Vitamin E infused highly cross-linked cemented cups in total hip arthroplasty show good wear pattern and stabilize satisfactorily : a randomized, controlled RSA trial with 5-year follow-up
(2022) In Acta Orthopaedica 93. p.249-255- Abstract
Background and purpose - Vitamin E infused highly cross-linked polyethylene (VEPE) was introduced in order to enhance oxidative resistance in highly cross-linked polyethylene cups in total hip arthroplasty (THA). We have, with a follow-up of 5 years, evaluated wear characteristics of 2 identically designed cemented cups with the only difference being the material, VEPE or ultra-high molecular weight polyethylene (UHMWPE). Furthermore, we report cup migration and clinical outcome. Patients and methods - 48 patients with primary osteoarthritis were randomized to either UHMWPE or VEPE cups. Patients were followed with radiostereometric analysis (RSA) from the first postoperative day, at 3 months, 1, 2, and 5 years as well as with... (More)
Background and purpose - Vitamin E infused highly cross-linked polyethylene (VEPE) was introduced in order to enhance oxidative resistance in highly cross-linked polyethylene cups in total hip arthroplasty (THA). We have, with a follow-up of 5 years, evaluated wear characteristics of 2 identically designed cemented cups with the only difference being the material, VEPE or ultra-high molecular weight polyethylene (UHMWPE). Furthermore, we report cup migration and clinical outcome. Patients and methods - 48 patients with primary osteoarthritis were randomized to either UHMWPE or VEPE cups. Patients were followed with radiostereometric analysis (RSA) from the first postoperative day, at 3 months, 1, 2, and 5 years as well as with hip-specific outcome questionnaires. Results - At 3 months the mean proximal head penetration for UHMWPE was 0.07 mm (95% CI 0.03-0.11) and for VEPE 0.06 mm (-0.01 to 0.13). Thereafter, there was a continuous annual wear of 0.08 mm/year, up to 0.46 mm (0.36-0.57) at 5 years, for the UHMWPE cup. The VEPE cup showed low annual wear of 0.01 mm/year, up to 0.09 mm (0.02-0.16) at 5 years. In the first 3 months the UHMWPE cup migrated cranially 0.08 mm (0.03-0.13) whereas the VEPE cup migrated 0.17 mm (0.10-0.24), Thereafter, they showed similar migration patterns with stabilization between 2 and 5 years up to 0.21 mm (0.04-0.39) and 0.24 mm (0.13-0.36) respectively. The HOOS remained good up to 5 years, and no cup was revised. Interpretation - Compared with otherwise identical UHMWPE cups the cemented VEPE cup shows statistically significant reduction of wear up to 5 years and both cup types stabilize well with good clinical outcome.
(Less)
- author
- Bergvinsson, Halldor
LU
; Zampelis, Vasilis
LU
; Sundberg, Martin
LU
; Tjörnstrand, Jon
LU
and Flivik, Gunnar
LU
- organization
- publishing date
- 2022
- type
- Contribution to journal
- publication status
- published
- subject
- in
- Acta Orthopaedica
- volume
- 93
- pages
- 249 - 255
- publisher
- Taylor & Francis
- external identifiers
-
- scopus:85123663063
- pmid:35048993
- ISSN
- 1745-3682
- DOI
- 10.2340/17453674.2022.1517
- language
- English
- LU publication?
- yes
- id
- ba114e98-ec47-40ca-a738-ec4187c4949c
- date added to LUP
- 2022-01-31 00:46:14
- date last changed
- 2025-10-20 07:49:30
@article{ba114e98-ec47-40ca-a738-ec4187c4949c,
abstract = {{<p>Background and purpose - Vitamin E infused highly cross-linked polyethylene (VEPE) was introduced in order to enhance oxidative resistance in highly cross-linked polyethylene cups in total hip arthroplasty (THA). We have, with a follow-up of 5 years, evaluated wear characteristics of 2 identically designed cemented cups with the only difference being the material, VEPE or ultra-high molecular weight polyethylene (UHMWPE). Furthermore, we report cup migration and clinical outcome. Patients and methods - 48 patients with primary osteoarthritis were randomized to either UHMWPE or VEPE cups. Patients were followed with radiostereometric analysis (RSA) from the first postoperative day, at 3 months, 1, 2, and 5 years as well as with hip-specific outcome questionnaires. Results - At 3 months the mean proximal head penetration for UHMWPE was 0.07 mm (95% CI 0.03-0.11) and for VEPE 0.06 mm (-0.01 to 0.13). Thereafter, there was a continuous annual wear of 0.08 mm/year, up to 0.46 mm (0.36-0.57) at 5 years, for the UHMWPE cup. The VEPE cup showed low annual wear of 0.01 mm/year, up to 0.09 mm (0.02-0.16) at 5 years. In the first 3 months the UHMWPE cup migrated cranially 0.08 mm (0.03-0.13) whereas the VEPE cup migrated 0.17 mm (0.10-0.24), Thereafter, they showed similar migration patterns with stabilization between 2 and 5 years up to 0.21 mm (0.04-0.39) and 0.24 mm (0.13-0.36) respectively. The HOOS remained good up to 5 years, and no cup was revised. Interpretation - Compared with otherwise identical UHMWPE cups the cemented VEPE cup shows statistically significant reduction of wear up to 5 years and both cup types stabilize well with good clinical outcome.</p>}},
author = {{Bergvinsson, Halldor and Zampelis, Vasilis and Sundberg, Martin and Tjörnstrand, Jon and Flivik, Gunnar}},
issn = {{1745-3682}},
language = {{eng}},
pages = {{249--255}},
publisher = {{Taylor & Francis}},
series = {{Acta Orthopaedica}},
title = {{Vitamin E infused highly cross-linked cemented cups in total hip arthroplasty show good wear pattern and stabilize satisfactorily : a randomized, controlled RSA trial with 5-year follow-up}},
url = {{http://dx.doi.org/10.2340/17453674.2022.1517}},
doi = {{10.2340/17453674.2022.1517}},
volume = {{93}},
year = {{2022}},
}