Outcomes and performance of the ACURATE neo2 transcatheter heart valve in clinical practice : one-yearresults of the ACURATE neo2 PMCF Study

Kim, Won Keun ; Möllmann, Helge ; Montorfano, Matteo ; Ellert-Gregersen, Julia ; Rudolph, Tanja K. ; Van Mieghem, Nicolas M. ; Hilker, Michael ; Amat-Santos, Ignacio ; Terkelsen, Christian Juhl and Petronio, Anna Sonia , et al. (2024) In EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology 20(1). p.85-94
Abstract

BACKGROUND: Transcatheter aortic valve implantation is an effective treatment for patients with aortic stenosis; however, complications related to paravalvular leakage (PVL) persist, including increased risk of mortality, cardiovascular mortality, and rehospitalisation. AIMS: We sought to evaluate the clinical outcomes and valve performance at 1 year in patients with severe aortic stenosis treated with the ACURATE neo2 valve in a post-market clinical setting. METHODS: Valve Academic Research Consortium-2 safety events were assessed up to 1 year. Independent core laboratories evaluated echocardiographic measures of valve performance and hypoattenuated leaflet thickening (HALT; as measured by four-dimensional computed tomography).... (More)

BACKGROUND: Transcatheter aortic valve implantation is an effective treatment for patients with aortic stenosis; however, complications related to paravalvular leakage (PVL) persist, including increased risk of mortality, cardiovascular mortality, and rehospitalisation. AIMS: We sought to evaluate the clinical outcomes and valve performance at 1 year in patients with severe aortic stenosis treated with the ACURATE neo2 valve in a post-market clinical setting. METHODS: Valve Academic Research Consortium-2 safety events were assessed up to 1 year. Independent core laboratories evaluated echocardiographic measures of valve performance and hypoattenuated leaflet thickening (HALT; as measured by four-dimensional computed tomography). RESULTS: The study enrolled 250 patients (64% female; mean age: 81 years; baseline Society of Thoracic Surgeons risk score: 2.9±2.0%); 246 patients were implanted with ACURATE neo2. All-cause mortality was 0.8% at 30 days and 5.1% at 1 year. The 1-year rates for stroke and disabling stroke were 3.0% and 1.3%, respectively. Overall, HALT of >50% leaflet involvement of at least one leaflet was present in 9% of patients at 30 days and in 12% of patients at 1 year. No association was observed between the presence of HALT and 1-year clinical or haemodynamic outcomes. Early haemodynamic improvements were maintained up to 1 year (mean aortic valve gradient: 47.6±14.5 mmHg at baseline, 7.6±3.2 mmHg at 1 year; mean aortic valve area: 0.7±0.2 cm2 at baseline, 1.7±0.4 cm2 at 1 year). At 1 year, 99% of patients had mild or no/trace PVL (<1% had moderate PVL; no patient had severe PVL). CONCLUSIONS: The study outcomes confirm favourable performance and safety up to 1 year in patients treated with ACURATE neo2 in routine clinical practice. (ClinicalTrials.gov: NCT04655248).

(Less)
Please use this url to cite or link to this publication:
author
; ; ; ; ; ; ; ; and , et al. (More)
; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; ; and (Less)
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology
volume
20
issue
1
pages
10 pages
publisher
Société Europa Edition
external identifiers
  • pmid:37982152
  • scopus:85181539269
ISSN
1969-6213
DOI
10.4244/EIJ-D-23-00823
language
English
LU publication?
yes
id
0a13ded5-c075-48dd-8857-816e5753e913
date added to LUP
2024-02-06 13:58:07
date last changed
2024-04-22 21:02:10
@article{0a13ded5-c075-48dd-8857-816e5753e913,
  abstract     = {{<p>BACKGROUND: Transcatheter aortic valve implantation is an effective treatment for patients with aortic stenosis; however, complications related to paravalvular leakage (PVL) persist, including increased risk of mortality, cardiovascular mortality, and rehospitalisation. AIMS: We sought to evaluate the clinical outcomes and valve performance at 1 year in patients with severe aortic stenosis treated with the ACURATE neo2 valve in a post-market clinical setting. METHODS: Valve Academic Research Consortium-2 safety events were assessed up to 1 year. Independent core laboratories evaluated echocardiographic measures of valve performance and hypoattenuated leaflet thickening (HALT; as measured by four-dimensional computed tomography). RESULTS: The study enrolled 250 patients (64% female; mean age: 81 years; baseline Society of Thoracic Surgeons risk score: 2.9±2.0%); 246 patients were implanted with ACURATE neo2. All-cause mortality was 0.8% at 30 days and 5.1% at 1 year. The 1-year rates for stroke and disabling stroke were 3.0% and 1.3%, respectively. Overall, HALT of &gt;50% leaflet involvement of at least one leaflet was present in 9% of patients at 30 days and in 12% of patients at 1 year. No association was observed between the presence of HALT and 1-year clinical or haemodynamic outcomes. Early haemodynamic improvements were maintained up to 1 year (mean aortic valve gradient: 47.6±14.5 mmHg at baseline, 7.6±3.2 mmHg at 1 year; mean aortic valve area: 0.7±0.2 cm2 at baseline, 1.7±0.4 cm2 at 1 year). At 1 year, 99% of patients had mild or no/trace PVL (&lt;1% had moderate PVL; no patient had severe PVL). CONCLUSIONS: The study outcomes confirm favourable performance and safety up to 1 year in patients treated with ACURATE neo2 in routine clinical practice. (ClinicalTrials.gov: NCT04655248).</p>}},
  author       = {{Kim, Won Keun and Möllmann, Helge and Montorfano, Matteo and Ellert-Gregersen, Julia and Rudolph, Tanja K. and Van Mieghem, Nicolas M. and Hilker, Michael and Amat-Santos, Ignacio and Terkelsen, Christian Juhl and Petronio, Anna Sonia and Stella, Pieter and Götberg, Matthias and Rück, Andreas and Kasel, A. Markus and Trillo, Ramiro and Appleby, Clare and Barbanti, Marco and Blanke, Philipp and Asch, Federico M. and Modolo, Rodrigo and Allocco, Dominic J. and Tamburino, Corrado}},
  issn         = {{1969-6213}},
  language     = {{eng}},
  number       = {{1}},
  pages        = {{85--94}},
  publisher    = {{Société Europa Edition}},
  series       = {{EuroIntervention : journal of EuroPCR in collaboration with the Working Group on Interventional Cardiology of the European Society of Cardiology}},
  title        = {{Outcomes and performance of the ACURATE neo2 transcatheter heart valve in clinical practice : one-yearresults of the ACURATE neo2 PMCF Study}},
  url          = {{http://dx.doi.org/10.4244/EIJ-D-23-00823}},
  doi          = {{10.4244/EIJ-D-23-00823}},
  volume       = {{20}},
  year         = {{2024}},
}