Aim: This paper describes the quality assurance (QA) work performed in the Swedish multicenter ARTSCAN (Accelerated RadioTherapy of Squamous cell CArcinomas in the head and Neck) trial to guarantee high quality in a multicenter study which involved modern radiotherapy such as 3DCRT or IMRT. Materials and methods: The study was closed in June 2006 with 750 randomised patients. Radiation therapy-related data for every patient were sent by each participating centre to the QA office where all trial data were reviewed, analysed and stored. In case of any deviation from the protocol, an interactive process was started between the QA office and the local responsible clinician and/or physicist to increase the compliance to the protocol for future... (More)
Aim: This paper describes the quality assurance (QA) work performed in the Swedish multicenter ARTSCAN (Accelerated RadioTherapy of Squamous cell CArcinomas in the head and Neck) trial to guarantee high quality in a multicenter study which involved modern radiotherapy such as 3DCRT or IMRT. Materials and methods: The study was closed in June 2006 with 750 randomised patients. Radiation therapy-related data for every patient were sent by each participating centre to the QA office where all trial data were reviewed, analysed and stored. In case of any deviation from the protocol, an interactive process was started between the QA office and the local responsible clinician and/or physicist to increase the compliance to the protocol for future randomised patients. Meetings and workshops were held on a regular basis for discussions on various trial-related issues and for the QA office to report on updated results. Results and discussion: This review covers the 734 patients out of a total of 750 who had entered the study. Deviations early in the study were corrected so that the overall compliance to the protocol was very high. There were only negligible variations in doses and dose distributions to target volumes for each specific site and stage. The quality of the treatments was high. Furthermore, an extensive database of treatment parameters was accumulated for future dose-volume vs. endpoint evaluations. Conclusions: This comprehensive QA programme increased the probability to draw firm conclusions from our study and may serve as a concept for QA work in future radiotherapy trials where comparatively small effects are searched for in a heterogeneous tumour population. (Less)
Please use this url to cite or link to this publication:
author
, et al. (More)
(Less)
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
head and neck, quality assurance, multicentre trial, radiotherapy, 3DCRT, IMRT
in
Radiotherapy and Oncology
volume
87
issue
2
pages
290 - 299
publisher
Elsevier
external identifiers
  • wos:000256774800020
  • scopus:43049165323
ISSN
1879-0887
DOI
10.1016/j.radonc.2007.12.005
language
English
LU publication?
yes
id
d28dbbf3-44c9-46a8-bf61-f7e1286abdcf (old id 1187293)
date added to LUP
2016-04-01 11:50:46
date last changed
2019-11-13 01:52:11
@article{d28dbbf3-44c9-46a8-bf61-f7e1286abdcf,
  abstract     = {Aim: This paper describes the quality assurance (QA) work performed in the Swedish multicenter ARTSCAN (Accelerated RadioTherapy of Squamous cell CArcinomas in the head and Neck) trial to guarantee high quality in a multicenter study which involved modern radiotherapy such as 3DCRT or IMRT. Materials and methods: The study was closed in June 2006 with 750 randomised patients. Radiation therapy-related data for every patient were sent by each participating centre to the QA office where all trial data were reviewed, analysed and stored. In case of any deviation from the protocol, an interactive process was started between the QA office and the local responsible clinician and/or physicist to increase the compliance to the protocol for future randomised patients. Meetings and workshops were held on a regular basis for discussions on various trial-related issues and for the QA office to report on updated results. Results and discussion: This review covers the 734 patients out of a total of 750 who had entered the study. Deviations early in the study were corrected so that the overall compliance to the protocol was very high. There were only negligible variations in doses and dose distributions to target volumes for each specific site and stage. The quality of the treatments was high. Furthermore, an extensive database of treatment parameters was accumulated for future dose-volume vs. endpoint evaluations. Conclusions: This comprehensive QA programme increased the probability to draw firm conclusions from our study and may serve as a concept for QA work in future radiotherapy trials where comparatively small effects are searched for in a heterogeneous tumour population.},
  author       = {Johansson, Karl-Axel and Nilsson, Per and Zackrisson, Bjorn and Ohlson, Birgitta and Kjellén, Elisabeth and Mercke, Claes and Alvarez-Fonseca, Mauricio and Billstrom, Anette and Bjork-Eriksson, Thomas and Bjor, Ove and Ekberg, Lars and Friesland, Signe and Karlsson, Magnus and Lagerlund, Magnus and Lundkvist, Lena and Lofroth, Per-Olov and Löfvander-Thapper, Kerstin and Nilsson, Alta and Nyman, Jan and Persson, Essie and Reizenstein, Johan and Rosenbrand, Hans-Olov and Wikund, Fredrik and Wittgren, Lena},
  issn         = {1879-0887},
  language     = {eng},
  number       = {2},
  pages        = {290--299},
  publisher    = {Elsevier},
  series       = {Radiotherapy and Oncology},
  title        = {The quality assurance process for the ARTSCAN head and neck study - A practical interactive approach for QA in 3DCRT and IMRT},
  url          = {http://dx.doi.org/10.1016/j.radonc.2007.12.005},
  doi          = {10.1016/j.radonc.2007.12.005},
  volume       = {87},
  year         = {2008},
}