Iodine contrast iso-attenuating with diagnostic gadolinium doses in CTA and angiography results in ultra-low iodine doses. A way to avoid both CIN and NSF in azotemic patients?
(2011) In European Radiology 21. p.326-336- Abstract
- OBJECTIVES: To establish iodine (I) contrast medium (CM) doses iso-attenuating with gadolinium (Gd) CM doses regarded diagnostic in CTA and percutaneous catheter-angiography/vascular interventions (PCA/PVI) in azotemic patients. METHODS: CT Hounsfield units (HU) were measured in 20-mL syringes containing 0.01/0.02,/0.05/0.1 mmol/mL of iodine or gadolinium atoms and placed in phantoms. Relative contrast were measured in 20-mL syringes filled with iohexol at 35/50/70/90/110/140 mg I/mL and 0.5 M gadodiamide using radiofluoroscopy (RF), digital radiography (DX) and x-ray angiography (XA) systems. Clinical doses of Gd-CM at CTA/PCA/PVI were reviewed. RESULTS: At CT 91-116 and 104-125 mg I/mL in the chest and abdominal phantoms, respectively,... (More)
- OBJECTIVES: To establish iodine (I) contrast medium (CM) doses iso-attenuating with gadolinium (Gd) CM doses regarded diagnostic in CTA and percutaneous catheter-angiography/vascular interventions (PCA/PVI) in azotemic patients. METHODS: CT Hounsfield units (HU) were measured in 20-mL syringes containing 0.01/0.02,/0.05/0.1 mmol/mL of iodine or gadolinium atoms and placed in phantoms. Relative contrast were measured in 20-mL syringes filled with iohexol at 35/50/70/90/110/140 mg I/mL and 0.5 M gadodiamide using radiofluoroscopy (RF), digital radiography (DX) and x-ray angiography (XA) systems. Clinical doses of Gd-CM at CTA/PCA/PVI were reviewed. RESULTS: At CT 91-116 and 104-125 mg I/mL in the chest and abdominal phantoms, respectively, were iso-attenuating with 0.5 M Gd at 80-140 kVp. At RF/DX/XA systems 35-90 mg I/mL were iso-attenuating with 0.5 M gadodiamide at 60-115 kVp. Clinically, 60 mL 91-125 mg I/mL (5.5-7.5 gram-iodine) at 80-140 kVp CTA and 60 mL of 35-90 mg I/mL (2.1-5.4 gram-iodine) at 60-115 kVp PCA/PVI would be iso-attenuating with 60 mL 0.5 M Gd-CM (=0.4 mmol Gd/kg in a 75-kg person). CONCLUSIONS: Meticulous examination technique and judicious use of ultra-low I-CM doses iso-attenuating with diagnostic Gd-CM doses in CTA and PCA/PVI may minimise the risk of nephrotoxicity in azotemic patients, while there is no risk of NSF. (Less)
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/1664906
- author
- Nyman, Ulf ; Elmståhl, Barbara LU ; Geijer, Håkan ; Leander, Peter LU ; Almén, Torsten LU and Nilsson, Mats LU
- organization
- publishing date
- 2011
- type
- Contribution to journal
- publication status
- published
- subject
- in
- European Radiology
- volume
- 21
- pages
- 326 - 336
- publisher
- Springer
- external identifiers
-
- wos:000287079300014
- pmid:20803203
- scopus:79951955235
- pmid:20803203
- ISSN
- 0938-7994
- DOI
- 10.1007/s00330-010-1924-2
- language
- English
- LU publication?
- yes
- id
- 785fe11b-09b7-4800-a976-dcc32271fa65 (old id 1664906)
- alternative location
- http://www.ncbi.nlm.nih.gov/pubmed/20803203?dopt=Abstract
- date added to LUP
- 2016-04-04 07:10:49
- date last changed
- 2022-05-01 03:30:01
@article{785fe11b-09b7-4800-a976-dcc32271fa65, abstract = {{OBJECTIVES: To establish iodine (I) contrast medium (CM) doses iso-attenuating with gadolinium (Gd) CM doses regarded diagnostic in CTA and percutaneous catheter-angiography/vascular interventions (PCA/PVI) in azotemic patients. METHODS: CT Hounsfield units (HU) were measured in 20-mL syringes containing 0.01/0.02,/0.05/0.1 mmol/mL of iodine or gadolinium atoms and placed in phantoms. Relative contrast were measured in 20-mL syringes filled with iohexol at 35/50/70/90/110/140 mg I/mL and 0.5 M gadodiamide using radiofluoroscopy (RF), digital radiography (DX) and x-ray angiography (XA) systems. Clinical doses of Gd-CM at CTA/PCA/PVI were reviewed. RESULTS: At CT 91-116 and 104-125 mg I/mL in the chest and abdominal phantoms, respectively, were iso-attenuating with 0.5 M Gd at 80-140 kVp. At RF/DX/XA systems 35-90 mg I/mL were iso-attenuating with 0.5 M gadodiamide at 60-115 kVp. Clinically, 60 mL 91-125 mg I/mL (5.5-7.5 gram-iodine) at 80-140 kVp CTA and 60 mL of 35-90 mg I/mL (2.1-5.4 gram-iodine) at 60-115 kVp PCA/PVI would be iso-attenuating with 60 mL 0.5 M Gd-CM (=0.4 mmol Gd/kg in a 75-kg person). CONCLUSIONS: Meticulous examination technique and judicious use of ultra-low I-CM doses iso-attenuating with diagnostic Gd-CM doses in CTA and PCA/PVI may minimise the risk of nephrotoxicity in azotemic patients, while there is no risk of NSF.}}, author = {{Nyman, Ulf and Elmståhl, Barbara and Geijer, Håkan and Leander, Peter and Almén, Torsten and Nilsson, Mats}}, issn = {{0938-7994}}, language = {{eng}}, pages = {{326--336}}, publisher = {{Springer}}, series = {{European Radiology}}, title = {{Iodine contrast iso-attenuating with diagnostic gadolinium doses in CTA and angiography results in ultra-low iodine doses. A way to avoid both CIN and NSF in azotemic patients?}}, url = {{http://dx.doi.org/10.1007/s00330-010-1924-2}}, doi = {{10.1007/s00330-010-1924-2}}, volume = {{21}}, year = {{2011}}, }