Determination of glucose in diluted blood with a thermal flow injection analysis biosensor

Harborn, Ulrika ; Xie, Bin LU and Danielsson, Bengt LU (1994) In Analytical Letters 27(14). p.2639-2645
Abstract

The glucose concentration in diluted whole blood has been measured, using a miniaturized thermal biosensor based on the enzyme thermistor principle. The biosensor is a small flow injection system. A sample volume of 20μl is injected into a flow of 50μ1/min. The heat produced when the sample passes the enzyme column is measured with a thermistor connected to a Wheatstone bridge. The enzyme column contains glucose oxidase and catalase co-immobilized on a solid support material. Samples of whole blood usually cause problems in flow-systems. The blood cells tend to block the enzyme column and the back pressure increases. We have tested a supeiporous agarose material as enzyme support material using tenfold diluted samples of whole human... (More)

The glucose concentration in diluted whole blood has been measured, using a miniaturized thermal biosensor based on the enzyme thermistor principle. The biosensor is a small flow injection system. A sample volume of 20μl is injected into a flow of 50μ1/min. The heat produced when the sample passes the enzyme column is measured with a thermistor connected to a Wheatstone bridge. The enzyme column contains glucose oxidase and catalase co-immobilized on a solid support material. Samples of whole blood usually cause problems in flow-systems. The blood cells tend to block the enzyme column and the back pressure increases. We have tested a supeiporous agarose material as enzyme support material using tenfold diluted samples of whole human blood. The blood was collected from the finger-tip and diluted with buffer containing an anticoagulant and sodium fluoride. The number of samples possible to inject and the accuracy compared to the Boehringer Mannheim Reflolux have been determined. At least 100 ten-fold diluted blood samples could be injected on a micro-column of superporous agarose. The obtained glucose concentration correlated well with the one obtained with the reference instrument.

(Less)
Please use this url to cite or link to this publication:
author
; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
glucose, glucose oxidase, thermal biosensor, whole blood
in
Analytical Letters
volume
27
issue
14
pages
7 pages
publisher
Taylor & Francis
external identifiers
  • scopus:0028011649
ISSN
0003-2719
DOI
10.1080/00032719408005996
language
English
LU publication?
yes
id
a47ee2db-6a0e-420b-9ae7-372ed2d0bb37
date added to LUP
2021-10-27 16:00:22
date last changed
2021-11-01 11:46:52
@article{a47ee2db-6a0e-420b-9ae7-372ed2d0bb37,
  abstract     = {{<p>The glucose concentration in diluted whole blood has been measured, using a miniaturized thermal biosensor based on the enzyme thermistor principle. The biosensor is a small flow injection system. A sample volume of 20μl is injected into a flow of 50μ1/min. The heat produced when the sample passes the enzyme column is measured with a thermistor connected to a Wheatstone bridge. The enzyme column contains glucose oxidase and catalase co-immobilized on a solid support material. Samples of whole blood usually cause problems in flow-systems. The blood cells tend to block the enzyme column and the back pressure increases. We have tested a supeiporous agarose material as enzyme support material using tenfold diluted samples of whole human blood. The blood was collected from the finger-tip and diluted with buffer containing an anticoagulant and sodium fluoride. The number of samples possible to inject and the accuracy compared to the Boehringer Mannheim Reflolux have been determined. At least 100 ten-fold diluted blood samples could be injected on a micro-column of superporous agarose. The obtained glucose concentration correlated well with the one obtained with the reference instrument.</p>}},
  author       = {{Harborn, Ulrika and Xie, Bin and Danielsson, Bengt}},
  issn         = {{0003-2719}},
  keywords     = {{glucose; glucose oxidase; thermal biosensor; whole blood}},
  language     = {{eng}},
  number       = {{14}},
  pages        = {{2639--2645}},
  publisher    = {{Taylor & Francis}},
  series       = {{Analytical Letters}},
  title        = {{Determination of glucose in diluted blood with a thermal flow injection analysis biosensor}},
  url          = {{http://dx.doi.org/10.1080/00032719408005996}},
  doi          = {{10.1080/00032719408005996}},
  volume       = {{27}},
  year         = {{1994}},
}