Stationary tissue background correction increases the precision of clinical evaluation of intra-cardiac shunts by cardiovascular magnetic resonance

Nickander, Jannike ; Lundin, Magnus ; Abdula, Goran ; Jenner, Jonas ; Maret, Eva ; Sörensson, Peder ; Heiberg, Einar LU ; Sigfridsson, Andreas and Ugander, Martin LU (2020) In Scientific Reports 10(1).
Abstract

We aimed to evaluate the clinical utility of stationary tissue background phase correction for affecting precision in the measurement of Qp/Qs by cardiovascular magnetic resonance (CMR). We enrolled consecutive patients (n = 91) referred for CMR at 1.5T without suspicion of cardiac shunt, and patients (n = 10) with verified cardiac shunts in this retrospective study. All patients underwent phase contrast flow quantification in the ascending aorta and pulmonary trunk. Flow was quantified using two semi-automatic software platforms (SyngoVia VA30, Vendor 1; Segment 2.0R4534, Vendor 2). Measurements were performed both uncorrected and corrected for linear (Vendor 1 and Vendor 2) or quadratic (Vendor 2) background phase. The proportion of... (More)

We aimed to evaluate the clinical utility of stationary tissue background phase correction for affecting precision in the measurement of Qp/Qs by cardiovascular magnetic resonance (CMR). We enrolled consecutive patients (n = 91) referred for CMR at 1.5T without suspicion of cardiac shunt, and patients (n = 10) with verified cardiac shunts in this retrospective study. All patients underwent phase contrast flow quantification in the ascending aorta and pulmonary trunk. Flow was quantified using two semi-automatic software platforms (SyngoVia VA30, Vendor 1; Segment 2.0R4534, Vendor 2). Measurements were performed both uncorrected and corrected for linear (Vendor 1 and Vendor 2) or quadratic (Vendor 2) background phase. The proportion of patients outside the normal range of Qp/Qs was compared using the McNemar's test. Compared to uncorrected measurements, there were fewer patients with a Qp/Qs outside the normal range following linear correction using Vendor 1 (10% vs 18%, p < 0.001), and Vendor 2 (10% vs 18%, p < 0.001), and following quadratic correction using Vendor 2 (7% vs 18%, p < 0.001). No patient with known shunt was reclassified as normal following stationary background correction. Therefore, we conclude that stationary tissue background correction reduces the number of patients with a Qp/Qs ratio outside the normal range in a consecutive clinical population, while simultaneously not reclassifying any patient with known cardiac shunts as having a normal Qp/Qs. Stationary tissue background correction may be used in clinical patients to increase diagnostic precision.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; ; ; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Scientific Reports
volume
10
issue
1
article number
5053
publisher
Nature Publishing Group
external identifiers
  • pmid:32193468
  • scopus:85082049962
ISSN
2045-2322
DOI
10.1038/s41598-020-61812-7
language
English
LU publication?
yes
id
9f87fc08-9530-4092-92d9-6be70d86e695
date added to LUP
2020-04-03 12:01:44
date last changed
2024-04-03 05:33:46
@article{9f87fc08-9530-4092-92d9-6be70d86e695,
  abstract     = {{<p>We aimed to evaluate the clinical utility of stationary tissue background phase correction for affecting precision in the measurement of Qp/Qs by cardiovascular magnetic resonance (CMR). We enrolled consecutive patients (n = 91) referred for CMR at 1.5T without suspicion of cardiac shunt, and patients (n = 10) with verified cardiac shunts in this retrospective study. All patients underwent phase contrast flow quantification in the ascending aorta and pulmonary trunk. Flow was quantified using two semi-automatic software platforms (SyngoVia VA30, Vendor 1; Segment 2.0R4534, Vendor 2). Measurements were performed both uncorrected and corrected for linear (Vendor 1 and Vendor 2) or quadratic (Vendor 2) background phase. The proportion of patients outside the normal range of Qp/Qs was compared using the McNemar's test. Compared to uncorrected measurements, there were fewer patients with a Qp/Qs outside the normal range following linear correction using Vendor 1 (10% vs 18%, p &lt; 0.001), and Vendor 2 (10% vs 18%, p &lt; 0.001), and following quadratic correction using Vendor 2 (7% vs 18%, p &lt; 0.001). No patient with known shunt was reclassified as normal following stationary background correction. Therefore, we conclude that stationary tissue background correction reduces the number of patients with a Qp/Qs ratio outside the normal range in a consecutive clinical population, while simultaneously not reclassifying any patient with known cardiac shunts as having a normal Qp/Qs. Stationary tissue background correction may be used in clinical patients to increase diagnostic precision.</p>}},
  author       = {{Nickander, Jannike and Lundin, Magnus and Abdula, Goran and Jenner, Jonas and Maret, Eva and Sörensson, Peder and Heiberg, Einar and Sigfridsson, Andreas and Ugander, Martin}},
  issn         = {{2045-2322}},
  language     = {{eng}},
  month        = {{03}},
  number       = {{1}},
  publisher    = {{Nature Publishing Group}},
  series       = {{Scientific Reports}},
  title        = {{Stationary tissue background correction increases the precision of clinical evaluation of intra-cardiac shunts by cardiovascular magnetic resonance}},
  url          = {{http://dx.doi.org/10.1038/s41598-020-61812-7}},
  doi          = {{10.1038/s41598-020-61812-7}},
  volume       = {{10}},
  year         = {{2020}},
}