Hemodynamic impact of abdominal aortic aneurysm stent-graft implantation-induced stenosis

Aristokleous, Nicolas LU orcid ; Kontopodis, Nikolaos G. ; Tzirakis, Konstantinos ; Ioannou, Christos V. and Papaharilaou, Yannis (2016) In Medical & Biological Engineering & Computing 54(10). p.1523-1532
Abstract

The current study aims to computationally evaluate the hemodynamic impact of a novel sealing mechanism employed by a recently developed endograft (Ovation TriVascular Stent-Graft System) for endovascular aneurysm repair. The exploitation of two inflatable O-rings to achieve sealing may be advantageous in terms of accommodating challenging anatomies, but comes at a price of a marked inflow stenosis. Here, four representative patient cases of inflow stenosis ranging from 30 to 80 % were analyzed. Lumen surface models were constructed from 1 month post-operative computed tomography images and then used to numerically compute the complex endograft flow field. Our results highlight coexistence of stenotic wall regions exposed to high shear... (More)

The current study aims to computationally evaluate the hemodynamic impact of a novel sealing mechanism employed by a recently developed endograft (Ovation TriVascular Stent-Graft System) for endovascular aneurysm repair. The exploitation of two inflatable O-rings to achieve sealing may be advantageous in terms of accommodating challenging anatomies, but comes at a price of a marked inflow stenosis. Here, four representative patient cases of inflow stenosis ranging from 30 to 80 % were analyzed. Lumen surface models were constructed from 1 month post-operative computed tomography images and then used to numerically compute the complex endograft flow field. Our results highlight coexistence of stenotic wall regions exposed to high shear rate and post-stenotic recirculation zones. These conditions may implicate platelet activation and predispose thrombus formation and thromboembolic complications. A clinically insignificant cycle-averaged pressure drop along the inflow stenosis and further in the endograft main body legs was predicted (range 0.01–1.72 mmHg) which was, however, notable at peak systole (range 3.52–19.73 mmHg). Although the functional impact of the endograft stenosis at rest flow conditions may appear insignificant, increased flow rate during exercise is expected to strongly accentuate the observed effects. Pressure drop in the endograft legs was attributed to suboptimal, based on Murray’s scaling law, cross-sectional area ratio between trunk and legs of the device.

(Less)
Please use this url to cite or link to this publication:
author
; ; ; and
publishing date
type
Contribution to journal
publication status
published
subject
keywords
Abdominal aortic aneurysm, Endovascular graft, Image-based CFD, Thrombosis
in
Medical & Biological Engineering & Computing
volume
54
issue
10
pages
10 pages
publisher
Springer
external identifiers
  • pmid:26676685
  • scopus:84949971402
ISSN
0140-0118
DOI
10.1007/s11517-015-1425-1
language
English
LU publication?
no
id
fd8ac495-de99-4f67-aad8-4bde518ac260
date added to LUP
2019-05-14 09:24:19
date last changed
2024-02-15 01:12:52
@article{fd8ac495-de99-4f67-aad8-4bde518ac260,
  abstract     = {{<p>The current study aims to computationally evaluate the hemodynamic impact of a novel sealing mechanism employed by a recently developed endograft (Ovation TriVascular Stent-Graft System) for endovascular aneurysm repair. The exploitation of two inflatable O-rings to achieve sealing may be advantageous in terms of accommodating challenging anatomies, but comes at a price of a marked inflow stenosis. Here, four representative patient cases of inflow stenosis ranging from 30 to 80 % were analyzed. Lumen surface models were constructed from 1 month post-operative computed tomography images and then used to numerically compute the complex endograft flow field. Our results highlight coexistence of stenotic wall regions exposed to high shear rate and post-stenotic recirculation zones. These conditions may implicate platelet activation and predispose thrombus formation and thromboembolic complications. A clinically insignificant cycle-averaged pressure drop along the inflow stenosis and further in the endograft main body legs was predicted (range 0.01–1.72 mmHg) which was, however, notable at peak systole (range 3.52–19.73 mmHg). Although the functional impact of the endograft stenosis at rest flow conditions may appear insignificant, increased flow rate during exercise is expected to strongly accentuate the observed effects. Pressure drop in the endograft legs was attributed to suboptimal, based on Murray’s scaling law, cross-sectional area ratio between trunk and legs of the device.</p>}},
  author       = {{Aristokleous, Nicolas and Kontopodis, Nikolaos G. and Tzirakis, Konstantinos and Ioannou, Christos V. and Papaharilaou, Yannis}},
  issn         = {{0140-0118}},
  keywords     = {{Abdominal aortic aneurysm; Endovascular graft; Image-based CFD; Thrombosis}},
  language     = {{eng}},
  month        = {{10}},
  number       = {{10}},
  pages        = {{1523--1532}},
  publisher    = {{Springer}},
  series       = {{Medical & Biological Engineering & Computing}},
  title        = {{Hemodynamic impact of abdominal aortic aneurysm stent-graft implantation-induced stenosis}},
  url          = {{http://dx.doi.org/10.1007/s11517-015-1425-1}},
  doi          = {{10.1007/s11517-015-1425-1}},
  volume       = {{54}},
  year         = {{2016}},
}