Initial tests of a new phantom for investigation of spatial resolution, partial volume effect and detectability in nuclear medicine tomography

Söderberg, Marcus LU orcid ; Engeland, Uwe ; Mattsson, Sören LU ; Gernot, Ebel and Leide Svegborn, Sigrid LU (2011) International Conference in Image Optimisation in Nuclear Medicine 317(012017). p.1-7
Abstract
Abstract in Undetermined
A new phantom has been designed that can provide simultaneously different target to background activity ratios with a linearly changing diameter of lesions. The purpose of the study was to describe and perform initial measurements with the phantom aimed to characterize different nuclear medicine tomographic systems and reconstruction algorithms in their performance and behaviour concerning partial volume effect (PVE) and detectability by varying the acquisition parameters and the count statistics. The phantom has an external vessel whose outline is half-cylindrical and allows it to be incorporated into an anthropomorphic thorax phantom. The phantom itself contains 16 fillable cones with an inner diameter... (More)
Abstract in Undetermined
A new phantom has been designed that can provide simultaneously different target to background activity ratios with a linearly changing diameter of lesions. The purpose of the study was to describe and perform initial measurements with the phantom aimed to characterize different nuclear medicine tomographic systems and reconstruction algorithms in their performance and behaviour concerning partial volume effect (PVE) and detectability by varying the acquisition parameters and the count statistics. The phantom has an external vessel whose outline is half-cylindrical and allows it to be incorporated into an anthropomorphic thorax phantom. The phantom itself contains 16 fillable cones with an inner diameter linearly decreasing from 16 mm to 2 mm and a wall thickness of 1 mm acrylic glass. They as well as the outer vessel were separately filled with (99m)Tc-and (18)F-solutions respectively of different activity concentrations. The phantom was easy to fill and air bubbles could easily be avoided. Images taken using a SPECT/CT and a PET/CT system are presented as well as evaluations of PVE. The new phantom seems to be useful for comparison and optimisation of different acquisition and reconstruction parameters in nuclear medicine tomographic studies and for comparisons between various tomographic units. (Less)
Please use this url to cite or link to this publication:
author
; ; ; and
organization
publishing date
type
Chapter in Book/Report/Conference proceeding
publication status
published
subject
host publication
Journal of Physics Conference Series
volume
317
issue
012017
pages
1 - 7
publisher
IOP Publishing
conference name
International Conference in Image Optimisation in Nuclear Medicine
conference location
Ayia Napa, Cyprus
conference dates
0001-01-02
external identifiers
  • wos:000299413100017
  • scopus:81355135535
DOI
10.1088/1742-6596/317/1/012017
language
English
LU publication?
yes
id
5ec42343-46f0-4b90-8a41-c329452830a7 (old id 2439999)
date added to LUP
2016-04-04 11:29:27
date last changed
2022-01-29 21:58:38
@inproceedings{5ec42343-46f0-4b90-8a41-c329452830a7,
  abstract     = {{Abstract in Undetermined<br/>A new phantom has been designed that can provide simultaneously different target to background activity ratios with a linearly changing diameter of lesions. The purpose of the study was to describe and perform initial measurements with the phantom aimed to characterize different nuclear medicine tomographic systems and reconstruction algorithms in their performance and behaviour concerning partial volume effect (PVE) and detectability by varying the acquisition parameters and the count statistics. The phantom has an external vessel whose outline is half-cylindrical and allows it to be incorporated into an anthropomorphic thorax phantom. The phantom itself contains 16 fillable cones with an inner diameter linearly decreasing from 16 mm to 2 mm and a wall thickness of 1 mm acrylic glass. They as well as the outer vessel were separately filled with (99m)Tc-and (18)F-solutions respectively of different activity concentrations. The phantom was easy to fill and air bubbles could easily be avoided. Images taken using a SPECT/CT and a PET/CT system are presented as well as evaluations of PVE. The new phantom seems to be useful for comparison and optimisation of different acquisition and reconstruction parameters in nuclear medicine tomographic studies and for comparisons between various tomographic units.}},
  author       = {{Söderberg, Marcus and Engeland, Uwe and Mattsson, Sören and Gernot, Ebel and Leide Svegborn, Sigrid}},
  booktitle    = {{Journal of Physics Conference Series}},
  language     = {{eng}},
  number       = {{012017}},
  pages        = {{1--7}},
  publisher    = {{IOP Publishing}},
  title        = {{Initial tests of a new phantom for investigation of spatial resolution, partial volume effect and detectability in nuclear medicine tomography}},
  url          = {{http://dx.doi.org/10.1088/1742-6596/317/1/012017}},
  doi          = {{10.1088/1742-6596/317/1/012017}},
  volume       = {{317}},
  year         = {{2011}},
}