Development of an alkali chloride vapour-generating apparatus for calibration of ultraviolet absorption measurements

Leffler, T. LU ; Brackmann, C. LU ; Berg, M. LU ; Aldén, M. LU and Li, Z. S. LU (2017) In Review of Scientific Instruments 88(2).
Abstract

A novel design of alkali chloride vapour-generating cell has been developed, which can serve as a calibration cell for quantitative ultraviolet absorption concentration measurements and meticulous spectral investigations of alkali compounds. The calibration cell was designed to provide alkali vapour of well-controlled concentrations and temperatures, and consisted of a sealed quartz cell measuring 0.4 m in length with a temperature-controlled reservoir containing solid alkali salt. The cell was placed in a furnace and the alkali vapours generated from the reservoir have direct access to the measuring chamber. Investigations of potassium chloride (KCl) were made on sublimated vapour at temperatures 650, 700, 750, 780, and 800 °C while... (More)

A novel design of alkali chloride vapour-generating cell has been developed, which can serve as a calibration cell for quantitative ultraviolet absorption concentration measurements and meticulous spectral investigations of alkali compounds. The calibration cell was designed to provide alkali vapour of well-controlled concentrations and temperatures, and consisted of a sealed quartz cell measuring 0.4 m in length with a temperature-controlled reservoir containing solid alkali salt. The cell was placed in a furnace and the alkali vapours generated from the reservoir have direct access to the measuring chamber. Investigations of potassium chloride (KCl) were made on sublimated vapour at temperatures 650, 700, 750, 780, and 800 °C while the reservoir temperature was kept 50 °C lower to avoid condensation. The cell provides stable KCl vapour pressures, and the furnace provides a homogenous temperature profile along the cell. KCl vapour pressures are well characterised and conform the base for determination of the KCl concentration in the cell. The alkali chloride levels matched the concentration range of the absorption setup and indicated a previously employed calibration method to overestimate KCl concentrations. The KCl absorption cross sections for wavelengths λ=197.6 nm and λ=246.2 nm were calculated to be 3.4 × 10−17 and 2.9 × 10−17 cm2/molecule, respectively. The absorption cross section spectra did not show any structural differences with increasing temperature, which could indicate influence of dimers or significant changes of the population in the KCl vibrational states. The KCl absorption cross sections thus did not show any temperature dependence in the temperature region of 700-800 °C. Moreover, the applicability of the calibration cell for measurement of other alkali chlorides and hydroxides is discussed.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Review of Scientific Instruments
volume
88
issue
2
article number
023112
publisher
American Institute of Physics (AIP)
external identifiers
  • scopus:85013102350
  • pmid:28249467
  • wos:000395902700012
ISSN
0034-6748
DOI
10.1063/1.4975590
language
English
LU publication?
yes
id
b3dda922-1614-45c4-9775-4440c3733a29
date added to LUP
2017-03-01 12:03:18
date last changed
2024-04-14 06:15:30
@article{b3dda922-1614-45c4-9775-4440c3733a29,
  abstract     = {{<p>A novel design of alkali chloride vapour-generating cell has been developed, which can serve as a calibration cell for quantitative ultraviolet absorption concentration measurements and meticulous spectral investigations of alkali compounds. The calibration cell was designed to provide alkali vapour of well-controlled concentrations and temperatures, and consisted of a sealed quartz cell measuring 0.4 m in length with a temperature-controlled reservoir containing solid alkali salt. The cell was placed in a furnace and the alkali vapours generated from the reservoir have direct access to the measuring chamber. Investigations of potassium chloride (KCl) were made on sublimated vapour at temperatures 650, 700, 750, 780, and 800 °C while the reservoir temperature was kept 50 °C lower to avoid condensation. The cell provides stable KCl vapour pressures, and the furnace provides a homogenous temperature profile along the cell. KCl vapour pressures are well characterised and conform the base for determination of the KCl concentration in the cell. The alkali chloride levels matched the concentration range of the absorption setup and indicated a previously employed calibration method to overestimate KCl concentrations. The KCl absorption cross sections for wavelengths λ=197.6 nm and λ=246.2 nm were calculated to be 3.4 × 10<sup>−17</sup> and 2.9 × 10<sup>−17</sup> cm<sup>2</sup>/molecule, respectively. The absorption cross section spectra did not show any structural differences with increasing temperature, which could indicate influence of dimers or significant changes of the population in the KCl vibrational states. The KCl absorption cross sections thus did not show any temperature dependence in the temperature region of 700-800 °C. Moreover, the applicability of the calibration cell for measurement of other alkali chlorides and hydroxides is discussed.</p>}},
  author       = {{Leffler, T. and Brackmann, C. and Berg, M. and Aldén, M. and Li, Z. S.}},
  issn         = {{0034-6748}},
  language     = {{eng}},
  month        = {{02}},
  number       = {{2}},
  publisher    = {{American Institute of Physics (AIP)}},
  series       = {{Review of Scientific Instruments}},
  title        = {{Development of an alkali chloride vapour-generating apparatus for calibration of ultraviolet absorption measurements}},
  url          = {{http://dx.doi.org/10.1063/1.4975590}},
  doi          = {{10.1063/1.4975590}},
  volume       = {{88}},
  year         = {{2017}},
}