Performance of a Personal Low-Weight Aerosol Sampler with Particle Size Fractionation and Time Resolution
(1983) The Tenth Annual Conference of the Association for Aerosol Research 14(3). p.409-413- Abstract
- A low-weight aerosol sampler has been developed and tested. Internal losses for both sub-micron and
large particles ( < 10 tam) are of the order of 5 %. The size fractionation patterns indicate that the sampler may be
used to collect a separate respirable fraction in good agreement with the BMRC standard and a coarse fraction
deternuned by the inlet design.
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/1762785
- author
- Malmqvist, Klas LU ; Bohgard, Mats LU ; Johansson, Gerd LU and Akselsson, Roland LU
- organization
- publishing date
- 1983
- type
- Chapter in Book/Report/Conference proceeding
- publication status
- published
- subject
- keywords
- aerosol sampler, size fractionation, time resolution
- host publication
- Journal of Aerosol Science
- volume
- 14
- issue
- 3
- pages
- 409 - 413
- conference name
- The Tenth Annual Conference of the Association for Aerosol Research
- conference location
- Bologna, Italy
- conference dates
- 1982-10-14 - 1982-10-17
- external identifiers
-
- scopus:0020543615
- DOI
- 10.1016/0021-8502(83)90149-0
- language
- English
- LU publication?
- yes
- additional info
- The information about affiliations in this record was updated in December 2015. The record was previously connected to the following departments: Ergonomics and Aerosol Technology (011025002), Division of Occupational and Environmental Medicine (013078001), Nuclear Physics (Faculty of Technology) (011013007)
- id
- bc844a52-f884-464e-addd-46093c5a3cec (old id 1762785)
- date added to LUP
- 2016-04-04 13:34:06
- date last changed
- 2021-01-03 10:34:16
@inproceedings{bc844a52-f884-464e-addd-46093c5a3cec, abstract = {{A low-weight aerosol sampler has been developed and tested. Internal losses for both sub-micron and<br/><br> large particles ( < 10 tam) are of the order of 5 %. The size fractionation patterns indicate that the sampler may be<br/><br> used to collect a separate respirable fraction in good agreement with the BMRC standard and a coarse fraction<br/><br> deternuned by the inlet design.}}, author = {{Malmqvist, Klas and Bohgard, Mats and Johansson, Gerd and Akselsson, Roland}}, booktitle = {{Journal of Aerosol Science}}, keywords = {{aerosol sampler; size fractionation; time resolution}}, language = {{eng}}, number = {{3}}, pages = {{409--413}}, title = {{Performance of a Personal Low-Weight Aerosol Sampler with Particle Size Fractionation and Time Resolution}}, url = {{http://dx.doi.org/10.1016/0021-8502(83)90149-0}}, doi = {{10.1016/0021-8502(83)90149-0}}, volume = {{14}}, year = {{1983}}, }