MANAGEMENT OF PATIENT DOSES FROM DIGITAL X-RAY CHEST SCREENING EXAMINATIONS.

Vodovatov, A V LU ; Drozdov, A A ; Telnova, A U and Bernhardsson, Christian LU orcid (2016) In Radiation Protection Dosimetry 169(1-4). p.232-239
Abstract
An anthropomorphic phantom study was carried out in 2013-14 in two hospitals, one located in Russia (Mariinsky Hospital, Saint Petersburg) and the other in Sweden (Skåne University Hospital, Malmö). The aim of the study was to investigate the possibilities to reduce the patient dose from digital X-ray chest screening examinations. The existing chest imaging protocols were adjusted by changing the tube voltage, total filtration and grid in order to determine the most dose-effective combination of the examination parameters. It was possible to achieve up to 50 % dose-area product (DAP) and 30 % effective dose reduction by raising the tube voltage from 100 to 125 or 150 kV, and simultaneously decrease the total filtration to the minimum... (More)
An anthropomorphic phantom study was carried out in 2013-14 in two hospitals, one located in Russia (Mariinsky Hospital, Saint Petersburg) and the other in Sweden (Skåne University Hospital, Malmö). The aim of the study was to investigate the possibilities to reduce the patient dose from digital X-ray chest screening examinations. The existing chest imaging protocols were adjusted by changing the tube voltage, total filtration and grid in order to determine the most dose-effective combination of the examination parameters. It was possible to achieve up to 50 % dose-area product (DAP) and 30 % effective dose reduction by raising the tube voltage from 100 to 125 or 150 kV, and simultaneously decrease the total filtration to the minimum allowed by the X-ray unit (3 mm Al). The absence of a grid allowed to further reduce the DAP and effective dose by up to 80 %. Comparison between Russian and Swedish X-ray units showed the same trend in DAP and effective dose reduction, but the absolute dose values were lower by almost a factor of 10 for the Swedish units due to different image receptors and automatic exposure control settings. (Less)
Please use this url to cite or link to this publication:
author
; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Radiation Protection Dosimetry
volume
169
issue
1-4
pages
232 - 239
publisher
Oxford University Press
external identifiers
  • pmid:26769906
  • pmid:26769906
  • scopus:84979025769
  • wos:000383492100035
ISSN
1742-3406
DOI
10.1093/rpd/ncv535
project
Radiation protection in Eastern Europe
language
English
LU publication?
yes
id
ead7569d-8018-4012-951a-4f1f8cf850be (old id 8589054)
alternative location
http://www.ncbi.nlm.nih.gov/pubmed/26769906?dopt=Abstract
date added to LUP
2016-04-04 09:23:00
date last changed
2022-07-24 22:07:05
@article{ead7569d-8018-4012-951a-4f1f8cf850be,
  abstract     = {{An anthropomorphic phantom study was carried out in 2013-14 in two hospitals, one located in Russia (Mariinsky Hospital, Saint Petersburg) and the other in Sweden (Skåne University Hospital, Malmö). The aim of the study was to investigate the possibilities to reduce the patient dose from digital X-ray chest screening examinations. The existing chest imaging protocols were adjusted by changing the tube voltage, total filtration and grid in order to determine the most dose-effective combination of the examination parameters. It was possible to achieve up to 50 % dose-area product (DAP) and 30 % effective dose reduction by raising the tube voltage from 100 to 125 or 150 kV, and simultaneously decrease the total filtration to the minimum allowed by the X-ray unit (3 mm Al). The absence of a grid allowed to further reduce the DAP and effective dose by up to 80 %. Comparison between Russian and Swedish X-ray units showed the same trend in DAP and effective dose reduction, but the absolute dose values were lower by almost a factor of 10 for the Swedish units due to different image receptors and automatic exposure control settings.}},
  author       = {{Vodovatov, A V and Drozdov, A A and Telnova, A U and Bernhardsson, Christian}},
  issn         = {{1742-3406}},
  language     = {{eng}},
  number       = {{1-4}},
  pages        = {{232--239}},
  publisher    = {{Oxford University Press}},
  series       = {{Radiation Protection Dosimetry}},
  title        = {{MANAGEMENT OF PATIENT DOSES FROM DIGITAL X-RAY CHEST SCREENING EXAMINATIONS.}},
  url          = {{http://dx.doi.org/10.1093/rpd/ncv535}},
  doi          = {{10.1093/rpd/ncv535}},
  volume       = {{169}},
  year         = {{2016}},
}