Optimization of Radiation Exposure and Image Quality of the Cone-beam O-arm Intraoperative Imaging System in Spinal Surgery.

Abul-Kasim, Kasim LU ; Söderberg, Marcus LU orcid ; Selariu, Eufrozina LU ; Gunnarsson, Mikael LU ; Kherad, Mehrsa LU and Ohlin, Acke LU (2012) In Journal of Spinal Disorders and Techniques 25(1). p.52-58
Abstract
STUDY DESIGN: Retrospective study. OBJECTIVES: To optimize the radiation doses and image quality for the cone-beam O-arm surgical imaging system in spinal surgery. SUMMARY OF BACKGROUND: Neurovascular compromise has been reported after screw misplacement during thoracic pedicle screw insertion. The use of O-arm with or without navigation system during spinal surgery has been shown to lower the rate of screw misplacement. The main drawback of such imaging surgical systems is the high radiation exposure. METHODS: Chest phantom and cadaveric pig spine were examined on the O-arm with different scan settings: 2 were recommended by the O-arm manufacturer (120 kV/320 mAs, and 120 kV/128 mAs), and 3 low-dose settings (80 kV/80 mAs, 80 kV/40 mAs,... (More)
STUDY DESIGN: Retrospective study. OBJECTIVES: To optimize the radiation doses and image quality for the cone-beam O-arm surgical imaging system in spinal surgery. SUMMARY OF BACKGROUND: Neurovascular compromise has been reported after screw misplacement during thoracic pedicle screw insertion. The use of O-arm with or without navigation system during spinal surgery has been shown to lower the rate of screw misplacement. The main drawback of such imaging surgical systems is the high radiation exposure. METHODS: Chest phantom and cadaveric pig spine were examined on the O-arm with different scan settings: 2 were recommended by the O-arm manufacturer (120 kV/320 mAs, and 120 kV/128 mAs), and 3 low-dose settings (80 kV/80 mAs, 80 kV/40 mAs, and 60 kV/40 mAs). The radiation doses were estimated by Monte Carlo calculations. Objective evaluation of image quality included interobserver agreement in the measurement of pedicular width in chest phantom and assessment of screw placement in cadaveric pig spine. RESULTS: The effective dose/cm for 120 kV/320 mAs scan was 13, 26, and 69 times higher than those delivered with 80 kV/80 mAs, 80 kV/40 mAs, and 60 kV/40 mAs scans, respectively. Images with 60 kV/40 mAs were unreliable. Images with 80 kV/80 mAs were considered reliable with good interobserver agreement when measuring the pedicular width (random error 0.38 mm and intraclass correlation coefficient 0.979) and almost perfect agreement when evaluating the screw placement (κ value 0.86). CONCLUSIONS: The radiation doses of the O-arm system can be reduced 5 to 13 times without negative impact on image quality with regard to information required for spinal surgery. (Less)
Please use this url to cite or link to this publication:
author
; ; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Journal of Spinal Disorders and Techniques
volume
25
issue
1
pages
52 - 58
publisher
Lippincott Williams & Wilkins
external identifiers
  • wos:000300046700009
  • pmid:21423057
  • scopus:84856752846
ISSN
1539-2465
DOI
10.1097/BSD.0b013e318211fdea
project
Image quality / radiation doses for CT examinations
Image quality / radiation doses at cone-beam CT (CBCT) examinations
language
English
LU publication?
yes
id
d69b4198-c03e-4329-8ea2-573ba5ec7c40 (old id 1883660)
alternative location
http://www.ncbi.nlm.nih.gov/pubmed/21423057?dopt=Abstract
date added to LUP
2016-04-04 09:17:33
date last changed
2022-05-16 23:48:51
@article{d69b4198-c03e-4329-8ea2-573ba5ec7c40,
  abstract     = {{STUDY DESIGN: Retrospective study. OBJECTIVES: To optimize the radiation doses and image quality for the cone-beam O-arm surgical imaging system in spinal surgery. SUMMARY OF BACKGROUND: Neurovascular compromise has been reported after screw misplacement during thoracic pedicle screw insertion. The use of O-arm with or without navigation system during spinal surgery has been shown to lower the rate of screw misplacement. The main drawback of such imaging surgical systems is the high radiation exposure. METHODS: Chest phantom and cadaveric pig spine were examined on the O-arm with different scan settings: 2 were recommended by the O-arm manufacturer (120 kV/320 mAs, and 120 kV/128 mAs), and 3 low-dose settings (80 kV/80 mAs, 80 kV/40 mAs, and 60 kV/40 mAs). The radiation doses were estimated by Monte Carlo calculations. Objective evaluation of image quality included interobserver agreement in the measurement of pedicular width in chest phantom and assessment of screw placement in cadaveric pig spine. RESULTS: The effective dose/cm for 120 kV/320 mAs scan was 13, 26, and 69 times higher than those delivered with 80 kV/80 mAs, 80 kV/40 mAs, and 60 kV/40 mAs scans, respectively. Images with 60 kV/40 mAs were unreliable. Images with 80 kV/80 mAs were considered reliable with good interobserver agreement when measuring the pedicular width (random error 0.38 mm and intraclass correlation coefficient 0.979) and almost perfect agreement when evaluating the screw placement (κ value 0.86). CONCLUSIONS: The radiation doses of the O-arm system can be reduced 5 to 13 times without negative impact on image quality with regard to information required for spinal surgery.}},
  author       = {{Abul-Kasim, Kasim and Söderberg, Marcus and Selariu, Eufrozina and Gunnarsson, Mikael and Kherad, Mehrsa and Ohlin, Acke}},
  issn         = {{1539-2465}},
  language     = {{eng}},
  number       = {{1}},
  pages        = {{52--58}},
  publisher    = {{Lippincott Williams & Wilkins}},
  series       = {{Journal of Spinal Disorders and Techniques}},
  title        = {{Optimization of Radiation Exposure and Image Quality of the Cone-beam O-arm Intraoperative Imaging System in Spinal Surgery.}},
  url          = {{http://dx.doi.org/10.1097/BSD.0b013e318211fdea}},
  doi          = {{10.1097/BSD.0b013e318211fdea}},
  volume       = {{25}},
  year         = {{2012}},
}