Computerized profile perimetry in glaucoma

Heijl, A LU and Drance, S M (1980) In Archives of Ophthalmology 98(12). p.201-2199
Abstract

The new profile testing mode of the COMPETER automatic perimeter was clinically tested and compared with careful manual profile perimetry on the Tübinger perimeter. Each of 110 patients with glaucomatous field defects, patients with suspected glaucoma, and normal subjects had one meridian tested. All of the 55 patients who showed an abnormal result on the manual profile also had abnormal findings on the automatic test. Two (4%) of the profiles labeled as normal on initial manual perimetry were identified by the automatic perimeter to be abnormal, and these defects were confirmed on repeated manual profile perimetry. The automatic perimetry gave five (9%) false-positive results in the 53 normal manual profiles, thus giving a specificity... (More)

The new profile testing mode of the COMPETER automatic perimeter was clinically tested and compared with careful manual profile perimetry on the Tübinger perimeter. Each of 110 patients with glaucomatous field defects, patients with suspected glaucoma, and normal subjects had one meridian tested. All of the 55 patients who showed an abnormal result on the manual profile also had abnormal findings on the automatic test. Two (4%) of the profiles labeled as normal on initial manual perimetry were identified by the automatic perimeter to be abnormal, and these defects were confirmed on repeated manual profile perimetry. The automatic perimetry gave five (9%) false-positive results in the 53 normal manual profiles, thus giving a specificity of 91%. In two of these cases, the automatic profile indicated defects in areas where the manual profile appeared normal, but there was abnormality in the manual kinetic perimetry in the affected area. The automatic profiles of the COMPETER automatic perimeter are accurate, sensitive, time-saving, and clinically useful.

(Less)
Please use this url to cite or link to this publication:
author
and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
Adolescent, Adult, Aged, Computers, False Positive Reactions, Glaucoma/diagnosis, Humans, Middle Aged, Vision Disorders/diagnosis, Visual Field Tests/methods, Visual Fields
in
Archives of Ophthalmology
volume
98
issue
12
pages
3 pages
publisher
American Medical Association
external identifiers
  • pmid:7447773
  • scopus:0019215063
ISSN
0003-9950
language
English
LU publication?
yes
id
ba8ce5e1-0f8b-470e-95c4-0a0dc941d060
date added to LUP
2019-06-30 16:36:56
date last changed
2024-01-01 14:46:18
@article{ba8ce5e1-0f8b-470e-95c4-0a0dc941d060,
  abstract     = {{<p>The new profile testing mode of the COMPETER automatic perimeter was clinically tested and compared with careful manual profile perimetry on the Tübinger perimeter. Each of 110 patients with glaucomatous field defects, patients with suspected glaucoma, and normal subjects had one meridian tested. All of the 55 patients who showed an abnormal result on the manual profile also had abnormal findings on the automatic test. Two (4%) of the profiles labeled as normal on initial manual perimetry were identified by the automatic perimeter to be abnormal, and these defects were confirmed on repeated manual profile perimetry. The automatic perimetry gave five (9%) false-positive results in the 53 normal manual profiles, thus giving a specificity of 91%. In two of these cases, the automatic profile indicated defects in areas where the manual profile appeared normal, but there was abnormality in the manual kinetic perimetry in the affected area. The automatic profiles of the COMPETER automatic perimeter are accurate, sensitive, time-saving, and clinically useful.</p>}},
  author       = {{Heijl, A and Drance, S M}},
  issn         = {{0003-9950}},
  keywords     = {{Adolescent; Adult; Aged; Computers; False Positive Reactions; Glaucoma/diagnosis; Humans; Middle Aged; Vision Disorders/diagnosis; Visual Field Tests/methods; Visual Fields}},
  language     = {{eng}},
  number       = {{12}},
  pages        = {{201--2199}},
  publisher    = {{American Medical Association}},
  series       = {{Archives of Ophthalmology}},
  title        = {{Computerized profile perimetry in glaucoma}},
  volume       = {{98}},
  year         = {{1980}},
}