On the reduction of imaging time-points for dosimetry in radionuclide therapy

Gustafsson, Johan LU orcid and Taprogge, Jan (2025) In EJNMMI Physics 12(1).
Abstract

Background: The aim was to develop a theoretical framework for how errors in estimated activities propagate to a dispersion in time-integrated activity in radionuclide-therapy dosimetry and how this affects the comparison of radionuclide-therapy dosimetry schemes. Methods: Formulae for the variance of relative errors of estimated time-integrated activities and relative differences in time-integrated activities between measurement schemes when one or more time-points are removed were derived using the law of propagation of uncertainty for a population of time-activity-curve parameters. The formulae were derived under the assumptions of fixed coefficients of variation for estimated activities, and underlying mono-exponential curves.... (More)

Background: The aim was to develop a theoretical framework for how errors in estimated activities propagate to a dispersion in time-integrated activity in radionuclide-therapy dosimetry and how this affects the comparison of radionuclide-therapy dosimetry schemes. Methods: Formulae for the variance of relative errors of estimated time-integrated activities and relative differences in time-integrated activities between measurement schemes when one or more time-points are removed were derived using the law of propagation of uncertainty for a population of time-activity-curve parameters. The formulae were derived under the assumptions of fixed coefficients of variation for estimated activities, and underlying mono-exponential curves. Analytical predictions were compared with results from numerical simulations and data for kidneys, liver, and spleen from a data-set of 18 patients treated with 177Lu-DOTA-TATE. Results: The dispersion in time-integrated activity is minimized if the time-points used for curve fitting have a large dispersion and are centered over the mean of τ=λeff-1 over the population, where λeff is the effective decay constant (i.e., the sum of the biological and physical decay constants). For large dispersions of decay constants in the population, the centering of time-points becomes gradually less important. The analytical expressions replicated the main trends from the numerical simulations. Furthermore, the analytical expressions predicted correctly the optimal reduced imaging schedule in 9 of 12 pairwise comparisons between schedules for patients. Conclusions: The dispersion of errors and deviations in estimated time-activity curves can be predicted using simple formulae. These formulae have the potential to be used for optimization of dosimetry measurement schemes for established and new radiopharmaceuticals as long as the mean and dispersion of biological half-lives are known in the patient population.

(Less)
Please use this url to cite or link to this publication:
author
and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
Curve fitting, Dosimetry, Radionuclide therapy, Uncertainty
in
EJNMMI Physics
volume
12
issue
1
article number
14
publisher
Springer
external identifiers
  • pmid:39909959
  • scopus:85218207962
ISSN
2197-7364
DOI
10.1186/s40658-025-00721-y
language
English
LU publication?
yes
id
0708d084-9869-44a3-8a1f-a3425f4faa81
date added to LUP
2025-06-09 09:23:47
date last changed
2025-06-10 03:00:02
@article{0708d084-9869-44a3-8a1f-a3425f4faa81,
  abstract     = {{<p>Background: The aim was to develop a theoretical framework for how errors in estimated activities propagate to a dispersion in time-integrated activity in radionuclide-therapy dosimetry and how this affects the comparison of radionuclide-therapy dosimetry schemes. Methods: Formulae for the variance of relative errors of estimated time-integrated activities and relative differences in time-integrated activities between measurement schemes when one or more time-points are removed were derived using the law of propagation of uncertainty for a population of time-activity-curve parameters. The formulae were derived under the assumptions of fixed coefficients of variation for estimated activities, and underlying mono-exponential curves. Analytical predictions were compared with results from numerical simulations and data for kidneys, liver, and spleen from a data-set of 18 patients treated with <sup>177</sup>Lu-DOTA-TATE. Results: The dispersion in time-integrated activity is minimized if the time-points used for curve fitting have a large dispersion and are centered over the mean of τ=λeff-1 over the population, where λeff is the effective decay constant (i.e., the sum of the biological and physical decay constants). For large dispersions of decay constants in the population, the centering of time-points becomes gradually less important. The analytical expressions replicated the main trends from the numerical simulations. Furthermore, the analytical expressions predicted correctly the optimal reduced imaging schedule in 9 of 12 pairwise comparisons between schedules for patients. Conclusions: The dispersion of errors and deviations in estimated time-activity curves can be predicted using simple formulae. These formulae have the potential to be used for optimization of dosimetry measurement schemes for established and new radiopharmaceuticals as long as the mean and dispersion of biological half-lives are known in the patient population.</p>}},
  author       = {{Gustafsson, Johan and Taprogge, Jan}},
  issn         = {{2197-7364}},
  keywords     = {{Curve fitting; Dosimetry; Radionuclide therapy; Uncertainty}},
  language     = {{eng}},
  number       = {{1}},
  publisher    = {{Springer}},
  series       = {{EJNMMI Physics}},
  title        = {{On the reduction of imaging time-points for dosimetry in radionuclide therapy}},
  url          = {{http://dx.doi.org/10.1186/s40658-025-00721-y}},
  doi          = {{10.1186/s40658-025-00721-y}},
  volume       = {{12}},
  year         = {{2025}},
}