kirurgi och kritisk sjukdom
language
English
LU publication?
yes
id
da31135f-29bb-47f6-b628-6c97c6a81d56 (old id 4601925)
date added to LUP
2016-04-01 13:10:16
date last changed
2022-02-11 19:40:17
@article{da31135f-29bb-47f6-b628-6c97c6a81d56,
  abstract     = {{Plasma fibrinogen concentration is important for coagulopathy assessment, and is most commonly measured using the Clauss method. Several factors, including device type and reagent, have been shown to affect results, The study objective was to evaluate performance and repeatability of the Clauss method and to assess differences between measurements performed during and after cardiopulmonary bypass (CPB), by testing plasma samples from patients undergoing cardiac surgery with CPB. Samples were collected from 30 patients before surgery, approximately 20 minutes before weaning from CPB, and 5 minutes after CPB and protamine. Fibrinogen concentration was determined using the Clauss method at six quality-controlled specialised laboratories, according to accredited standard operating procedures. Regarding within-centre agreement for Clauss measurement, mean differences between duplicate measurements were between 0.00 g/l and 0.15 g/l, with intervals for 95% limits of agreement for mean Bland-Altman differences up to 1.3 g/l. Regarding between-centre agreement, some mean differences between pairs of centres were above 0.5 g/l. Differences of up to similar to 2 g/l were observed with individual samples. Increased variability was observed between centres, with inter-class correlation values below 0.5 suggesting only fair agreement. There were no significant differences in fibrinogen concentration before weaning from CPB and after CPB for most centres and methods. In conclusion, considerable differences exist between Clauss-based plasma fibrinogen measured using different detection methods. Nevertheless, the similarity between measurements shortly before weaning from CPB and after CPB within centres suggests that on-pump measurements could provide an early estimation of fibrinogen deficit after CPB and thus guidance for haemostatic therapy.}},
  author       = {{Solomon, Cristina and Baryshnikova, Ekaterina and Tripodi, Armando and Schlimp, Christoph J. and Schoechl, Herbert and Cadamuro, Janne and Winstedt, Dag and Asmis, Lars and Ranucci, Marco}},
  issn         = {{0340-6245}},
  keywords     = {{Blood coagulation tests; clinical laboratory techniques; fibrinogen; nephelometry and turbidimetry; patient care management}},
  language     = {{eng}},
  number       = {{1}},
  pages        = {{109--117}},
  publisher    = {{Schattauer GmbH}},
  series       = {{Thrombosis and Haemostasis}},
  title        = {{Fibrinogen measurement in cardiac surgery with cardiopulmonary bypass: Analysis of repeatability and agreement of Clauss method within and between six different laboratories}},
  url          = {{http://dx.doi.org/10.1160/TH13-12-0997}},
  doi          = {{10.1160/TH13-12-0997}},
  volume       = {{112}},
  year         = {{2014}},
}