A simple and fast liquid-liquid extraction method for the determination of 2,2 ',4,4 ',5,5 '-hexachlorobiphenyl (CB-153) and 1,1-dichloro-2,2-bis(p-chlorophenyl)-ethylene (p,p '-DDE) from human serum for epidemiological studies on type 2 diabetes

Rantakokko, Panu; Kiviranta, Hannu; Rylander, Lars LU ; Rignell-Hydbom, Anna LU and Vartiainen, Terttu (2009) In Journal of Chromatography A 1216(6). p.897-901
Abstract
A simple and fast method is presented to be used for example in studies on the relationship between serum levels of persistent organic pollutants and type 2 diabetes mellitus. Method is based on liquid-liquid extraction and gas chromatography coupled with high.-resolution mass spectrometry. In the sample pretreatment special attention was paid to minimize the number of sample manipulation steps and the amounts of organic solvents needed. Compounds analyzed were 2,2',4,4',5,5'-hexachlorobiphenyl (PCB-153) and 1,1-dichloro-2,2-bis (p-chlorophenyl)-ethylene (p,p'-DDE), the major metabolite of DDT. The method included extraction and cleanup of 0.2 rill of serum in a single test tube and subsequent analysis of the extract from 0.2 ml final... (More)
A simple and fast method is presented to be used for example in studies on the relationship between serum levels of persistent organic pollutants and type 2 diabetes mellitus. Method is based on liquid-liquid extraction and gas chromatography coupled with high.-resolution mass spectrometry. In the sample pretreatment special attention was paid to minimize the number of sample manipulation steps and the amounts of organic solvents needed. Compounds analyzed were 2,2',4,4',5,5'-hexachlorobiphenyl (PCB-153) and 1,1-dichloro-2,2-bis (p-chlorophenyl)-ethylene (p,p'-DDE), the major metabolite of DDT. The method included extraction and cleanup of 0.2 rill of serum in a single test tube and subsequent analysis of the extract from 0.2 ml final volume. Validation was conducted to explore the performance of the method. The limits of detection for p,p'-DDE and PCB-153 based on the standard deviation of the blank samples were 4.3 and 3.1 pg/ml, respectively. Repeatability was less than 2.5% at three concentration levels tested and recovery from Certified Reference Material SRM 1589a was 84% for p,p'-DDE and 87% for PCB-153 of the certified values, respectively. Serum samples from the AMAP intercalibration round 2008-2 were also analyzed, and results were 101-116% of the assigned values. The presented method was used for an epidemiological study with more than 700 serum samples from a type 2 diabetes cohort from Sweden. (C) 2008 Elsevier B.V. All rights reserved. (Less)
Please use this url to cite or link to this publication:
author
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
PCB-153, p, p '-DDE, organic pollutants (POPs), Persistent, Type 2 diabetes, Liquid-liquid extraction, Human serum
in
Journal of Chromatography A
volume
1216
issue
6
pages
897 - 901
publisher
Elsevier
external identifiers
  • wos:000263087700003
  • scopus:58149512054
ISSN
0021-9673
DOI
10.1016/j.chroma.2008.12.013
language
English
LU publication?
yes
id
5abc76a7-e371-4202-945e-9393908ba806 (old id 1311432)
date added to LUP
2009-03-18 08:51:18
date last changed
2017-01-01 06:14:25
@article{5abc76a7-e371-4202-945e-9393908ba806,
  abstract     = {A simple and fast method is presented to be used for example in studies on the relationship between serum levels of persistent organic pollutants and type 2 diabetes mellitus. Method is based on liquid-liquid extraction and gas chromatography coupled with high.-resolution mass spectrometry. In the sample pretreatment special attention was paid to minimize the number of sample manipulation steps and the amounts of organic solvents needed. Compounds analyzed were 2,2',4,4',5,5'-hexachlorobiphenyl (PCB-153) and 1,1-dichloro-2,2-bis (p-chlorophenyl)-ethylene (p,p'-DDE), the major metabolite of DDT. The method included extraction and cleanup of 0.2 rill of serum in a single test tube and subsequent analysis of the extract from 0.2 ml final volume. Validation was conducted to explore the performance of the method. The limits of detection for p,p'-DDE and PCB-153 based on the standard deviation of the blank samples were 4.3 and 3.1 pg/ml, respectively. Repeatability was less than 2.5% at three concentration levels tested and recovery from Certified Reference Material SRM 1589a was 84% for p,p'-DDE and 87% for PCB-153 of the certified values, respectively. Serum samples from the AMAP intercalibration round 2008-2 were also analyzed, and results were 101-116% of the assigned values. The presented method was used for an epidemiological study with more than 700 serum samples from a type 2 diabetes cohort from Sweden. (C) 2008 Elsevier B.V. All rights reserved.},
  author       = {Rantakokko, Panu and Kiviranta, Hannu and Rylander, Lars and Rignell-Hydbom, Anna and Vartiainen, Terttu},
  issn         = {0021-9673},
  keyword      = {PCB-153,p,p '-DDE,organic pollutants (POPs),Persistent,Type 2 diabetes,Liquid-liquid extraction,Human serum},
  language     = {eng},
  number       = {6},
  pages        = {897--901},
  publisher    = {Elsevier},
  series       = {Journal of Chromatography A},
  title        = {A simple and fast liquid-liquid extraction method for the determination of 2,2 ',4,4 ',5,5 '-hexachlorobiphenyl (CB-153) and 1,1-dichloro-2,2-bis(p-chlorophenyl)-ethylene (p,p '-DDE) from human serum for epidemiological studies on type 2 diabetes},
  url          = {http://dx.doi.org/10.1016/j.chroma.2008.12.013},
  volume       = {1216},
  year         = {2009},
}