Evaluation of a routine hematology analyzer for quality control of leukoreduced plasma

Petersson, Annika LU and Ekblom, Kim LU (2019) In Transfusion 59(10). p.3214-3218
Abstract

BACKGROUND: Quality control of residual white blood cells (WBCs) and red blood cells (RBCs) in leukoreduced plasma is mandatory. Although technological advances have been made, analysis of quality controls using routine hematology analyzers has not generally been introduced. The aim of this study was to evaluate if the routine hematology analyzer Sysmex XN-10, (Sysmex Nordic ApS) could be used for quality control of residual WBCs and RBCs in leukoreduced plasma. STUDY DESIGN AND METHODS: Linearity, accuracy, and precision were established for two Sysmex XN-10 analyzers using spiked donor plasma. ADAM rWBC (NanoEnTek) and manual counting in the Bürker chamber (NanoEnTek) were reference methods for WBCs and RBCs, respectively. Twenty-five... (More)

BACKGROUND: Quality control of residual white blood cells (WBCs) and red blood cells (RBCs) in leukoreduced plasma is mandatory. Although technological advances have been made, analysis of quality controls using routine hematology analyzers has not generally been introduced. The aim of this study was to evaluate if the routine hematology analyzer Sysmex XN-10, (Sysmex Nordic ApS) could be used for quality control of residual WBCs and RBCs in leukoreduced plasma. STUDY DESIGN AND METHODS: Linearity, accuracy, and precision were established for two Sysmex XN-10 analyzers using spiked donor plasma. ADAM rWBC (NanoEnTek) and manual counting in the Bürker chamber (NanoEnTek) were reference methods for WBCs and RBCs, respectively. Twenty-five consecutive leukoreduced donor plasma samples were also tested. RESULTS: For WBCs, the linearity criteria were met for the ADAM rWBC, but not for the Sysmex XN-10 instruments. Precision on both Sysmex XN-10 instruments was accurate only at 6 cells/μL, and accuracy was consistently acceptable only at 5 to 6 cells/μL. The precision and accuracy of the ADAM rWBC were acceptable at 2 to 6 cells/μL. For RBCs, both Sysmex XN-10 instruments and manual counting in the Bürker chamber were linear and fulfilled the precision criteria. Accuracy was acceptable for both Sysmex instruments at 6 to 12 × 109 WBCs/L but fluctuated within the study's measuring range for the Bürker chamber. No false-positive results were seen in the 25 consecutive donor plasma samples tested. CONCLUSION: For quality control purposes of leukoreduced plasma, the Sysmex XN-10 analyzer is suitable for the enumeration of residual RBCs but not of residual WBCs.

(Less)
Please use this url to cite or link to this publication:
author
and
publishing date
type
Contribution to journal
publication status
published
in
Transfusion
volume
59
issue
10
pages
3214 - 3218
publisher
Wiley-Blackwell
external identifiers
  • scopus:85070323823
  • pmid:31400008
ISSN
0041-1132
DOI
10.1111/trf.15481
language
English
LU publication?
no
additional info
Publisher Copyright: © 2019 AABB
id
094d3add-ffa3-4def-b584-7aadcd8ed2f0
date added to LUP
2026-02-17 14:11:59
date last changed
2026-02-18 07:43:36
@article{094d3add-ffa3-4def-b584-7aadcd8ed2f0,
  abstract     = {{<p>BACKGROUND: Quality control of residual white blood cells (WBCs) and red blood cells (RBCs) in leukoreduced plasma is mandatory. Although technological advances have been made, analysis of quality controls using routine hematology analyzers has not generally been introduced. The aim of this study was to evaluate if the routine hematology analyzer Sysmex XN-10, (Sysmex Nordic ApS) could be used for quality control of residual WBCs and RBCs in leukoreduced plasma. STUDY DESIGN AND METHODS: Linearity, accuracy, and precision were established for two Sysmex XN-10 analyzers using spiked donor plasma. ADAM rWBC (NanoEnTek) and manual counting in the Bürker chamber (NanoEnTek) were reference methods for WBCs and RBCs, respectively. Twenty-five consecutive leukoreduced donor plasma samples were also tested. RESULTS: For WBCs, the linearity criteria were met for the ADAM rWBC, but not for the Sysmex XN-10 instruments. Precision on both Sysmex XN-10 instruments was accurate only at 6 cells/μL, and accuracy was consistently acceptable only at 5 to 6 cells/μL. The precision and accuracy of the ADAM rWBC were acceptable at 2 to 6 cells/μL. For RBCs, both Sysmex XN-10 instruments and manual counting in the Bürker chamber were linear and fulfilled the precision criteria. Accuracy was acceptable for both Sysmex instruments at 6 to 12 × 10<sup>9</sup> WBCs/L but fluctuated within the study's measuring range for the Bürker chamber. No false-positive results were seen in the 25 consecutive donor plasma samples tested. CONCLUSION: For quality control purposes of leukoreduced plasma, the Sysmex XN-10 analyzer is suitable for the enumeration of residual RBCs but not of residual WBCs.</p>}},
  author       = {{Petersson, Annika and Ekblom, Kim}},
  issn         = {{0041-1132}},
  language     = {{eng}},
  month        = {{10}},
  number       = {{10}},
  pages        = {{3214--3218}},
  publisher    = {{Wiley-Blackwell}},
  series       = {{Transfusion}},
  title        = {{Evaluation of a routine hematology analyzer for quality control of leukoreduced plasma}},
  url          = {{http://dx.doi.org/10.1111/trf.15481}},
  doi          = {{10.1111/trf.15481}},
  volume       = {{59}},
  year         = {{2019}},
}