Look-Locker FAIR TrueFISP for arterial spin labelling on mouse at 9.4 T

Gottschalk, Michael LU orcid (2020) In NMR in Biomedicine 33(3). p.1-11
Abstract

Pulsed arterial spin labelling remains a non-invasive and highly used method for the study of rodent cerebral blood flow (CBF). Flow-sensitive alternating inversion recovery (FAIR) is one of the most commonly used MR-sequences for this purpose and exists with many different strategies to record the images. This study investigates Look-Locker (LL) TrueFISP readout for FAIR as an alternative to the standard EPI readout, which is provided by the manufacturer. The aim was to show the improved image quality using TrueFISP and to verify the reproducibility of the determination of the cerebral blood flow values. The measurement of many inversion points also allowed to investigate the influence of the correct blood relaxation rate on the fit of... (More)

Pulsed arterial spin labelling remains a non-invasive and highly used method for the study of rodent cerebral blood flow (CBF). Flow-sensitive alternating inversion recovery (FAIR) is one of the most commonly used MR-sequences for this purpose and exists with many different strategies to record the images. This study investigates Look-Locker (LL) TrueFISP readout for FAIR as an alternative to the standard EPI readout, which is provided by the manufacturer. The aim was to show the improved image quality using TrueFISP and to verify the reproducibility of the determination of the cerebral blood flow values. The measurement of many inversion points also allowed to investigate the influence of the correct blood relaxation rate on the fit of the CBF data. For the LL-FAIR TrueFISP an in-house written method was created. The method was tested on a group of C57BL/6 mice at the field strength of 9.4 T. The results show CBF maps with less distortion than for EPI and the values found are in good agreement with the literature. A comparison of the CBF values found with EPI and LL-TrueFISP shows very small differences, most being not significant. In conclusion, the method presented gives equivalent CBF maps in comparison to standard FAIR-EPI. Both methods have the same measurement time. TrueFISP has the advantage to EPI of producing undistorted images over larger areas of the mouse brain. It is advisable to check the value of the blood relaxation rate by measurement or to estimate it as a fitting parameter.

(Less)
Please use this url to cite or link to this publication:
author
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
NMR in Biomedicine
volume
33
issue
3
article number
e4191
pages
1 - 11
publisher
John Wiley & Sons Inc.
external identifiers
  • pmid:31829485
  • scopus:85076352804
ISSN
0952-3480
DOI
10.1002/nbm.4191
language
English
LU publication?
yes
additional info
© 2019 John Wiley & Sons, Ltd.
id
2e24037c-39ac-4041-a372-e9541368f7e3
date added to LUP
2020-02-06 12:34:01
date last changed
2024-03-04 13:04:43
@article{2e24037c-39ac-4041-a372-e9541368f7e3,
  abstract     = {{<p>Pulsed arterial spin labelling remains a non-invasive and highly used method for the study of rodent cerebral blood flow (CBF). Flow-sensitive alternating inversion recovery (FAIR) is one of the most commonly used MR-sequences for this purpose and exists with many different strategies to record the images. This study investigates Look-Locker (LL) TrueFISP readout for FAIR as an alternative to the standard EPI readout, which is provided by the manufacturer. The aim was to show the improved image quality using TrueFISP and to verify the reproducibility of the determination of the cerebral blood flow values. The measurement of many inversion points also allowed to investigate the influence of the correct blood relaxation rate on the fit of the CBF data. For the LL-FAIR TrueFISP an in-house written method was created. The method was tested on a group of C57BL/6 mice at the field strength of 9.4 T. The results show CBF maps with less distortion than for EPI and the values found are in good agreement with the literature. A comparison of the CBF values found with EPI and LL-TrueFISP shows very small differences, most being not significant. In conclusion, the method presented gives equivalent CBF maps in comparison to standard FAIR-EPI. Both methods have the same measurement time. TrueFISP has the advantage to EPI of producing undistorted images over larger areas of the mouse brain. It is advisable to check the value of the blood relaxation rate by measurement or to estimate it as a fitting parameter.</p>}},
  author       = {{Gottschalk, Michael}},
  issn         = {{0952-3480}},
  language     = {{eng}},
  number       = {{3}},
  pages        = {{1--11}},
  publisher    = {{John Wiley & Sons Inc.}},
  series       = {{NMR in Biomedicine}},
  title        = {{Look-Locker FAIR TrueFISP for arterial spin labelling on mouse at 9.4 T}},
  url          = {{http://dx.doi.org/10.1002/nbm.4191}},
  doi          = {{10.1002/nbm.4191}},
  volume       = {{33}},
  year         = {{2020}},
}