The vacuum ultraviolet scanning photoelectron microscope at MAX-LAB

Johansson, U. LU ; Nyholm, R. LU ; Törnevik, C. and Flodström, A. (1995) In Review of Scientific Instruments 66(2). p.1398-1400
Abstract

The performance of a scanning photoelectron microscope at the MAX I storage ring in Lund is presented. The microscope utilizes undulator radiation in the energy range 15-150 eV and is comprised of a plane-grating monochromator with a Kirkpatrick-Baez objective and a gracing incidence ellipsoidal focusing mirror. The instrument with its high photon flux, 109-1010 photons/s, and narrow bandwidth, better than 0.2 eV, is excellently suited for high-resolution core-level spectroscopy and imaging of samples with lateral inhomogeneties in the micrometer range.

Please use this url to cite or link to this publication:
author
; ; and
organization
publishing date
type
Contribution to journal
publication status
published
in
Review of Scientific Instruments
volume
66
issue
2
pages
3 pages
publisher
American Institute of Physics (AIP)
external identifiers
  • scopus:0001414102
ISSN
0034-6748
DOI
10.1063/1.1145986
language
English
LU publication?
yes
id
cda8af80-a90b-45c5-818e-a40e3e3b0411
date added to LUP
2016-04-29 11:01:53
date last changed
2021-01-03 03:22:11
@article{cda8af80-a90b-45c5-818e-a40e3e3b0411,
  abstract     = {{<p>The performance of a scanning photoelectron microscope at the MAX I storage ring in Lund is presented. The microscope utilizes undulator radiation in the energy range 15-150 eV and is comprised of a plane-grating monochromator with a Kirkpatrick-Baez objective and a gracing incidence ellipsoidal focusing mirror. The instrument with its high photon flux, 10<sup>9</sup>-10<sup>10</sup> photons/s, and narrow bandwidth, better than 0.2 eV, is excellently suited for high-resolution core-level spectroscopy and imaging of samples with lateral inhomogeneties in the micrometer range.</p>}},
  author       = {{Johansson, U. and Nyholm, R. and Törnevik, C. and Flodström, A.}},
  issn         = {{0034-6748}},
  language     = {{eng}},
  number       = {{2}},
  pages        = {{1398--1400}},
  publisher    = {{American Institute of Physics (AIP)}},
  series       = {{Review of Scientific Instruments}},
  title        = {{The vacuum ultraviolet scanning photoelectron microscope at MAX-LAB}},
  url          = {{http://dx.doi.org/10.1063/1.1145986}},
  doi          = {{10.1063/1.1145986}},
  volume       = {{66}},
  year         = {{1995}},
}