Heart Rate Estimation from Dual Pressure Sensors of a Dialysis Machine
(2015) Computing in Cardiology 42. p.29-32- Abstract
- Dialysis patients often suffer from cardiovascular dis- eases, motivating the use of continuous monitoring of car- diac activity in clinical routine. Cardiac pressure pulses propagate through the vascular system and enter the ex- tracorporeal blood circuit of a dialysis machine, where the pulses are captured by pressure sensors. The cardiac pulses are obscured by the much stronger pressure pulses originating from the peristaltic blood pump. We have pre- viously shown that a cardiac signal can be extracted from the venous pressure signal. However, that method has been found to perform less well at very low cardiac pressure pulse amplitudes. In the present study, we propose a novel method which addresses this issue by using the signals from... (More)
- Dialysis patients often suffer from cardiovascular dis- eases, motivating the use of continuous monitoring of car- diac activity in clinical routine. Cardiac pressure pulses propagate through the vascular system and enter the ex- tracorporeal blood circuit of a dialysis machine, where the pulses are captured by pressure sensors. The cardiac pulses are obscured by the much stronger pressure pulses originating from the peristaltic blood pump. We have pre- viously shown that a cardiac signal can be extracted from the venous pressure signal. However, that method has been found to perform less well at very low cardiac pressure pulse amplitudes. In the present study, we propose a novel method which addresses this issue by using the signals from both the arterial and the venous pressure sensors. The method is compared to the previous method on clini- cal data using a photoplethysmogram as reference. The re- sults suggests that heart rate can be estimated more accu- rately from pressure signals with lower cardiac signal am- plitude when both arterial and venous pressure are used, compared to when only the venous signal is used. (Less)
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/8571390
- author
- Holmer, Mattias LU ; Sandberg, Frida LU ; Solem, Kristian ; Olde, Bo and Sörnmo, Leif LU
- organization
- publishing date
- 2015
- type
- Chapter in Book/Report/Conference proceeding
- publication status
- published
- subject
- host publication
- 2015 Computing in Cardiology Conference (CinC)
- editor
- Murray, Allan
- volume
- 42
- pages
- 4 pages
- publisher
- IEEE - Institute of Electrical and Electronics Engineers Inc.
- conference name
- Computing in Cardiology
- conference location
- Nice, France
- conference dates
- 2015-09-06
- external identifiers
-
- scopus:84964033736
- ISSN
- 2325-8861
- DOI
- 10.1109/CIC.2015.7408578
- language
- English
- LU publication?
- yes
- id
- 5edc1e40-a22d-4564-9fce-145867c680ef (old id 8571390)
- alternative location
- http://www.cinc.org/archives/2015/pdf/0029.pdf
- date added to LUP
- 2016-04-04 09:12:36
- date last changed
- 2022-01-29 08:49:54
@inproceedings{5edc1e40-a22d-4564-9fce-145867c680ef, abstract = {{Dialysis patients often suffer from cardiovascular dis- eases, motivating the use of continuous monitoring of car- diac activity in clinical routine. Cardiac pressure pulses propagate through the vascular system and enter the ex- tracorporeal blood circuit of a dialysis machine, where the pulses are captured by pressure sensors. The cardiac pulses are obscured by the much stronger pressure pulses originating from the peristaltic blood pump. We have pre- viously shown that a cardiac signal can be extracted from the venous pressure signal. However, that method has been found to perform less well at very low cardiac pressure pulse amplitudes. In the present study, we propose a novel method which addresses this issue by using the signals from both the arterial and the venous pressure sensors. The method is compared to the previous method on clini- cal data using a photoplethysmogram as reference. The re- sults suggests that heart rate can be estimated more accu- rately from pressure signals with lower cardiac signal am- plitude when both arterial and venous pressure are used, compared to when only the venous signal is used.}}, author = {{Holmer, Mattias and Sandberg, Frida and Solem, Kristian and Olde, Bo and Sörnmo, Leif}}, booktitle = {{2015 Computing in Cardiology Conference (CinC)}}, editor = {{Murray, Allan}}, issn = {{2325-8861}}, language = {{eng}}, pages = {{29--32}}, publisher = {{IEEE - Institute of Electrical and Electronics Engineers Inc.}}, title = {{Heart Rate Estimation from Dual Pressure Sensors of a Dialysis Machine}}, url = {{http://dx.doi.org/10.1109/CIC.2015.7408578}}, doi = {{10.1109/CIC.2015.7408578}}, volume = {{42}}, year = {{2015}}, }