Characterization of a back-illuminated CMOS camera for soft x-ray coherent scattering

Desjardins, Kewin ; Popescu, Horia ; Mercère, Pascal ; Menneglier, Claude ; Gaudemer, Roland ; Thånell, Karina LU and Jaouen, Nicolas (2019) 13th International Conference on Synchrotron Radiation Instrumentation, SRI 2018 2054.
Abstract

A commercial scientific camera has been adapted and characterized at the SOLEIL Synchrotron with the aim to improve the acquisition capabilities on the soft X-ray coherent scattering experimental station at SEXTANTS beamline. This device is equipped by the last generation of back side illuminated scientific CMOS (BSI-sCMOS) of 2048 by 2048 pixels of 11 μm2 able to acquire low noise images with a frame rate up to 48 Hz. The camera's performance measurements have been done and shows a good level of readout noise, a large full-well capacity, a medium dark current and a good homogeneity, respectively, 1.6 e- rms (in High Gain mode), 80 000 e- (in Low Gain mode),<5 e-/pixel/s and ∼ 1%. The quantum efficiency (QE) measurement... (More)

A commercial scientific camera has been adapted and characterized at the SOLEIL Synchrotron with the aim to improve the acquisition capabilities on the soft X-ray coherent scattering experimental station at SEXTANTS beamline. This device is equipped by the last generation of back side illuminated scientific CMOS (BSI-sCMOS) of 2048 by 2048 pixels of 11 μm2 able to acquire low noise images with a frame rate up to 48 Hz. The camera's performance measurements have been done and shows a good level of readout noise, a large full-well capacity, a medium dark current and a good homogeneity, respectively, 1.6 e- rms (in High Gain mode), 80 000 e- (in Low Gain mode),<5 e-/pixel/s and ∼ 1%. The quantum efficiency (QE) measurement has been performed at the soft x-ray branch of the METROLOGIE beamline and gives a relatively good agreement with the expected theoretical values. Finally, the demonstration of the camera's efficiency and of the gain in useful time measurement related to the high frame rate have been performed with a series of Airy patterns images compared with an image recorded using the standard BSI-CDD already in operation at the SEXTANTS beamline.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; ; ; and
organization
publishing date
type
Chapter in Book/Report/Conference proceeding
publication status
published
subject
host publication
Proceedings of the 13th International Conference on Synchrotron Radiation Instrumentation, SRI 2018
editor
Gwo, Shangjr ; Chuang, Tzu-Hung ; Lee, Yao-Chang ; Chainani, Ashish Atma ; Chuang, Yu-Chun ; Wang, Chun-Chieh ; Huang, Di-Jing ; Wei, Der-Hsin ; Luo, Gwo-Huei ; Lee, Jyh-Fu ; Lin, Hong-Ji ; Ying, Gung-Chian ; Cheng, Cheng-Maw ; Lin, Jhih-Min ; Chen, Jin-Ming ; Chou, Ping-Jung ; Huang, Ching-Shiang ; Yang, Yaw-Wen ; Jeng, U-Ser ; Kuo, Chin-Cheng ; Sheu, Hwo-Shuenn ; Chung, Shih-Chun ; Lau, Wai-Keung ; Lee, Hsin-Yi and Tang, Mau-Tsu
volume
2054
article number
060066
publisher
American Institute of Physics (AIP)
conference name
13th International Conference on Synchrotron Radiation Instrumentation, SRI 2018
conference location
Taipei, Taiwan
conference dates
2018-06-11 - 2018-06-15
external identifiers
  • scopus:85060480634
ISBN
9780735417823
DOI
10.1063/1.5084697
language
English
LU publication?
yes
id
9fd0ad55-026a-4782-a668-fbf6706d9364
date added to LUP
2019-02-05 13:42:36
date last changed
2022-04-25 20:49:27
@inproceedings{9fd0ad55-026a-4782-a668-fbf6706d9364,
  abstract     = {{<p>A commercial scientific camera has been adapted and characterized at the SOLEIL Synchrotron with the aim to improve the acquisition capabilities on the soft X-ray coherent scattering experimental station at SEXTANTS beamline. This device is equipped by the last generation of back side illuminated scientific CMOS (BSI-sCMOS) of 2048 by 2048 pixels of 11 μm<sup>2</sup> able to acquire low noise images with a frame rate up to 48 Hz. The camera's performance measurements have been done and shows a good level of readout noise, a large full-well capacity, a medium dark current and a good homogeneity, respectively, 1.6 e- rms (in High Gain mode), 80 000 e- (in Low Gain mode),&lt;5 e-/pixel/s and ∼ 1%. The quantum efficiency (QE) measurement has been performed at the soft x-ray branch of the METROLOGIE beamline and gives a relatively good agreement with the expected theoretical values. Finally, the demonstration of the camera's efficiency and of the gain in useful time measurement related to the high frame rate have been performed with a series of Airy patterns images compared with an image recorded using the standard BSI-CDD already in operation at the SEXTANTS beamline.</p>}},
  author       = {{Desjardins, Kewin and Popescu, Horia and Mercère, Pascal and Menneglier, Claude and Gaudemer, Roland and Thånell, Karina and Jaouen, Nicolas}},
  booktitle    = {{Proceedings of the 13th International Conference on Synchrotron Radiation Instrumentation, SRI 2018}},
  editor       = {{Gwo, Shangjr and Chuang, Tzu-Hung and Lee, Yao-Chang and Chainani, Ashish Atma and Chuang, Yu-Chun and Wang, Chun-Chieh and Huang, Di-Jing and Wei, Der-Hsin and Luo, Gwo-Huei and Lee, Jyh-Fu and Lin, Hong-Ji and Ying, Gung-Chian and Cheng, Cheng-Maw and Lin, Jhih-Min and Chen, Jin-Ming and Chou, Ping-Jung and Huang, Ching-Shiang and Yang, Yaw-Wen and Jeng, U-Ser and Kuo, Chin-Cheng and Sheu, Hwo-Shuenn and Chung, Shih-Chun and Lau, Wai-Keung and Lee, Hsin-Yi and Tang, Mau-Tsu}},
  isbn         = {{9780735417823}},
  language     = {{eng}},
  month        = {{01}},
  publisher    = {{American Institute of Physics (AIP)}},
  title        = {{Characterization of a back-illuminated CMOS camera for soft x-ray coherent scattering}},
  url          = {{http://dx.doi.org/10.1063/1.5084697}},
  doi          = {{10.1063/1.5084697}},
  volume       = {{2054}},
  year         = {{2019}},
}