A new collimator simulation in SIMIND based on the Delta-Scattering technique

Ljungberg, Michael LU ; Larsson, A and Johansson, L (2004) 2004 Nuclear Science Symposium, Medical Imaging Conference, Symposium on Nuclear Power Systems and the 14th International Workshop on Room Temperature Semiconductor X- and Gamma- Ray Detectors 6. p.3584-3588
Abstract
To use conventional ray tracing methods in Monte Carlo simulation of the collimator in a scintillation camera system can be time consuming. It is however necessary to take collimator interactions into account when simulating radionuclides emitting high-energy photons that can penetrate the septa in the collimator. In this work a statistical collimator algorithm, based on the Delta-Scattering method, is evaluated using <sup>123</sup>L The evaluation is performed by comparing results from Monte Carlo simulations and measurements for a scintillation camera system, using point sources and an RSD phantom. A good agreement can be seen for both images and energy spectra.
Please use this url to cite or link to this publication:
author
; and
organization
publishing date
type
Chapter in Book/Report/Conference proceeding
publication status
published
subject
keywords
Collimator algorithm, Scintialltion camera system, Collimator simulation, Delta-Scattering method
host publication
IEEE Nuclear Science Symposium Conference Record
volume
6
pages
3584 - 3588
publisher
IEEE - Institute of Electrical and Electronics Engineers Inc.
conference name
2004 Nuclear Science Symposium, Medical Imaging Conference, Symposium on Nuclear Power Systems and the 14th International Workshop on Room Temperature Semiconductor X- and Gamma- Ray Detectors
conference location
Rome, Italy
conference dates
2004-10-16 - 2004-10-22
external identifiers
  • wos:000232002105053
  • scopus:23844527205
ISSN
1095-7863
DOI
10.1109/NSSMIC.2004.1466659
language
English
LU publication?
yes
id
82cdb020-fa7d-4f9c-b197-c6ac2df151a3 (old id 613781)
date added to LUP
2016-04-01 15:36:16
date last changed
2022-03-14 19:05:22
@inproceedings{82cdb020-fa7d-4f9c-b197-c6ac2df151a3,
  abstract     = {{To use conventional ray tracing methods in Monte Carlo simulation of the collimator in a scintillation camera system can be time consuming. It is however necessary to take collimator interactions into account when simulating radionuclides emitting high-energy photons that can penetrate the septa in the collimator. In this work a statistical collimator algorithm, based on the Delta-Scattering method, is evaluated using &lt;sup&gt;123&lt;/sup&gt;L The evaluation is performed by comparing results from Monte Carlo simulations and measurements for a scintillation camera system, using point sources and an RSD phantom. A good agreement can be seen for both images and energy spectra.}},
  author       = {{Ljungberg, Michael and Larsson, A and Johansson, L}},
  booktitle    = {{IEEE Nuclear Science Symposium Conference Record}},
  issn         = {{1095-7863}},
  keywords     = {{Collimator algorithm; Scintialltion camera system; Collimator simulation; Delta-Scattering method}},
  language     = {{eng}},
  pages        = {{3584--3588}},
  publisher    = {{IEEE - Institute of Electrical and Electronics Engineers Inc.}},
  title        = {{A new collimator simulation in SIMIND based on the Delta-Scattering technique}},
  url          = {{http://dx.doi.org/10.1109/NSSMIC.2004.1466659}},
  doi          = {{10.1109/NSSMIC.2004.1466659}},
  volume       = {{6}},
  year         = {{2004}},
}