Determining Heart Activity Present in the Pressure Sensors of a Dialysis Machine

Holmer, Mattias; Grigonyte, Egle; Solem, Kristian; Olde, Bo, et al. (2013). Determining Heart Activity Present in the Pressure Sensors of a Dialysis Machine. Murray, Alan (Ed.). Computing in Cardiology, 217 - 220. Computing in Cardiology 2013. Zaragoza, Spain
Download:
URL:
Conference Proceeding/Paper | Published | English
Authors:
Holmer, Mattias ; Grigonyte, Egle ; Solem, Kristian ; Olde, Bo , et al.
Editors:
Murray, Alan
Department:
Department of Biomedical Engineering
Departments at LTH
Abstract:
Determination of heart status during dialysis can im- prove patient monitoring. Pressure sensors in the dialysis machine measures the heart pulses that propagates in the body and enter the extracorporeal blood circuit. A peri- staltic blood pump, located in the same circuit, introduces strong periodic pressure pulses that interfere with the much weaker cardiac component. These signal characteristics make the extraction of the heart activity challenging. In the present study, we explore the possibility to extract and analyze the cardiac component using simulated data. The accuracy of the timing of each heartbeat is analyzed. Ad- ditionally, the heart component is extracted from patient pressure recordings, and compared to the heart rate com- puted from a photoplethysmogram. The results show that heart timings can be accurately determined using the pres- sure sensors of a dialysis machine.