Masses detection in breast tomosynthesis and digital mammography: a model observer study
(2009) SPIE Medical Imaging, 2009 7263.- Abstract
- In this study, we adapt and apply model observers within the framework of realistic detection tasks in breast tomosynthesis (BT). We use images consisting of realistic masses digitally embedded in real patient anatomical backgrounds, and we adapt specific model observers that have been previously applied to digital mammography (DM). We design alternative forced-choice experiments (AFC) studies for DM and BT tasks in the signal known exactly but variable (SKEV) framework. We compare performance of various linear model observers (non-prewhitening matched filter with an eye filter, and several channelized Hotelling observers (CHO) against human. A good agreement in performance between human and model observers can be obtained when an... (More)
- In this study, we adapt and apply model observers within the framework of realistic detection tasks in breast tomosynthesis (BT). We use images consisting of realistic masses digitally embedded in real patient anatomical backgrounds, and we adapt specific model observers that have been previously applied to digital mammography (DM). We design alternative forced-choice experiments (AFC) studies for DM and BT tasks in the signal known exactly but variable (SKEV) framework. We compare performance of various linear model observers (non-prewhitening matched filter with an eye filter, and several channelized Hotelling observers (CHO) against human. A good agreement in performance between human and model observers can be obtained when an appropriate internal noise level is adopted. Models achieve the same detection performance across BT and DM with about three times less projected signal intensity in BT than in DM (humans: 3.8), due to the anatomical noise reduction in BT. We suggest that, in the future, model observers can potentially be used as an objective tool for automating the optimization of BT acquisition parameters or reconstruction algorithms, or narrowing a wide span of possible parameter combinations, without requiring human observers studies. (Less)
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/3045825
- author
- Castella, Cyril ; Ruschin, Mark LU ; Eckstein, Miguel P ; Kinkel, K ; Verdun, Francis R ; Tingberg, Anders LU and Bochud, Francois O
- organization
- publishing date
- 2009
- type
- Chapter in Book/Report/Conference proceeding
- publication status
- published
- subject
- host publication
- [Host publication title missing]
- volume
- 7263
- publisher
- SPIE
- conference name
- SPIE Medical Imaging, 2009
- conference location
- Orlando, FL, United States
- conference dates
- 2009-02-07 - 2009-02-12
- external identifiers
-
- wos:000306174400023
- scopus:67249101322
- DOI
- 10.1117/12.811131
- language
- English
- LU publication?
- yes
- id
- bcff7851-500a-49c9-bf57-38c089237155 (old id 3045825)
- alternative location
- http://proceedings.spiedigitallibrary.org/proceeding.aspx?articleid=1335925
- http://infoscience.epfl.ch/record/133431/files/EPFL_TH4347.pdf#page=79
- date added to LUP
- 2016-04-04 11:21:53
- date last changed
- 2024-04-13 16:21:44
@inproceedings{bcff7851-500a-49c9-bf57-38c089237155, abstract = {{In this study, we adapt and apply model observers within the framework of realistic detection tasks in breast tomosynthesis (BT). We use images consisting of realistic masses digitally embedded in real patient anatomical backgrounds, and we adapt specific model observers that have been previously applied to digital mammography (DM). We design alternative forced-choice experiments (AFC) studies for DM and BT tasks in the signal known exactly but variable (SKEV) framework. We compare performance of various linear model observers (non-prewhitening matched filter with an eye filter, and several channelized Hotelling observers (CHO) against human. A good agreement in performance between human and model observers can be obtained when an appropriate internal noise level is adopted. Models achieve the same detection performance across BT and DM with about three times less projected signal intensity in BT than in DM (humans: 3.8), due to the anatomical noise reduction in BT. We suggest that, in the future, model observers can potentially be used as an objective tool for automating the optimization of BT acquisition parameters or reconstruction algorithms, or narrowing a wide span of possible parameter combinations, without requiring human observers studies.}}, author = {{Castella, Cyril and Ruschin, Mark and Eckstein, Miguel P and Kinkel, K and Verdun, Francis R and Tingberg, Anders and Bochud, Francois O}}, booktitle = {{[Host publication title missing]}}, language = {{eng}}, publisher = {{SPIE}}, title = {{Masses detection in breast tomosynthesis and digital mammography: a model observer study}}, url = {{http://dx.doi.org/10.1117/12.811131}}, doi = {{10.1117/12.811131}}, volume = {{7263}}, year = {{2009}}, }