On-line Heart Rate Monitoring Using the Extra-corporeal Pressure Sensors of a Dialysis Machine

Holmer, Mattias; Grigonyte, Egle; Olde, Bo; Sternby, Jan, et al. (2012). On-line Heart Rate Monitoring Using the Extra-corporeal Pressure Sensors of a Dialysis Machine Biomedical engineering – 2012 : proceedings of international conference. Biomedical engineering – 2012. Kaunas, Lithuania
Download:
Conference Proceeding/Paper | Published | English
Authors:
Holmer, Mattias ; Grigonyte, Egle ; Olde, Bo ; Sternby, Jan , et al.
Department:
Department of Biomedical Engineering
Nephrology
Departments at LTH
Abstract:
Heart rate can be extracted from the extracorporeal venous pressure signal of a dialysis machine. The results are of comparable accuracy and reliability to the ones obtained by the PPG reference signal from a pulse oximeter. Difficulties that occur during heart rate estimation were determined, and some can be overcome by a slight adjustment of blood flow rate. The described techniques, after being implemented into the dialysis machines, would help to improve current hemodialysis safety throughout the treatment by on-line monitoring of the cardiac activity.