Migration and wear of a hydroxyapatite-coated hip prosthesis. A controlled roentgen stereophotogrammetric study
(1996) In Journal of Bone and Joint Surgery: British Volume 78(1). p.85-91- Abstract
- A consecutive series of 30 total hip replacements using a hydroxyapatite (HA)-coated, modular implant (Omnifit) was followed clinically and by roentgen stereophotogrammetric analysis for two years and compared with two control groups, one of 27 cemented Charnley sockets and one of 40 cemented Charnley stems. Omnifit sockets with a central gap between the dome of the socket and the acetabular bone in the postoperative radiographs, migrated less than sockets without such gaps (p = 0.01). After adjustment for patient-related factors (age, gender and weight), no significant difference was found between the two prostheses with respect to micromotion and wear. We conclude that the early fixation of the HA-coated Omnifit prosthesis compares with... (More)
- A consecutive series of 30 total hip replacements using a hydroxyapatite (HA)-coated, modular implant (Omnifit) was followed clinically and by roentgen stereophotogrammetric analysis for two years and compared with two control groups, one of 27 cemented Charnley sockets and one of 40 cemented Charnley stems. Omnifit sockets with a central gap between the dome of the socket and the acetabular bone in the postoperative radiographs, migrated less than sockets without such gaps (p = 0.01). After adjustment for patient-related factors (age, gender and weight), no significant difference was found between the two prostheses with respect to micromotion and wear. We conclude that the early fixation of the HA-coated Omnifit prosthesis compares with that of the cemented Charnley prosthesis. (Less)
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/1110746
- author
- Önsten, Ingemar LU ; Carlsson, Åke S ; Sanzén, Lennart LU and Besjakov, Jack LU
- organization
- publishing date
- 1996
- type
- Contribution to journal
- publication status
- published
- subject
- in
- Journal of Bone and Joint Surgery: British Volume
- volume
- 78
- issue
- 1
- pages
- 85 - 91
- publisher
- British Editorial Society of Bone & Joint Surgery
- external identifiers
-
- pmid:8898134
- scopus:0030071476
- ISSN
- 2044-5377
- language
- English
- LU publication?
- yes
- additional info
- The information about affiliations in this record was updated in December 2015. The record was previously connected to the following departments: Medical Radiology Unit (013241410), Reconstructive Surgery (013240300)
- id
- d433e8b6-3896-4d86-882a-339a01ff1931 (old id 1110746)
- date added to LUP
- 2016-04-01 11:51:02
- date last changed
- 2022-03-13 01:39:21
@article{d433e8b6-3896-4d86-882a-339a01ff1931, abstract = {{A consecutive series of 30 total hip replacements using a hydroxyapatite (HA)-coated, modular implant (Omnifit) was followed clinically and by roentgen stereophotogrammetric analysis for two years and compared with two control groups, one of 27 cemented Charnley sockets and one of 40 cemented Charnley stems. Omnifit sockets with a central gap between the dome of the socket and the acetabular bone in the postoperative radiographs, migrated less than sockets without such gaps (p = 0.01). After adjustment for patient-related factors (age, gender and weight), no significant difference was found between the two prostheses with respect to micromotion and wear. We conclude that the early fixation of the HA-coated Omnifit prosthesis compares with that of the cemented Charnley prosthesis.}}, author = {{Önsten, Ingemar and Carlsson, Åke S and Sanzén, Lennart and Besjakov, Jack}}, issn = {{2044-5377}}, language = {{eng}}, number = {{1}}, pages = {{85--91}}, publisher = {{British Editorial Society of Bone & Joint Surgery}}, series = {{Journal of Bone and Joint Surgery: British Volume}}, title = {{Migration and wear of a hydroxyapatite-coated hip prosthesis. A controlled roentgen stereophotogrammetric study}}, volume = {{78}}, year = {{1996}}, }