A sub-picosecond accumulating streak camera for x-rays

Naylor, G. A; Scheidt, K; Larsson, Jörgen LU ; Wulff, M and Filhol, J. M (2001) In Measurement Science & Technology 12(11). p.1858-1864
Abstract
An x-ray streak camera system uses a laser triggered photo-conductive switch to synchronize the high voltage ramp applied to its sweep plates to the photo-excitation of a sample. This technique allows the jitter. between successive sweeps to be kept below 100 fs compared to over 1 ps:using classical methods. By accumulating a stable pulsed x-ray signal over many sweeps, a very high dynamic range can be achieved whilst maintaining a sub-picosecond time resolution. The ultimate time resolution is limited by the dispersion of electrons in the streak tube. Whilst triggering with a laser at 900 Hz we have achieved a time resolution (FWHM) of 640 fs at a detection wavelength of 267 nm for a 60 s accumulation period, corresponding to over 50 000... (More)
An x-ray streak camera system uses a laser triggered photo-conductive switch to synchronize the high voltage ramp applied to its sweep plates to the photo-excitation of a sample. This technique allows the jitter. between successive sweeps to be kept below 100 fs compared to over 1 ps:using classical methods. By accumulating a stable pulsed x-ray signal over many sweeps, a very high dynamic range can be achieved whilst maintaining a sub-picosecond time resolution. The ultimate time resolution is limited by the dispersion of electrons in the streak tube. Whilst triggering with a laser at 900 Hz we have achieved a time resolution (FWHM) of 640 fs at a detection wavelength of 267 nm for a 60 s accumulation period, corresponding to over 50 000 sweeps. (Less)
Please use this url to cite or link to this publication:
author
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Measurement Science & Technology
volume
12
issue
11
pages
1858 - 1864
publisher
IOP Publishing
external identifiers
  • Scopus:0035504754
ISSN
0957-0233
DOI
10.1088/0957-0233/12/11/314
language
English
LU publication?
yes
id
c20509d5-e75c-4888-b864-f402313ba33b (old id 2259026)
date added to LUP
2012-02-17 23:01:51
date last changed
2016-10-13 04:29:42
@misc{c20509d5-e75c-4888-b864-f402313ba33b,
  abstract     = {An x-ray streak camera system uses a laser triggered photo-conductive switch to synchronize the high voltage ramp applied to its sweep plates to the photo-excitation of a sample. This technique allows the jitter. between successive sweeps to be kept below 100 fs compared to over 1 ps:using classical methods. By accumulating a stable pulsed x-ray signal over many sweeps, a very high dynamic range can be achieved whilst maintaining a sub-picosecond time resolution. The ultimate time resolution is limited by the dispersion of electrons in the streak tube. Whilst triggering with a laser at 900 Hz we have achieved a time resolution (FWHM) of 640 fs at a detection wavelength of 267 nm for a 60 s accumulation period, corresponding to over 50 000 sweeps.},
  author       = {Naylor, G. A and Scheidt, K and Larsson, Jörgen and Wulff, M and Filhol, J. M},
  issn         = {0957-0233},
  language     = {eng},
  number       = {11},
  pages        = {1858--1864},
  publisher    = {ARRAY(0x9d8d388)},
  series       = {Measurement Science & Technology},
  title        = {A sub-picosecond accumulating streak camera for x-rays},
  url          = {http://dx.doi.org/10.1088/0957-0233/12/11/314},
  volume       = {12},