Development of a physical 3D anthropomorphic breast phantom

Carton, Ann Katherine ; Bakic, Predrag LU ; Ullberg, Christer ; Derand, Helen and Maidment, Andrew D.A. (2011) In Medical Physics 38(2). p.891-896
Abstract

Purpose: Develop a technique to fabricate a 3D anthropomorphic breast phantom with known ground truth for image quality assessment of 2D and 3D breast x-ray imaging systems. Methods: The phantom design is based on an existing computer model that can generate breast voxel phantoms of varying composition, size, and shape. The physical phantom is produced in two steps. First, the portion of the voxel phantom consisting of the glandular tissue, skin, and Cooper's ligaments is separated into sections. These sections are then fabricated by high-resolution rapid prototyping using a single material with 50% glandular equivalence. The remaining adipose compartments are then filled using an epoxy-based resin (EBR) with 100% adipose equivalence.... (More)

Purpose: Develop a technique to fabricate a 3D anthropomorphic breast phantom with known ground truth for image quality assessment of 2D and 3D breast x-ray imaging systems. Methods: The phantom design is based on an existing computer model that can generate breast voxel phantoms of varying composition, size, and shape. The physical phantom is produced in two steps. First, the portion of the voxel phantom consisting of the glandular tissue, skin, and Cooper's ligaments is separated into sections. These sections are then fabricated by high-resolution rapid prototyping using a single material with 50% glandular equivalence. The remaining adipose compartments are then filled using an epoxy-based resin (EBR) with 100% adipose equivalence. The phantom sections are stacked to form the physical anthropomorphic phantom. Results: The authors fabricated a prototype phantom corresponding to a 450 ml breast with 45% dense tissue, deformed to a 5 cm compressed thickness. Both the rapid prototype (RP) and EBR phantom materials are radiographically uniform. The coefficient of variation (CoV) of the relative attenuation between RP and EBR phantom samples was <1% and the CoV of the signal intensity within RP and EBR phantom samples was <1.5% on average. Digital mammography and reconstructed digital breast tomosynthesis images of the authors' phantom were reviewed by two radiologists; they reported that the images are similar in appearance to clinical images, noting there are still artifacts from air bubbles in the EBR. Conclusions: The authors have developed a technique to produce 3D anthropomorphic breast phantoms with known ground truth, yielding highly realistic x-ray images. Such phantoms may serve both qualitative and quantitative performance assessments for 2D and 3D breast x-ray imaging systems.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; and
publishing date
type
Contribution to journal
publication status
published
subject
keywords
anthropomorphic breast phantom, breast x-ray imaging, digital breast tomosynthesis, image quality assessment, mammography
in
Medical Physics
volume
38
issue
2
pages
6 pages
publisher
American Association of Physicists in Medicine
external identifiers
  • scopus:79551654718
ISSN
0094-2405
DOI
10.1118/1.3533896
language
English
LU publication?
no
id
80990f63-6328-4027-8de5-43f7d94101ab
date added to LUP
2020-11-07 13:16:49
date last changed
2022-04-19 01:53:04
@article{80990f63-6328-4027-8de5-43f7d94101ab,
  abstract     = {{<p>Purpose: Develop a technique to fabricate a 3D anthropomorphic breast phantom with known ground truth for image quality assessment of 2D and 3D breast x-ray imaging systems. Methods: The phantom design is based on an existing computer model that can generate breast voxel phantoms of varying composition, size, and shape. The physical phantom is produced in two steps. First, the portion of the voxel phantom consisting of the glandular tissue, skin, and Cooper's ligaments is separated into sections. These sections are then fabricated by high-resolution rapid prototyping using a single material with 50% glandular equivalence. The remaining adipose compartments are then filled using an epoxy-based resin (EBR) with 100% adipose equivalence. The phantom sections are stacked to form the physical anthropomorphic phantom. Results: The authors fabricated a prototype phantom corresponding to a 450 ml breast with 45% dense tissue, deformed to a 5 cm compressed thickness. Both the rapid prototype (RP) and EBR phantom materials are radiographically uniform. The coefficient of variation (CoV) of the relative attenuation between RP and EBR phantom samples was &lt;1% and the CoV of the signal intensity within RP and EBR phantom samples was &lt;1.5% on average. Digital mammography and reconstructed digital breast tomosynthesis images of the authors' phantom were reviewed by two radiologists; they reported that the images are similar in appearance to clinical images, noting there are still artifacts from air bubbles in the EBR. Conclusions: The authors have developed a technique to produce 3D anthropomorphic breast phantoms with known ground truth, yielding highly realistic x-ray images. Such phantoms may serve both qualitative and quantitative performance assessments for 2D and 3D breast x-ray imaging systems.</p>}},
  author       = {{Carton, Ann Katherine and Bakic, Predrag and Ullberg, Christer and Derand, Helen and Maidment, Andrew D.A.}},
  issn         = {{0094-2405}},
  keywords     = {{anthropomorphic breast phantom; breast x-ray imaging; digital breast tomosynthesis; image quality assessment; mammography}},
  language     = {{eng}},
  number       = {{2}},
  pages        = {{891--896}},
  publisher    = {{American Association of Physicists in Medicine}},
  series       = {{Medical Physics}},
  title        = {{Development of a physical 3D anthropomorphic breast phantom}},
  url          = {{http://dx.doi.org/10.1118/1.3533896}},
  doi          = {{10.1118/1.3533896}},
  volume       = {{38}},
  year         = {{2011}},
}