Enhanced color ratio calibration approach for two-dimensional surface thermometry using laser-induced phosphorescence

Knappe, Christoph LU ; Lindén, Johannes LU orcid ; Richter, Mattias LU and Aldén, Marcus LU (2013) In Measurement Science & Technology 24(8).
Abstract
An alternative concept for calibrating spectral image intensity ratios is described, which provides simple, but effective means of compensation for systematic errors, caused by nonlinearities in signal detection. The method relies on image segmentation by means of

signal thresholds, where pixels are organized into different subgroups according to their corresponding signal count value. Instead of defining one global spectral ratio per calibration

temperature, the phosphor-coated target surface is illuminated inhomogeneously and resulting image ratios are calibrated individually for each pixel intensity subgroup. This allows the

exploitation of high intensity regions on the camera chip which offer great precision... (More)
An alternative concept for calibrating spectral image intensity ratios is described, which provides simple, but effective means of compensation for systematic errors, caused by nonlinearities in signal detection. The method relies on image segmentation by means of

signal thresholds, where pixels are organized into different subgroups according to their corresponding signal count value. Instead of defining one global spectral ratio per calibration

temperature, the phosphor-coated target surface is illuminated inhomogeneously and resulting image ratios are calibrated individually for each pixel intensity subgroup. This allows the

exploitation of high intensity regions on the camera chip which offer great precision advantages, but suffer from systematic errors caused by signal nonlinearities. Temperature calibration data of BaMg2Al16O27:Eu in the temperature range between 270 and 470 K was

used to assess and compare the potential of both calibration approaches. In comparison to the conventional, e.g. global calibration approach, accuracy improvements of up to 39% were

gained even while keeping average signal intensities below 15% of the detector’s full dynamic range. Image ratio evaluations, based on segregated pixel subgroups, could help improve measurement accuracy also for other techniques, relying on the calibration of measured

quantities. In two-dimensional phosphor thermometry, it helps bridge the current precision gap between two-color ratio methods and more elaborate lifetime-imaging approaches. (Less)
Please use this url to cite or link to this publication:
author
; ; and
organization
publishing date
type
Contribution to journal
publication status
published
subject
keywords
laser-induced phosphorescence, thermographic phosphors, thermometry, ICCD detectors, BAM
in
Measurement Science & Technology
volume
24
issue
8
article number
085202
publisher
IOP Publishing
external identifiers
  • wos:000321943100022
  • scopus:84880774868
ISSN
0957-0233
DOI
10.1088/0957-0233/24/8/085202
language
English
LU publication?
yes
id
d7609b61-0e08-4fce-be60-e833c1a59ab3 (old id 4022183)
alternative location
http://iopscience.iop.org/0957-0233/24/8/085202
date added to LUP
2016-04-01 14:09:42
date last changed
2022-01-27 23:06:45
@article{d7609b61-0e08-4fce-be60-e833c1a59ab3,
  abstract     = {{An alternative concept for calibrating spectral image intensity ratios is described, which provides simple, but effective means of compensation for systematic errors, caused by nonlinearities in signal detection. The method relies on image segmentation by means of<br/><br>
signal thresholds, where pixels are organized into different subgroups according to their corresponding signal count value. Instead of defining one global spectral ratio per calibration<br/><br>
temperature, the phosphor-coated target surface is illuminated inhomogeneously and resulting image ratios are calibrated individually for each pixel intensity subgroup. This allows the<br/><br>
exploitation of high intensity regions on the camera chip which offer great precision advantages, but suffer from systematic errors caused by signal nonlinearities. Temperature calibration data of BaMg2Al16O27:Eu in the temperature range between 270 and 470 K was<br/><br>
used to assess and compare the potential of both calibration approaches. In comparison to the conventional, e.g. global calibration approach, accuracy improvements of up to 39% were<br/><br>
gained even while keeping average signal intensities below 15% of the detector’s full dynamic range. Image ratio evaluations, based on segregated pixel subgroups, could help improve measurement accuracy also for other techniques, relying on the calibration of measured<br/><br>
quantities. In two-dimensional phosphor thermometry, it helps bridge the current precision gap between two-color ratio methods and more elaborate lifetime-imaging approaches.}},
  author       = {{Knappe, Christoph and Lindén, Johannes and Richter, Mattias and Aldén, Marcus}},
  issn         = {{0957-0233}},
  keywords     = {{laser-induced phosphorescence; thermographic phosphors; thermometry; ICCD detectors; BAM}},
  language     = {{eng}},
  number       = {{8}},
  publisher    = {{IOP Publishing}},
  series       = {{Measurement Science & Technology}},
  title        = {{Enhanced color ratio calibration approach for two-dimensional surface thermometry using laser-induced phosphorescence}},
  url          = {{http://dx.doi.org/10.1088/0957-0233/24/8/085202}},
  doi          = {{10.1088/0957-0233/24/8/085202}},
  volume       = {{24}},
  year         = {{2013}},
}