Beam Adjustments for Unflattened X-ray Beam Modes for an Elekta Synergy Linear Accelerator

Blad, Börje LU and Dalaryd, Mårten LU (2012) 5th European Conference of the International Federation for Medical and Biological Engineering 37. p.683-685
Abstract
A flattening filter free x-ray beam mode was added to an Elekta linear accelerator by placing a flat copper plate in the filter carousel and recalibrating the electron steering servo, gun servo and dosimetry system. Machine configurations were saved onto a separate hard disk in order to remain separate from the clinical configuration. Profile measurements with a Schuster BMS diode array show that the beam is stable and start up performance is similar to the normal filtered beam.
Please use this url to cite or link to this publication:
author
and
organization
publishing date
type
Chapter in Book/Report/Conference proceeding
publication status
published
subject
keywords
Elekta Synergy linear accelerator, unflattened X-ray beam, beam, adjustments
host publication
IFMBE Proceedings
volume
37
pages
683 - 685
publisher
Springer
conference name
5th European Conference of the International Federation for Medical and Biological Engineering
conference location
Budapest, Hungary
conference dates
2011-09-14 - 2011-09-18
external identifiers
  • wos:000310938200177
  • scopus:80455140656
ISSN
1680-0737
ISBN
978-3-642-23508-5
DOI
10.1007/978-3-642-23508-5_177
language
English
LU publication?
yes
id
d9efb880-adca-453e-a09c-696d40fd1ce9 (old id 3379349)
date added to LUP
2016-04-01 13:55:54
date last changed
2022-01-27 21:54:53
@inproceedings{d9efb880-adca-453e-a09c-696d40fd1ce9,
  abstract     = {{A flattening filter free x-ray beam mode was added to an Elekta linear accelerator by placing a flat copper plate in the filter carousel and recalibrating the electron steering servo, gun servo and dosimetry system. Machine configurations were saved onto a separate hard disk in order to remain separate from the clinical configuration. Profile measurements with a Schuster BMS diode array show that the beam is stable and start up performance is similar to the normal filtered beam.}},
  author       = {{Blad, Börje and Dalaryd, Mårten}},
  booktitle    = {{IFMBE Proceedings}},
  isbn         = {{978-3-642-23508-5}},
  issn         = {{1680-0737}},
  keywords     = {{Elekta Synergy linear accelerator; unflattened X-ray beam; beam; adjustments}},
  language     = {{eng}},
  pages        = {{683--685}},
  publisher    = {{Springer}},
  title        = {{Beam Adjustments for Unflattened X-ray Beam Modes for an Elekta Synergy Linear Accelerator}},
  url          = {{http://dx.doi.org/10.1007/978-3-642-23508-5_177}},
  doi          = {{10.1007/978-3-642-23508-5_177}},
  volume       = {{37}},
  year         = {{2012}},
}