The effect of bone quality on the stability of ankle arthrodesis. A finite element study

Alonso-Vazquez, A ; Lauge-Pedersen, Henrik LU ; Lidgren, Lars LU and Taylor, M (2004) In Foot & Ankle International 25(11). p.840-850
Abstract
Background: Despite uniform operating techniques, lack of fusion still occurs after ankle arthrodesis. Differences in the biological healing potential may be a factor but the mechanical performance of the arthrodesis construct because of varying bone quality also may be important. Internal compression techniques are preferred because of higher union rates, shorter fusion times, and fewer complications. A three-screw configuration has been shown to be more stable than a two-screw configuration, but it is not obvious when it should be used. Methods: Three-dimensional finite element models of intact and flat-cut ankle arthrodeses were built, using two and three screws in different configurations. Poor bone quality was simulated by decreasing... (More)
Background: Despite uniform operating techniques, lack of fusion still occurs after ankle arthrodesis. Differences in the biological healing potential may be a factor but the mechanical performance of the arthrodesis construct because of varying bone quality also may be important. Internal compression techniques are preferred because of higher union rates, shorter fusion times, and fewer complications. A three-screw configuration has been shown to be more stable than a two-screw configuration, but it is not obvious when it should be used. Methods: Three-dimensional finite element models of intact and flat-cut ankle arthrodeses were built, using two and three screws in different configurations. Poor bone quality was simulated by decreasing Young's modulus of the bone. The constructs were loaded in torsion and dorsiflexion, and micromotions at the fusion site were measured. Results: Bone quality had a marked effect on the stability at the arthrodesis site. Inserting two screws at 30 degrees relative to the longitudinal axis of the tibia in an intact arthrodesis seemed the best option, especially as bone quality worsened. The addition of a third screw increased the stability at the arthrodesis site. Conclusions: Overall, intact joint surfaces and three-screw fixation, with the lateral and medial screws inserted produced the most stable arthrodesis constructs when bone quality was poor. Clinical Relevance. Ankle arthrodeses are technically demanding because of the shape and small size of the talus. Preoperative planning is an absolute necessity to determine placement and number of screws. This study shows that poor bone quality decreases the stability of the arthrodesis constructs, suggesting that an attempt should be made to create the most stable three-screw configuration. Finite element models can be used as an effective preoperative tool for planning screw number and placement. (Less)
Please use this url to cite or link to this publication:
author
; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
ankle arthrodesis, bone quality, finite element analysis
in
Foot & Ankle International
volume
25
issue
11
pages
840 - 850
publisher
SAGE Publications
external identifiers
  • wos:000225072000016
  • pmid:15574246
  • scopus:8444247531
ISSN
1944-7876
language
English
LU publication?
yes
id
902a905d-a404-45d4-9015-eabd2847b222 (old id 260980)
date added to LUP
2016-04-01 12:30:00
date last changed
2022-01-27 05:54:17
@article{902a905d-a404-45d4-9015-eabd2847b222,
  abstract     = {{Background: Despite uniform operating techniques, lack of fusion still occurs after ankle arthrodesis. Differences in the biological healing potential may be a factor but the mechanical performance of the arthrodesis construct because of varying bone quality also may be important. Internal compression techniques are preferred because of higher union rates, shorter fusion times, and fewer complications. A three-screw configuration has been shown to be more stable than a two-screw configuration, but it is not obvious when it should be used. Methods: Three-dimensional finite element models of intact and flat-cut ankle arthrodeses were built, using two and three screws in different configurations. Poor bone quality was simulated by decreasing Young's modulus of the bone. The constructs were loaded in torsion and dorsiflexion, and micromotions at the fusion site were measured. Results: Bone quality had a marked effect on the stability at the arthrodesis site. Inserting two screws at 30 degrees relative to the longitudinal axis of the tibia in an intact arthrodesis seemed the best option, especially as bone quality worsened. The addition of a third screw increased the stability at the arthrodesis site. Conclusions: Overall, intact joint surfaces and three-screw fixation, with the lateral and medial screws inserted produced the most stable arthrodesis constructs when bone quality was poor. Clinical Relevance. Ankle arthrodeses are technically demanding because of the shape and small size of the talus. Preoperative planning is an absolute necessity to determine placement and number of screws. This study shows that poor bone quality decreases the stability of the arthrodesis constructs, suggesting that an attempt should be made to create the most stable three-screw configuration. Finite element models can be used as an effective preoperative tool for planning screw number and placement.}},
  author       = {{Alonso-Vazquez, A and Lauge-Pedersen, Henrik and Lidgren, Lars and Taylor, M}},
  issn         = {{1944-7876}},
  keywords     = {{ankle arthrodesis; bone quality; finite element analysis}},
  language     = {{eng}},
  number       = {{11}},
  pages        = {{840--850}},
  publisher    = {{SAGE Publications}},
  series       = {{Foot & Ankle International}},
  title        = {{The effect of bone quality on the stability of ankle arthrodesis. A finite element study}},
  volume       = {{25}},
  year         = {{2004}},
}