Correlation Between Range of Motion and Implant Fracture: A 5 Year Follow-Up of 72 Joints in 18 Patients in a Randomized Study Comparing Swanson and Avanta/Sutter Mcp Silicone Prosthesis
(2009) In Journal of Hand Surgery: European Volume 34E(6). p.743-747- Abstract
- Eighteen out of 18 rheumatoid patients (at one centre of a two-centre 30 patient study previously reported) with a mean age of 56 years, and 72/72 operated joints were randomized to Avanta/Sutter or Swanson MCP prostheses and followed for 5 years. Both ulnar deviation and extension lag were improved already at 6 weeks and remained improved at 5 years. The Avanta prosthesis had a better range of motion (ROM) than the Swanson. Six of nine patients with Avanta/Sutter implants had at least one implant fracture compared to 1/9 patients with the Swanson implant (P=0.05) but fracture did not change the outcome subjectively. The ROM at 3 months correlated with the occurrence of an implant fracture at 5 years and a greater early ROM may be related... (More)
- Eighteen out of 18 rheumatoid patients (at one centre of a two-centre 30 patient study previously reported) with a mean age of 56 years, and 72/72 operated joints were randomized to Avanta/Sutter or Swanson MCP prostheses and followed for 5 years. Both ulnar deviation and extension lag were improved already at 6 weeks and remained improved at 5 years. The Avanta prosthesis had a better range of motion (ROM) than the Swanson. Six of nine patients with Avanta/Sutter implants had at least one implant fracture compared to 1/9 patients with the Swanson implant (P=0.05) but fracture did not change the outcome subjectively. The ROM at 3 months correlated with the occurrence of an implant fracture at 5 years and a greater early ROM may be related to implant fracture. At 5 years patients remained satisfied and the deformities remained corrected. (Less)
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/1546629
- author
- Tägil, Magnus LU ; Geijer, Mats LU ; Malcus, P. and Kopylov, Philippe LU
- organization
- publishing date
- 2009
- type
- Contribution to journal
- publication status
- published
- subject
- keywords
- motion, range of, prosthetic fracture, MCP arthroplasty, rheumatoid arthritis, ulnar deviation, extension lag
- in
- Journal of Hand Surgery: European Volume
- volume
- 34E
- issue
- 6
- pages
- 743 - 747
- publisher
- SAGE Publications
- external identifiers
-
- wos:000273617800003
- scopus:74549206347
- ISSN
- 2043-6289
- DOI
- 10.1177/1753193409346793
- language
- English
- LU publication?
- yes
- id
- 2c5c4203-2641-4196-97c6-21eae55db076 (old id 1546629)
- date added to LUP
- 2016-04-01 12:12:07
- date last changed
- 2025-10-14 09:31:36
@article{2c5c4203-2641-4196-97c6-21eae55db076,
abstract = {{Eighteen out of 18 rheumatoid patients (at one centre of a two-centre 30 patient study previously reported) with a mean age of 56 years, and 72/72 operated joints were randomized to Avanta/Sutter or Swanson MCP prostheses and followed for 5 years. Both ulnar deviation and extension lag were improved already at 6 weeks and remained improved at 5 years. The Avanta prosthesis had a better range of motion (ROM) than the Swanson. Six of nine patients with Avanta/Sutter implants had at least one implant fracture compared to 1/9 patients with the Swanson implant (P=0.05) but fracture did not change the outcome subjectively. The ROM at 3 months correlated with the occurrence of an implant fracture at 5 years and a greater early ROM may be related to implant fracture. At 5 years patients remained satisfied and the deformities remained corrected.}},
author = {{Tägil, Magnus and Geijer, Mats and Malcus, P. and Kopylov, Philippe}},
issn = {{2043-6289}},
keywords = {{motion; range of; prosthetic fracture; MCP arthroplasty; rheumatoid arthritis; ulnar deviation; extension lag}},
language = {{eng}},
number = {{6}},
pages = {{743--747}},
publisher = {{SAGE Publications}},
series = {{Journal of Hand Surgery: European Volume}},
title = {{Correlation Between Range of Motion and Implant Fracture: A 5 Year Follow-Up of 72 Joints in 18 Patients in a Randomized Study Comparing Swanson and Avanta/Sutter Mcp Silicone Prosthesis}},
url = {{http://dx.doi.org/10.1177/1753193409346793}},
doi = {{10.1177/1753193409346793}},
volume = {{34E}},
year = {{2009}},
}