A sensitive and selective amperometric biosensor, based on NADH-oxidase (NADOx) working in tandem with horseradish peroxidase (HRP), which is electrically connected to graphite electrode via Os-redox polymer (RP), was developed. Two designs were considered for the biosensor optimization: (i) the first one, [G/HRP-NADOx], exploited the direct electron transfer between HRP and the graphite electrode (G); (ii) the second one, [G/RP-HR-P-NADOx], used R-P as mediator. In both cases the enzyme matrix was cross-linked using poly(ethylenglycol) diglycidyl ether. The biosensors sensitivities, estimated as the slopes of linear ranges from the calibration curves, recorded in a single line flow injection setup, showed that G/RP-HRP-NADOx electrode... (More)
A sensitive and selective amperometric biosensor, based on NADH-oxidase (NADOx) working in tandem with horseradish peroxidase (HRP), which is electrically connected to graphite electrode via Os-redox polymer (RP), was developed. Two designs were considered for the biosensor optimization: (i) the first one, [G/HRP-NADOx], exploited the direct electron transfer between HRP and the graphite electrode (G); (ii) the second one, [G/RP-HR-P-NADOx], used R-P as mediator. In both cases the enzyme matrix was cross-linked using poly(ethylenglycol) diglycidyl ether. The biosensors sensitivities, estimated as the slopes of linear ranges from the calibration curves, recorded in a single line flow injection setup, showed that G/RP-HRP-NADOx electrode (1.62 +/- 0.05 mA M-1) presents a higher efficiency than G/HRP-NADOx electrode (0.30 +/- 0.03 mA M-1). (Less)
Please use this url to cite or link to this publication:
author
; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
NADH detection, amperometric biosensors, NADOx, Os-redox polymer
in
Revue Roumaine de Chimie
volume
54
issue
9
pages
755 - 760
publisher
Romanian Academy / Academia Române
external identifiers
  • wos:000273850000007
  • scopus:75949091709
ISSN
0035-3930
language
English
LU publication?
yes
additional info
The information about affiliations in this record was updated in December 2015. The record was previously connected to the following departments: Biotechnology (LTH) (011001037), Analytical Chemistry (S/LTH) (011001004)
id
a459b5cb-a42f-4d86-b869-5f59c575e1a7 (old id 1546486)
date added to LUP
2016-04-01 14:01:01
date last changed
2022-01-27 22:22:34
@article{a459b5cb-a42f-4d86-b869-5f59c575e1a7,
  abstract     = {{A sensitive and selective amperometric biosensor, based on NADH-oxidase (NADOx) working in tandem with horseradish peroxidase (HRP), which is electrically connected to graphite electrode via Os-redox polymer (RP), was developed. Two designs were considered for the biosensor optimization: (i) the first one, [G/HRP-NADOx], exploited the direct electron transfer between HRP and the graphite electrode (G); (ii) the second one, [G/RP-HR-P-NADOx], used R-P as mediator. In both cases the enzyme matrix was cross-linked using poly(ethylenglycol) diglycidyl ether. The biosensors sensitivities, estimated as the slopes of linear ranges from the calibration curves, recorded in a single line flow injection setup, showed that G/RP-HRP-NADOx electrode (1.62 +/- 0.05 mA M-1) presents a higher efficiency than G/HRP-NADOx electrode (0.30 +/- 0.03 mA M-1).}},
  author       = {{Muresan, Laura and Nistor, Mihaela and Csöregi, Elisabeth and Popescu, Catalin}},
  issn         = {{0035-3930}},
  keywords     = {{NADH detection; amperometric biosensors; NADOx; Os-redox polymer}},
  language     = {{eng}},
  number       = {{9}},
  pages        = {{755--760}},
  publisher    = {{Romanian Academy / Academia Române}},
  series       = {{Revue Roumaine de Chimie}},
  title        = {{Reagentless Amperometric Biosensor For Nadh Detection}},
  volume       = {{54}},
  year         = {{2009}},
}