Signal alterations, artifacts and image distortion induced by a superparamagnetic contrast medium. A phantom study in a 0.3 tesla MR system
(1990) In Acta Radiologica 31(2). p.213-216- Abstract
- In a phantom study using a 0.3 tesla MR system the signal alterations, artifacts and image distortion induced by a bowel contrast medium containing superparamagnetic magnetite particles were evaluated. A concentration of 20 mg iron/l was most useful because it caused a significant signal reduction without disturbing artifacts and image distortion. The signal reduction was most pronounced on long TR/TE spin echo sequences. The appearance and distribution of artifacts are related to the direction of the static magnetic field and frequency encoding gradient. The size of artifacts is related to the concentration of the contrast medium and the band width used.
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/1105256
- author
- Holtås, Stig LU ; Wallengren, Nils-Olof LU ; Ericsson, A and Bach-Gansmo, T
- organization
- publishing date
- 1990
- type
- Contribution to journal
- publication status
- published
- subject
- in
- Acta Radiologica
- volume
- 31
- issue
- 2
- pages
- 213 - 216
- publisher
- SAGE Publications
- external identifiers
-
- pmid:2372467
- scopus:0025307065
- ISSN
- 1600-0455
- language
- English
- LU publication?
- yes
- id
- cdc438e4-5fce-4278-980b-2affa8ba1614 (old id 1105256)
- date added to LUP
- 2016-04-01 17:14:08
- date last changed
- 2021-01-03 09:56:26
@article{cdc438e4-5fce-4278-980b-2affa8ba1614, abstract = {{In a phantom study using a 0.3 tesla MR system the signal alterations, artifacts and image distortion induced by a bowel contrast medium containing superparamagnetic magnetite particles were evaluated. A concentration of 20 mg iron/l was most useful because it caused a significant signal reduction without disturbing artifacts and image distortion. The signal reduction was most pronounced on long TR/TE spin echo sequences. The appearance and distribution of artifacts are related to the direction of the static magnetic field and frequency encoding gradient. The size of artifacts is related to the concentration of the contrast medium and the band width used.}}, author = {{Holtås, Stig and Wallengren, Nils-Olof and Ericsson, A and Bach-Gansmo, T}}, issn = {{1600-0455}}, language = {{eng}}, number = {{2}}, pages = {{213--216}}, publisher = {{SAGE Publications}}, series = {{Acta Radiologica}}, title = {{Signal alterations, artifacts and image distortion induced by a superparamagnetic contrast medium. A phantom study in a 0.3 tesla MR system}}, volume = {{31}}, year = {{1990}}, }