Inter-subject variability and normal limits of the SITA Standard, SITA Fast, and the Humphrey Full Threshold computerized perimetry strategies, SITA STATPAC

Bengtsson, B LU and Heijl, A LU (1999) In Acta Ophthalmologica Scandinavica 77(2). p.9-125
Abstract

PURPOSE: To establish and evaluate inter-subject variability and normal threshold limits for the new SITA strategies and to compare them with those obtained with the traditional Humphrey Full Threshold algorithm.

METHODS: Data from 330 eyes of 330 normal subjects were collected at 10 centres in order to establish limits of normality for the new SITA strategies and thus, to make it possible to subject SITA fields to computer-assisted visual field analysis. Two visual field tests were obtained with each of the SITA Standard, SITA Fast, and the Full Threshold algorithms.

RESULTS: Inter-subject variance was 31% smaller with SITA Standard and 41% smaller with SITA Fast than with Full Threshold (p<0.0001). Age-dependent... (More)

PURPOSE: To establish and evaluate inter-subject variability and normal threshold limits for the new SITA strategies and to compare them with those obtained with the traditional Humphrey Full Threshold algorithm.

METHODS: Data from 330 eyes of 330 normal subjects were collected at 10 centres in order to establish limits of normality for the new SITA strategies and thus, to make it possible to subject SITA fields to computer-assisted visual field analysis. Two visual field tests were obtained with each of the SITA Standard, SITA Fast, and the Full Threshold algorithms.

RESULTS: Inter-subject variance was 31% smaller with SITA Standard and 41% smaller with SITA Fast than with Full Threshold (p<0.0001). Age-dependent decrease of differential light sensitivity was also significantly smaller with both SITA algorithms than with Full Threshold (p<0.0001), 23% and 25% respectively. Mean sensitivity was somewhat higher with both SITA Standard (29.5 dB) and SITA Fast (29.9 dB) as compared to Full Threshold (28.3 dB) (p<0.0001). Normal limits were tightened between 9 to 29% at different significance levels with SITA.

CONCLUSION: SITA test results from eyes with normal visual fields will on average be more even from centre to mid-periphery as compared with Full Threshold fields. They will also appear slightly lighter in grey-scale representations. Because of smaller inter-subject variance, shallower depressions are needed in SITA fields for statistical and clinical significance.

(Less)
Please use this url to cite or link to this publication:
author
and
organization
publishing date
type
Contribution to journal
publication status
published
keywords
Adult, Aged, Aged, 80 and over, Algorithms, Female, Humans, Male, Middle Aged, Reference Values, Reproducibility of Results, Sensitivity and Specificity, Visual Field Tests, Visual Fields
in
Acta Ophthalmologica Scandinavica
volume
77
issue
2
pages
9 - 125
publisher
Wiley
external identifiers
  • scopus:0032941718
  • pmid:10321523
ISSN
1395-3907
DOI
10.1034/j.1600-0420.1999.770201.x
language
English
LU publication?
yes
id
d8a75dbb-7b8b-4e72-a461-f9d11ed9d7e0
date added to LUP
2016-08-30 16:59:31
date last changed
2024-04-05 03:56:51
@article{d8a75dbb-7b8b-4e72-a461-f9d11ed9d7e0,
  abstract     = {{<p>PURPOSE: To establish and evaluate inter-subject variability and normal threshold limits for the new SITA strategies and to compare them with those obtained with the traditional Humphrey Full Threshold algorithm.</p><p>METHODS: Data from 330 eyes of 330 normal subjects were collected at 10 centres in order to establish limits of normality for the new SITA strategies and thus, to make it possible to subject SITA fields to computer-assisted visual field analysis. Two visual field tests were obtained with each of the SITA Standard, SITA Fast, and the Full Threshold algorithms.</p><p>RESULTS: Inter-subject variance was 31% smaller with SITA Standard and 41% smaller with SITA Fast than with Full Threshold (p&lt;0.0001). Age-dependent decrease of differential light sensitivity was also significantly smaller with both SITA algorithms than with Full Threshold (p&lt;0.0001), 23% and 25% respectively. Mean sensitivity was somewhat higher with both SITA Standard (29.5 dB) and SITA Fast (29.9 dB) as compared to Full Threshold (28.3 dB) (p&lt;0.0001). Normal limits were tightened between 9 to 29% at different significance levels with SITA.</p><p>CONCLUSION: SITA test results from eyes with normal visual fields will on average be more even from centre to mid-periphery as compared with Full Threshold fields. They will also appear slightly lighter in grey-scale representations. Because of smaller inter-subject variance, shallower depressions are needed in SITA fields for statistical and clinical significance.</p>}},
  author       = {{Bengtsson, B and Heijl, A}},
  issn         = {{1395-3907}},
  keywords     = {{Adult; Aged; Aged, 80 and over; Algorithms; Female; Humans; Male; Middle Aged; Reference Values; Reproducibility of Results; Sensitivity and Specificity; Visual Field Tests; Visual Fields}},
  language     = {{eng}},
  number       = {{2}},
  pages        = {{9--125}},
  publisher    = {{Wiley}},
  series       = {{Acta Ophthalmologica Scandinavica}},
  title        = {{Inter-subject variability and normal limits of the SITA Standard, SITA Fast, and the Humphrey Full Threshold computerized perimetry strategies, SITA STATPAC}},
  url          = {{http://dx.doi.org/10.1034/j.1600-0420.1999.770201.x}},
  doi          = {{10.1034/j.1600-0420.1999.770201.x}},
  volume       = {{77}},
  year         = {{1999}},
}