Blood compatibility of widely used central venous catheters; an experimental study

Thorarinsdottir, Hulda LU ; Kander, Thomas LU orcid ; Johansson, Dorota ; Nilsson, Bo ; Klarin, Bengt LU and Sanchez, Javier (2022) In Scientific Reports 12. p.1-10
Abstract

An inserted central venous catheter (CVC) is considered foreign material by the inert host defence systems and induce inflammation and thrombus formation. The objective of this study was to evaluate blood compatibility of six commonly used CVCs. Three coated and three uncoated CVC materials were tested in a modified Chandler loop model. Each catheter material circulated in blood from ten different healthy volunteers for 1 h. Blood cell counts and measurements of the inert host defence systems were performed on blood samples from the loop. All the tested catheters demonstrated impact on blood cells, contact coagulation, the complement system, or inflammatory markers, although the impact varied significantly. Of the catheters we... (More)

An inserted central venous catheter (CVC) is considered foreign material by the inert host defence systems and induce inflammation and thrombus formation. The objective of this study was to evaluate blood compatibility of six commonly used CVCs. Three coated and three uncoated CVC materials were tested in a modified Chandler loop model. Each catheter material circulated in blood from ten different healthy volunteers for 1 h. Blood cell counts and measurements of the inert host defence systems were performed on blood samples from the loop. All the tested catheters demonstrated impact on blood cells, contact coagulation, the complement system, or inflammatory markers, although the impact varied significantly. Of the catheters we evaluated, the most unfavourable blood compatibility profile was found for the polyurethane CVC coated with chlorohexidine and silver sulfadiazine. The greatest variation in blood compatibility between test runs was noted for the silicone dialysis catheter. Poor blood compatibility should be taken seriously but given the experimental design of the current study the clinical significance remains to be evaluated.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
Catheterization, Central Venous/adverse effects, Central Venous Catheters, Humans, Polyurethanes, Renal Dialysis, Silicones
in
Scientific Reports
volume
12
article number
8600
pages
1 - 10
publisher
Nature Publishing Group
external identifiers
  • pmid:35597879
  • scopus:85130364447
ISSN
2045-2322
DOI
10.1038/s41598-022-12564-z
language
English
LU publication?
yes
additional info
© 2022. The Author(s).
id
aa885391-caa1-42c1-9c69-878cd7588a17
date added to LUP
2022-05-26 22:37:49
date last changed
2024-06-13 15:30:48
@article{aa885391-caa1-42c1-9c69-878cd7588a17,
  abstract     = {{<p>An inserted central venous catheter (CVC) is considered foreign material by the inert host defence systems and induce inflammation and thrombus formation. The objective of this study was to evaluate blood compatibility of six commonly used CVCs. Three coated and three uncoated CVC materials were tested in a modified Chandler loop model. Each catheter material circulated in blood from ten different healthy volunteers for 1 h. Blood cell counts and measurements of the inert host defence systems were performed on blood samples from the loop. All the tested catheters demonstrated impact on blood cells, contact coagulation, the complement system, or inflammatory markers, although the impact varied significantly. Of the catheters we evaluated, the most unfavourable blood compatibility profile was found for the polyurethane CVC coated with chlorohexidine and silver sulfadiazine. The greatest variation in blood compatibility between test runs was noted for the silicone dialysis catheter. Poor blood compatibility should be taken seriously but given the experimental design of the current study the clinical significance remains to be evaluated.</p>}},
  author       = {{Thorarinsdottir, Hulda and Kander, Thomas and Johansson, Dorota and Nilsson, Bo and Klarin, Bengt and Sanchez, Javier}},
  issn         = {{2045-2322}},
  keywords     = {{Catheterization, Central Venous/adverse effects; Central Venous Catheters; Humans; Polyurethanes; Renal Dialysis; Silicones}},
  language     = {{eng}},
  pages        = {{1--10}},
  publisher    = {{Nature Publishing Group}},
  series       = {{Scientific Reports}},
  title        = {{Blood compatibility of widely used central venous catheters; an experimental study}},
  url          = {{http://dx.doi.org/10.1038/s41598-022-12564-z}},
  doi          = {{10.1038/s41598-022-12564-z}},
  volume       = {{12}},
  year         = {{2022}},
}