Beam Adjustments for Unflattened X-ray Beam Modes for an Elekta Synergy Linear Accelerator
(2012) 5th European Conference of the International Federation for Medical and Biological Engineering 37. p.683-685- Abstract
- A flattening filter free x-ray beam mode was added to an Elekta linear accelerator by placing a flat copper plate in the filter carousel and recalibrating the electron steering servo, gun servo and dosimetry system. Machine configurations were saved onto a separate hard disk in order to remain separate from the clinical configuration. Profile measurements with a Schuster BMS diode array show that the beam is stable and start up performance is similar to the normal filtered beam.
Please use this url to cite or link to this publication:
https://lup.lub.lu.se/record/3379349
- author
- Blad, Börje LU and Dalaryd, Mårten LU
- organization
- publishing date
- 2012
- type
- Chapter in Book/Report/Conference proceeding
- publication status
- published
- subject
- keywords
- Elekta Synergy linear accelerator, unflattened X-ray beam, beam, adjustments
- host publication
- IFMBE Proceedings
- volume
- 37
- pages
- 683 - 685
- publisher
- Springer
- conference name
- 5th European Conference of the International Federation for Medical and Biological Engineering
- conference location
- Budapest, Hungary
- conference dates
- 2011-09-14 - 2011-09-18
- external identifiers
-
- wos:000310938200177
- scopus:80455140656
- ISSN
- 1680-0737
- ISBN
- 978-3-642-23508-5
- DOI
- 10.1007/978-3-642-23508-5_177
- language
- English
- LU publication?
- yes
- id
- d9efb880-adca-453e-a09c-696d40fd1ce9 (old id 3379349)
- date added to LUP
- 2016-04-01 13:55:54
- date last changed
- 2025-04-04 14:17:11
@inproceedings{d9efb880-adca-453e-a09c-696d40fd1ce9, abstract = {{A flattening filter free x-ray beam mode was added to an Elekta linear accelerator by placing a flat copper plate in the filter carousel and recalibrating the electron steering servo, gun servo and dosimetry system. Machine configurations were saved onto a separate hard disk in order to remain separate from the clinical configuration. Profile measurements with a Schuster BMS diode array show that the beam is stable and start up performance is similar to the normal filtered beam.}}, author = {{Blad, Börje and Dalaryd, Mårten}}, booktitle = {{IFMBE Proceedings}}, isbn = {{978-3-642-23508-5}}, issn = {{1680-0737}}, keywords = {{Elekta Synergy linear accelerator; unflattened X-ray beam; beam; adjustments}}, language = {{eng}}, pages = {{683--685}}, publisher = {{Springer}}, title = {{Beam Adjustments for Unflattened X-ray Beam Modes for an Elekta Synergy Linear Accelerator}}, url = {{http://dx.doi.org/10.1007/978-3-642-23508-5_177}}, doi = {{10.1007/978-3-642-23508-5_177}}, volume = {{37}}, year = {{2012}}, }