Skip to main content

Lund University Publications

LUND UNIVERSITY LIBRARIES

4D dosimetry and its applications to pre-treatment quality control and real-time in vivo dosimetry of VMAT treatments

Nordström, Fredrik LU ; af Wetterstedt, S. and Bäck, Sven LU (2013) 7th International Conference on 3D Radiation Dosimetry (IC3DDose) 444. p.012021-012021
Abstract
In this study, a 4D dosimetry concept was developed. This concept included a method for calculation of 3D reference absorbed dose matrices at every control point of the delivery using a clinical treatment planning system (TPS). Further, the gamma evaluation method was extended to incorporate the 4th dimension of the TPS calculated dose distributions. The applications of the 4D dosimetry concept on pre-treatment quality control and real-time in vivo dosimetry were investigated.
Please use this url to cite or link to this publication:
author
; and
organization
publishing date
type
Chapter in Book/Report/Conference proceeding
publication status
published
subject
host publication
7th International Conference on 3D Radiation Dosimetry (IC3DDose)
volume
444
pages
012021 - 012021
publisher
IOP Publishing
conference name
7th International Conference on 3D Radiation Dosimetry (IC3DDose)
conference location
Sydney, Australia
conference dates
2012-11-04 - 2012-11-08
external identifiers
  • wos:000322968600021
  • scopus:84883407254
ISSN
1742-6596
1742-6588
DOI
10.1088/1742-6596/444/1/012021
language
English
LU publication?
yes
id
2d45d092-510e-400a-ac2a-8ca231a978ac (old id 4027082)
date added to LUP
2016-04-01 10:25:23
date last changed
2024-01-06 16:11:51
@inproceedings{2d45d092-510e-400a-ac2a-8ca231a978ac,
  abstract     = {{In this study, a 4D dosimetry concept was developed. This concept included a method for calculation of 3D reference absorbed dose matrices at every control point of the delivery using a clinical treatment planning system (TPS). Further, the gamma evaluation method was extended to incorporate the 4th dimension of the TPS calculated dose distributions. The applications of the 4D dosimetry concept on pre-treatment quality control and real-time in vivo dosimetry were investigated.}},
  author       = {{Nordström, Fredrik and af Wetterstedt, S. and Bäck, Sven}},
  booktitle    = {{7th International Conference on 3D Radiation Dosimetry (IC3DDose)}},
  issn         = {{1742-6596}},
  language     = {{eng}},
  pages        = {{012021--012021}},
  publisher    = {{IOP Publishing}},
  title        = {{4D dosimetry and its applications to pre-treatment quality control and real-time in vivo dosimetry of VMAT treatments}},
  url          = {{http://dx.doi.org/10.1088/1742-6596/444/1/012021}},
  doi          = {{10.1088/1742-6596/444/1/012021}},
  volume       = {{444}},
  year         = {{2013}},
}