Skip to main content

Lund University Publications

LUND UNIVERSITY LIBRARIES

OVERVIEW, PRACTICAL TIPS AND POTENTIAL PITFALLS OF USING AUTOMATIC EXPOSURE CONTROL IN CT: SIEMENS CARE DOSE 4D.

Söderberg, Marcus LU orcid (2016) In Radiation Protection Dosimetry
Abstract
Today, computed tomography (CT) systems routinely use automatic exposure control (AEC), which modulates the tube current. However, for optimal use, there are several aspects of an AEC system that need to be considered. The purpose of this study was to provide an overview of the Siemens CARE Dose 4D AEC system, discuss practical tips and demonstrate potential pitfalls. Two adult anthropomorphic phantoms were examined using two different Siemens CT systems. When optimising the CT radiation dose and image quality, the projection angle of the localiser, patient centring, protocol selection, scanning direction and the use of protective devices requires special attention.
Please use this url to cite or link to this publication:
author
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Radiation Protection Dosimetry
publisher
Oxford University Press
external identifiers
  • pmid:26567324
  • pmid:26567324
  • scopus:84978932732
  • wos:000383492100014
ISSN
1742-3406
DOI
10.1093/rpd/ncv459
project
Evaluation of automatic exposure control in CT
language
English
LU publication?
yes
id
404a9c4f-9756-4614-8b2b-96f0265a2d28 (old id 8235545)
alternative location
http://www.ncbi.nlm.nih.gov/pubmed/26567324?dopt=Abstract
date added to LUP
2016-04-04 09:28:24
date last changed
2022-05-09 05:13:31
@article{404a9c4f-9756-4614-8b2b-96f0265a2d28,
  abstract     = {{Today, computed tomography (CT) systems routinely use automatic exposure control (AEC), which modulates the tube current. However, for optimal use, there are several aspects of an AEC system that need to be considered. The purpose of this study was to provide an overview of the Siemens CARE Dose 4D AEC system, discuss practical tips and demonstrate potential pitfalls. Two adult anthropomorphic phantoms were examined using two different Siemens CT systems. When optimising the CT radiation dose and image quality, the projection angle of the localiser, patient centring, protocol selection, scanning direction and the use of protective devices requires special attention.}},
  author       = {{Söderberg, Marcus}},
  issn         = {{1742-3406}},
  language     = {{eng}},
  publisher    = {{Oxford University Press}},
  series       = {{Radiation Protection Dosimetry}},
  title        = {{OVERVIEW, PRACTICAL TIPS AND POTENTIAL PITFALLS OF USING AUTOMATIC EXPOSURE CONTROL IN CT: SIEMENS CARE DOSE 4D.}},
  url          = {{https://lup.lub.lu.se/search/files/8064167/5333759.pdf}},
  doi          = {{10.1093/rpd/ncv459}},
  year         = {{2016}},
}