Methods for counting residual leukocytes in leukocyte-depleted plasma—a comparison between a routine hematology instrument, the Nageotte chamber, flow cytometry, and a fluorescent microscopy analyzer

Petersson, Annika LU and Ekblom, Kim LU (2017) In Transfusion 57(5). p.1192-1198
Abstract

BACKGROUND: Counting very low levels of leukocytes is technically challenging but mandatory for quality control of leukocyte-depleted plasma. Established assays, such as flow cytometry and counting in the Nageotte chamber, are laborious and expensive. The aim of this study was to test two alternative assays, the cerebrospinal fluid program in the routine hematology analyzer ADVIA 2120 and a fluorescence microscopy analyzer, the ADAM-rWBC. STUDY DESIGN AND METHODS: Linearity, accuracy, and precision were established for the ADVIA 2120, the ADAM-rWBC analyzer and the Nageotte chamber with flow cytometry as the reference method. Two hundred consecutive leukocyte-depleted donor plasma samples were also tested. RESULTS: The ADAM-rWBC... (More)

BACKGROUND: Counting very low levels of leukocytes is technically challenging but mandatory for quality control of leukocyte-depleted plasma. Established assays, such as flow cytometry and counting in the Nageotte chamber, are laborious and expensive. The aim of this study was to test two alternative assays, the cerebrospinal fluid program in the routine hematology analyzer ADVIA 2120 and a fluorescence microscopy analyzer, the ADAM-rWBC. STUDY DESIGN AND METHODS: Linearity, accuracy, and precision were established for the ADVIA 2120, the ADAM-rWBC analyzer and the Nageotte chamber with flow cytometry as the reference method. Two hundred consecutive leukocyte-depleted donor plasma samples were also tested. RESULTS: The ADAM-rWBC analyzer and the Nageotte chamber fulfilled all quality requirements. Flow cytometry fulfilled the requirements for linearity and precision. The ADVIA 2120 analyzer did not fully reach the quality criteria, and flow cytometry did not reach quality criteria on accuracy. No false-positive results on donor plasma samples were recorded. CONCLUSION: The ADAM-rWBC is suitable for the purpose of quality control of residual leukocytes in leukocyte-depleted plasma. For the ADVIA 2120, further improvements and studies are needed to reach the quality requirements stated in this study.

(Less)
Please use this url to cite or link to this publication:
author
and
publishing date
type
Contribution to journal
publication status
published
in
Transfusion
volume
57
issue
5
pages
1192 - 1198
publisher
Wiley-Blackwell
external identifiers
  • scopus:85010931791
  • pmid:28130849
ISSN
0041-1132
DOI
10.1111/trf.14007
language
English
LU publication?
no
additional info
Publisher Copyright: © 2017 AABB
id
988c00e5-1175-454d-90ca-d271158ddbc6
date added to LUP
2026-02-17 14:10:02
date last changed
2026-02-18 07:49:09
@article{988c00e5-1175-454d-90ca-d271158ddbc6,
  abstract     = {{<p>BACKGROUND: Counting very low levels of leukocytes is technically challenging but mandatory for quality control of leukocyte-depleted plasma. Established assays, such as flow cytometry and counting in the Nageotte chamber, are laborious and expensive. The aim of this study was to test two alternative assays, the cerebrospinal fluid program in the routine hematology analyzer ADVIA 2120 and a fluorescence microscopy analyzer, the ADAM-rWBC. STUDY DESIGN AND METHODS: Linearity, accuracy, and precision were established for the ADVIA 2120, the ADAM-rWBC analyzer and the Nageotte chamber with flow cytometry as the reference method. Two hundred consecutive leukocyte-depleted donor plasma samples were also tested. RESULTS: The ADAM-rWBC analyzer and the Nageotte chamber fulfilled all quality requirements. Flow cytometry fulfilled the requirements for linearity and precision. The ADVIA 2120 analyzer did not fully reach the quality criteria, and flow cytometry did not reach quality criteria on accuracy. No false-positive results on donor plasma samples were recorded. CONCLUSION: The ADAM-rWBC is suitable for the purpose of quality control of residual leukocytes in leukocyte-depleted plasma. For the ADVIA 2120, further improvements and studies are needed to reach the quality requirements stated in this study.</p>}},
  author       = {{Petersson, Annika and Ekblom, Kim}},
  issn         = {{0041-1132}},
  language     = {{eng}},
  number       = {{5}},
  pages        = {{1192--1198}},
  publisher    = {{Wiley-Blackwell}},
  series       = {{Transfusion}},
  title        = {{Methods for counting residual leukocytes in leukocyte-depleted plasma—a comparison between a routine hematology instrument, the Nageotte chamber, flow cytometry, and a fluorescent microscopy analyzer}},
  url          = {{http://dx.doi.org/10.1111/trf.14007}},
  doi          = {{10.1111/trf.14007}},
  volume       = {{57}},
  year         = {{2017}},
}