Leaflet arrest in St Jude Medical and CarboMedics valves: an experimental study

Grattan, MT and Thulin, Lars LU (2004) In European Journal of Cardio-Thoracic Surgery 25(6). p.953-957
Abstract
Objective: Two patients who suffered acute failure of their St Jude Medical Masters aortic valve prostheses due to leaflet arrest that were unrelated to suture material are presented. It was hypothesized that the valves failed because force applied to bear upon the valve annulus caused the hinge mechanism to become restricted or to arrest. Methods: A study was designed to measure the force that would cause leaflet arrest in three sizes of St Jude Medical Masters valves and CarboMedics mechanical valves. A specially manufactured pushrod device was used to apply a variable force to the sewing cuff, low annulus level, or to the pivot guards of all the valves. Results: For every valve size tested, the St Jude Medical Masters valves required... (More)
Objective: Two patients who suffered acute failure of their St Jude Medical Masters aortic valve prostheses due to leaflet arrest that were unrelated to suture material are presented. It was hypothesized that the valves failed because force applied to bear upon the valve annulus caused the hinge mechanism to become restricted or to arrest. Methods: A study was designed to measure the force that would cause leaflet arrest in three sizes of St Jude Medical Masters valves and CarboMedics mechanical valves. A specially manufactured pushrod device was used to apply a variable force to the sewing cuff, low annulus level, or to the pivot guards of all the valves. Results: For every valve size tested, the St Jude Medical Masters valves required significantly less force applied than did the CarboMedics valves to cause one or both leaflets to arrest (P < 0.0004). Conclusions: A force applied on the valve ring of mechanical valves can cause leaflet arrest. The force required to arrest the leaflets is within an order of magnitude of that measured in other in vivo animal studies. We conclude that force on the valve rings in patients after aortic valve surgery could cause leaflet malfunction and even arrest in some patients. (C) 2004 Elsevier B.V. All rights reserved. (Less)
Please use this url to cite or link to this publication:
author
and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
valve malfunction, mechanical valve, valve replacement
in
European Journal of Cardio-Thoracic Surgery
volume
25
issue
6
pages
953 - 957
publisher
Oxford University Press
external identifiers
  • wos:000221996700011
  • pmid:15144994
  • scopus:2442445459
  • pmid:15144994
ISSN
1010-7940
DOI
10.1016/j.ejcts.2004.02.039
language
English
LU publication?
yes
id
3eb9c08f-35b1-4e68-abec-2cd80c312ca2 (old id 274903)
alternative location
http://ejcts.ctsnetjournals.org/cgi/content/abstract/25/6/953
date added to LUP
2016-04-01 12:15:15
date last changed
2022-01-27 01:06:23
@article{3eb9c08f-35b1-4e68-abec-2cd80c312ca2,
  abstract     = {{Objective: Two patients who suffered acute failure of their St Jude Medical Masters aortic valve prostheses due to leaflet arrest that were unrelated to suture material are presented. It was hypothesized that the valves failed because force applied to bear upon the valve annulus caused the hinge mechanism to become restricted or to arrest. Methods: A study was designed to measure the force that would cause leaflet arrest in three sizes of St Jude Medical Masters valves and CarboMedics mechanical valves. A specially manufactured pushrod device was used to apply a variable force to the sewing cuff, low annulus level, or to the pivot guards of all the valves. Results: For every valve size tested, the St Jude Medical Masters valves required significantly less force applied than did the CarboMedics valves to cause one or both leaflets to arrest (P &lt; 0.0004). Conclusions: A force applied on the valve ring of mechanical valves can cause leaflet arrest. The force required to arrest the leaflets is within an order of magnitude of that measured in other in vivo animal studies. We conclude that force on the valve rings in patients after aortic valve surgery could cause leaflet malfunction and even arrest in some patients. (C) 2004 Elsevier B.V. All rights reserved.}},
  author       = {{Grattan, MT and Thulin, Lars}},
  issn         = {{1010-7940}},
  keywords     = {{valve malfunction; mechanical valve; valve replacement}},
  language     = {{eng}},
  number       = {{6}},
  pages        = {{953--957}},
  publisher    = {{Oxford University Press}},
  series       = {{European Journal of Cardio-Thoracic Surgery}},
  title        = {{Leaflet arrest in St Jude Medical and CarboMedics valves: an experimental study}},
  url          = {{http://dx.doi.org/10.1016/j.ejcts.2004.02.039}},
  doi          = {{10.1016/j.ejcts.2004.02.039}},
  volume       = {{25}},
  year         = {{2004}},
}