Skip to main content

Lund University Publications

LUND UNIVERSITY LIBRARIES

Vitamin E infused highly cross-linked cemented cups in total hip arthroplasty show good wear pattern and stabilize satisfactorily : a randomized, controlled RSA trial with 5-year follow-up

Bergvinsson, Halldor LU orcid ; Zampelis, Vasilis LU ; Sundberg, Martin LU orcid ; Tjörnstrand, Jon LU and Flivik, Gunnar LU (2022) In Acta Orthopaedica 93. p.249-255
Abstract

Background and purpose - Vitamin E infused highly cross-linked polyethylene (VEPE) was introduced in order to enhance oxidative resistance in highly cross-linked polyethylene cups in total hip arthroplasty (THA). We have, with a follow-up of 5 years, evaluated wear characteristics of 2 identically designed cemented cups with the only difference being the material, VEPE or ultra-high molecular weight polyethylene (UHMWPE). Furthermore, we report cup migration and clinical outcome. Patients and methods - 48 patients with primary osteoarthritis were randomized to either UHMWPE or VEPE cups. Patients were followed with radiostereometric analysis (RSA) from the first postoperative day, at 3 months, 1, 2, and 5 years as well as with... (More)

Background and purpose - Vitamin E infused highly cross-linked polyethylene (VEPE) was introduced in order to enhance oxidative resistance in highly cross-linked polyethylene cups in total hip arthroplasty (THA). We have, with a follow-up of 5 years, evaluated wear characteristics of 2 identically designed cemented cups with the only difference being the material, VEPE or ultra-high molecular weight polyethylene (UHMWPE). Furthermore, we report cup migration and clinical outcome. Patients and methods - 48 patients with primary osteoarthritis were randomized to either UHMWPE or VEPE cups. Patients were followed with radiostereometric analysis (RSA) from the first postoperative day, at 3 months, 1, 2, and 5 years as well as with hip-specific outcome questionnaires. Results - At 3 months the mean proximal head penetration for UHMWPE was 0.07 mm (95% CI 0.03-0.11) and for VEPE 0.06 mm (-0.01 to 0.13). Thereafter, there was a continuous annual wear of 0.08 mm/year, up to 0.46 mm (0.36-0.57) at 5 years, for the UHMWPE cup. The VEPE cup showed low annual wear of 0.01 mm/year, up to 0.09 mm (0.02-0.16) at 5 years. In the first 3 months the UHMWPE cup migrated cranially 0.08 mm (0.03-0.13) whereas the VEPE cup migrated 0.17 mm (0.10-0.24), Thereafter, they showed similar migration patterns with stabilization between 2 and 5 years up to 0.21 mm (0.04-0.39) and 0.24 mm (0.13-0.36) respectively. The HOOS remained good up to 5 years, and no cup was revised. Interpretation - Compared with otherwise identical UHMWPE cups the cemented VEPE cup shows statistically significant reduction of wear up to 5 years and both cup types stabilize well with good clinical outcome.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
in
Acta Orthopaedica
volume
93
pages
249 - 255
publisher
Taylor & Francis
external identifiers
  • pmid:35048993
  • scopus:85123663063
ISSN
1745-3682
DOI
10.2340/17453674.2022.1517
language
English
LU publication?
yes
id
ba114e98-ec47-40ca-a738-ec4187c4949c
date added to LUP
2022-01-31 00:46:14
date last changed
2024-06-15 10:17:37
@article{ba114e98-ec47-40ca-a738-ec4187c4949c,
  abstract     = {{<p>Background and purpose - Vitamin E infused highly cross-linked polyethylene (VEPE) was introduced in order to enhance oxidative resistance in highly cross-linked polyethylene cups in total hip arthroplasty (THA). We have, with a follow-up of 5 years, evaluated wear characteristics of 2 identically designed cemented cups with the only difference being the material, VEPE or ultra-high molecular weight polyethylene (UHMWPE). Furthermore, we report cup migration and clinical outcome. Patients and methods - 48 patients with primary osteoarthritis were randomized to either UHMWPE or VEPE cups. Patients were followed with radiostereometric analysis (RSA) from the first postoperative day, at 3 months, 1, 2, and 5 years as well as with hip-specific outcome questionnaires. Results - At 3 months the mean proximal head penetration for UHMWPE was 0.07 mm (95% CI 0.03-0.11) and for VEPE 0.06 mm (-0.01 to 0.13). Thereafter, there was a continuous annual wear of 0.08 mm/year, up to 0.46 mm (0.36-0.57) at 5 years, for the UHMWPE cup. The VEPE cup showed low annual wear of 0.01 mm/year, up to 0.09 mm (0.02-0.16) at 5 years. In the first 3 months the UHMWPE cup migrated cranially 0.08 mm (0.03-0.13) whereas the VEPE cup migrated 0.17 mm (0.10-0.24), Thereafter, they showed similar migration patterns with stabilization between 2 and 5 years up to 0.21 mm (0.04-0.39) and 0.24 mm (0.13-0.36) respectively. The HOOS remained good up to 5 years, and no cup was revised. Interpretation - Compared with otherwise identical UHMWPE cups the cemented VEPE cup shows statistically significant reduction of wear up to 5 years and both cup types stabilize well with good clinical outcome.</p>}},
  author       = {{Bergvinsson, Halldor and Zampelis, Vasilis and Sundberg, Martin and Tjörnstrand, Jon and Flivik, Gunnar}},
  issn         = {{1745-3682}},
  language     = {{eng}},
  pages        = {{249--255}},
  publisher    = {{Taylor & Francis}},
  series       = {{Acta Orthopaedica}},
  title        = {{Vitamin E infused highly cross-linked cemented cups in total hip arthroplasty show good wear pattern and stabilize satisfactorily : a randomized, controlled RSA trial with 5-year follow-up}},
  url          = {{http://dx.doi.org/10.2340/17453674.2022.1517}},
  doi          = {{10.2340/17453674.2022.1517}},
  volume       = {{93}},
  year         = {{2022}},
}