Method for selecting optimal operation frequencies in bulk acoustophoretic devices

Vitali, Valentina LU and Core, Giulia LU (2018) BMEM01 20182
Department of Biomedical Engineering
Abstract
We report a new method to select the optimal actuation frequencies in bulk acoustophoretic devices, by means of differential impedance spectral analysis measured via the actuating piezoelectric transducer (piezo). The impedance spectra are rich of large spurious resonance peaks originating from bulk resonances not related to the channel resonance why direct measurement of the piezo impedance spectra is not a viable strategy. In our work we present for the first time that the resonance modes of the acoustophoretic channel can be clearly identified by sequentially measuring the impedance spectra of the acoustophoretic manifold when the channel is filled with two different fluids and calculate the normalized differential impedance spectra.... (More)
We report a new method to select the optimal actuation frequencies in bulk acoustophoretic devices, by means of differential impedance spectral analysis measured via the actuating piezoelectric transducer (piezo). The impedance spectra are rich of large spurious resonance peaks originating from bulk resonances not related to the channel resonance why direct measurement of the piezo impedance spectra is not a viable strategy. In our work we present for the first time that the resonance modes of the acoustophoretic channel can be clearly identified by sequentially measuring the impedance spectra of the acoustophoretic manifold when the channel is filled with two different fluids and calculate the normalized differential impedance spectra. The method was tested on simple and more complex chip designs that are nowadays used in lab-on-a-chip applications with good results. The developed differential impedance spectra method opens the route to an automatic tuning of acoustophoretic devices. (Less)
Popular Abstract
An innovative method for selecting optimal working conditions in acoustophoretic devices
Moving cells with sound? It is actually possible! This phenomenon is called "Acoustophoresis". How do you set the sound? Is it ok to use songs by Bob Dylan? We are sorry to say no, but we have developed an innovative method to identify the optimal working conditions for acoustophoretic devices (devices that exploit "Acoustophoresis").
"Acoustophoresis" is a technique that induces the motion of particles suspended in a solution by sound. The solution flows inside a small channel that has dimensions in the order of micrometers, exactly like a human hair, with a thickness that varies between 18 and 180 µm. Moreover, the microchannel is dug in a bigger... (More)
An innovative method for selecting optimal working conditions in acoustophoretic devices
Moving cells with sound? It is actually possible! This phenomenon is called "Acoustophoresis". How do you set the sound? Is it ok to use songs by Bob Dylan? We are sorry to say no, but we have developed an innovative method to identify the optimal working conditions for acoustophoretic devices (devices that exploit "Acoustophoresis").
"Acoustophoresis" is a technique that induces the motion of particles suspended in a solution by sound. The solution flows inside a small channel that has dimensions in the order of micrometers, exactly like a human hair, with a thickness that varies between 18 and 180 µm. Moreover, the microchannel is dug in a bigger structure called "microchip'' and this microchip, in turn, is glued to a piezoelectric transducer (piezo). A piezo is a component that generates oscillations and therefore sound waves. Since the piezo is glued to the microchip, the piezo causes the microchip to oscillate and sound waves will travel in it. In order to be able to move particles present in a solution, the microchannel must be in specific conditions, known as resonance conditions. In resonance conditions, a lot of energy is present in the microchannel and forces, called acoustic radiation forces, are able to move particles in specific locations depending on their physical properties. In particular, when particles are moved to the center of the channel, we will refer to it as particles focusing. Particles can therefore be manipulated and, if they have different properties, can be also separated. "Acoustophoresis" has been used in interesting applications such as the separation of lipid particles from red blood cells after open-heart surgeries, and the detection of circulating tumor cells.
Resonance conditions are necessary to manipulate particles in acoustophoretic devices. Nowadays, what it is usually done to determine when the microchannel is in resonance conditions is to observe when the particles focus, using a microscope. This procedure is time consuming and operator dependent. In our project, we wanted to avoid the use of the microscope and to use other tools to identify the resonance of the microchannel. In particular, we used impedance measurements. The impedance is an electrical characteristic of a system and it is a measure of the opposition that a system presents to a current when a voltage is applied. It can be easily measured using a machine called impedance analyser. Using these measurements, we found that it is possible to identify when the channel is in resonance conditions avoiding the use of time consuming visual inspections. This result is great, since it opens the route to a fast and automatic tuning of acoustophoretic devices. (Less)
Please use this url to cite or link to this publication:
author
Vitali, Valentina LU and Core, Giulia LU
supervisor
organization
course
BMEM01 20182
year
type
H2 - Master's Degree (Two Years)
subject
keywords
Lab-on-a-chip, microfluidics, acoustophoresis, impedance, measurements, resonance
language
English
additional info
2018-17
id
8959982
date added to LUP
2018-10-17 07:43:30
date last changed
2018-10-26 15:11:47
@misc{8959982,
  abstract     = {{We report a new method to select the optimal actuation frequencies in bulk acoustophoretic devices, by means of differential impedance spectral analysis measured via the actuating piezoelectric transducer (piezo). The impedance spectra are rich of large spurious resonance peaks originating from bulk resonances not related to the channel resonance why direct measurement of the piezo impedance spectra is not a viable strategy. In our work we present for the first time that the resonance modes of the acoustophoretic channel can be clearly identified by sequentially measuring the impedance spectra of the acoustophoretic manifold when the channel is filled with two different fluids and calculate the normalized differential impedance spectra. The method was tested on simple and more complex chip designs that are nowadays used in lab-on-a-chip applications with good results. The developed differential impedance spectra method opens the route to an automatic tuning of acoustophoretic devices.}},
  author       = {{Vitali, Valentina and Core, Giulia}},
  language     = {{eng}},
  note         = {{Student Paper}},
  title        = {{Method for selecting optimal operation frequencies in bulk acoustophoretic devices}},
  year         = {{2018}},
}