Detection of Atmospheric Hydrogen Using ECD Detector Doped with N2O

Necki, J.M. ; Heliasz, M. LU ; Rosiek, J. ; Pycia, M. ; Rozanski, K. ; Sliwka, L. and Bartyzel, J. (2009) In Chemia Analityczna 54(4). p.705-716
Abstract
Atmospheric concentrations of molecular hydrogen are commonly measured with the aid of commercial reduction gas analysers (RGA). An alternative technique of H2 detection has been proposed and tested. It is based on commercial automated gas chromatograph equipped with a μ-ECD detector doped with N2O. The main advantage of the presented method is its suitability for high-precision monitoring of atmospheric H2. Typical precision of single analysis, which was obtained in the tests of the system, is in the order of 1 nmol mol-1. This value is comparable to the precision offered by commercial RGA detectors. The μ-ECD detector investigated in this study revealed linear response in the range of H 2 concentrations encountered in atmospheric... (More)
Atmospheric concentrations of molecular hydrogen are commonly measured with the aid of commercial reduction gas analysers (RGA). An alternative technique of H2 detection has been proposed and tested. It is based on commercial automated gas chromatograph equipped with a μ-ECD detector doped with N2O. The main advantage of the presented method is its suitability for high-precision monitoring of atmospheric H2. Typical precision of single analysis, which was obtained in the tests of the system, is in the order of 1 nmol mol-1. This value is comparable to the precision offered by commercial RGA detectors. The μ-ECD detector investigated in this study revealed linear response in the range of H 2 concentrations encountered in atmospheric monitoring. The main disadvantage of the method is high sensitivity of the μ-ECD detector to the impurities in gases used in the analytical system. (Less)
Please use this url to cite or link to this publication:
author
; ; ; ; ; and
publishing date
type
Contribution to journal
publication status
published
in
Chemia Analityczna
volume
54
issue
4
pages
705 - 716
publisher
Polskie Towarzystwo Chemiczne - Polish Chemical Society
external identifiers
  • scopus:70450190064
ISSN
0009-2223
language
English
LU publication?
no
id
f34da09b-495a-4afa-be84-7a0633df46ad
date added to LUP
2019-05-20 11:53:29
date last changed
2022-04-26 00:05:22
@article{f34da09b-495a-4afa-be84-7a0633df46ad,
  abstract     = {{Atmospheric concentrations of molecular hydrogen are commonly measured with the aid of commercial reduction gas analysers (RGA). An alternative technique of H2 detection has been proposed and tested. It is based on commercial automated gas chromatograph equipped with a μ-ECD detector doped with N2O. The main advantage of the presented method is its suitability for high-precision monitoring of atmospheric H2. Typical precision of single analysis, which was obtained in the tests of the system, is in the order of 1 nmol mol-1. This value is comparable to the precision offered by commercial RGA detectors. The μ-ECD detector investigated in this study revealed linear response in the range of H 2 concentrations encountered in atmospheric monitoring. The main disadvantage of the method is high sensitivity of the μ-ECD detector to the impurities in gases used in the analytical system.}},
  author       = {{Necki, J.M. and Heliasz, M. and Rosiek, J. and Pycia, M. and Rozanski, K. and Sliwka, L. and Bartyzel, J.}},
  issn         = {{0009-2223}},
  language     = {{eng}},
  number       = {{4}},
  pages        = {{705--716}},
  publisher    = {{Polskie Towarzystwo Chemiczne - Polish Chemical Society}},
  series       = {{Chemia Analityczna}},
  title        = {{Detection of Atmospheric Hydrogen Using ECD Detector Doped with N2O}},
  volume       = {{54}},
  year         = {{2009}},
}