A MRI and polarized gases compatible respirator and gas administrator for the study of the small animal lung : volume measurement and control
(2010) In IEEE Transactions on Biomedical Engineering 57(7). p.9-1745- Abstract
We have developed over the past years an experimental magnetic resonance imaging (MRI) and polarized gases compatible mechanical respirator for the study of the small experimental animal. The respirator has been successfully used for experiments both in the MRI setting for polarized (3)He, (19)F, and proton imaging as well as for functional measurements of the lungs. The new main pneumatic valve with the two integrated sensors for simultaneous lung pressure and volume measurements and the proportional valve to set the tidal volume of the respiration are described. It is shown how the device measures and controls the tidal volume of the lungs.
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/1a926b23-bc24-4ad6-a066-8576f850d99b
- author
- Kyriazis, Angelos ; Perez de Alejo, Rigoberto ; Rodriguez, Ignacio ; Olsson, Lars E LU and Ruiz-Cabello, Jesus
- organization
- publishing date
- 2010-07
- type
- Contribution to journal
- publication status
- published
- subject
- keywords
- Animals, Biomedical Research, Gases, Lung Compliance, Lung Volume Measurements, Magnetic Resonance Imaging, Mice, Ventilators, Mechanical
- in
- IEEE Transactions on Biomedical Engineering
- volume
- 57
- issue
- 7
- pages
- 5 pages
- publisher
- IEEE - Institute of Electrical and Electronics Engineers Inc.
- external identifiers
-
- scopus:77953800531
- pmid:20176535
- ISSN
- 1558-2531
- DOI
- 10.1109/TBME.2010.2042596
- language
- English
- LU publication?
- yes
- id
- 1a926b23-bc24-4ad6-a066-8576f850d99b
- date added to LUP
- 2016-08-16 12:09:40
- date last changed
- 2024-06-14 12:05:45
@article{1a926b23-bc24-4ad6-a066-8576f850d99b, abstract = {{<p>We have developed over the past years an experimental magnetic resonance imaging (MRI) and polarized gases compatible mechanical respirator for the study of the small experimental animal. The respirator has been successfully used for experiments both in the MRI setting for polarized (3)He, (19)F, and proton imaging as well as for functional measurements of the lungs. The new main pneumatic valve with the two integrated sensors for simultaneous lung pressure and volume measurements and the proportional valve to set the tidal volume of the respiration are described. It is shown how the device measures and controls the tidal volume of the lungs.</p>}}, author = {{Kyriazis, Angelos and Perez de Alejo, Rigoberto and Rodriguez, Ignacio and Olsson, Lars E and Ruiz-Cabello, Jesus}}, issn = {{1558-2531}}, keywords = {{Animals; Biomedical Research; Gases; Lung Compliance; Lung Volume Measurements; Magnetic Resonance Imaging; Mice; Ventilators, Mechanical}}, language = {{eng}}, number = {{7}}, pages = {{9--1745}}, publisher = {{IEEE - Institute of Electrical and Electronics Engineers Inc.}}, series = {{IEEE Transactions on Biomedical Engineering}}, title = {{A MRI and polarized gases compatible respirator and gas administrator for the study of the small animal lung : volume measurement and control}}, url = {{http://dx.doi.org/10.1109/TBME.2010.2042596}}, doi = {{10.1109/TBME.2010.2042596}}, volume = {{57}}, year = {{2010}}, }