Wash-in and wash-out of sevoflurane in a test-lung model : A comparison between Aisys and FLOW-i
(2017) In F1000Research 6.- Abstract
Background: Modern anaesthesia workstations are reassuringly tight and are equipped with effective gas monitoring, thus providing good opportunities for low/minimal flow anaesthesia. A prerequisite for effective low flow anaesthesia is the possibility to rapidly increase and decrease gas concentrations in the circle system, thereby controlling the depth of anaesthesia. Methods: We studied the wash-in and wash-out of sevoflurane in the circle system with fixed fresh gas flow and vaporizer setting. We compared two modern anaesthesia work stations, the Aisys (GE, Madison, WI, USA) and FLOW-i (Maquet, Solna, Sweden) in a test lung model. Results: We found fresh-gas flow to have, as expected, a major influence on wash-in, as well as wash-out... (More)
Background: Modern anaesthesia workstations are reassuringly tight and are equipped with effective gas monitoring, thus providing good opportunities for low/minimal flow anaesthesia. A prerequisite for effective low flow anaesthesia is the possibility to rapidly increase and decrease gas concentrations in the circle system, thereby controlling the depth of anaesthesia. Methods: We studied the wash-in and wash-out of sevoflurane in the circle system with fixed fresh gas flow and vaporizer setting. We compared two modern anaesthesia work stations, the Aisys (GE, Madison, WI, USA) and FLOW-i (Maquet, Solna, Sweden) in a test lung model. Results: We found fresh-gas flow to have, as expected, a major influence on wash-in, as well as wash-out of sevoflurane. The wash-in time to reach a stable circle 1 MAC (2.1%) decreased from an average of 547 ± 83 seconds with a constant fresh gas flow of 300 ml/min and vaporizer setting of 8%, to a mean of 38 ± 6 seconds at a fresh gas flow of 4 L/min. There were only minor differences between the two works-stations tested; the Aisys was slightly faster at both 300 and 4 L/min flow. Time to further increase circle end-tidal concentration from 1-1.5 MAC showed likewise significant associations to fresh gas and decreased from 330 ± 24 seconds at 300 ml/min. to less than a minute at constant 4 L/min (17 ± 11 seconds), without anaesthetic machine difference. Wash-out was also fresh gas flow dependent and plateaued at 7.5 L/min. Conclusions: Circle system wash-in and wash-out show clear fresh gas dependency and varies somewhat between the Aisys and Flow-i. The circle saturation, reaching 1 MAC end-tidal or increasing from 1-1.5 MAC can be achieved with both work-stations within 1.5 minutes at a constant fresh gas flow of 2 and 4 L/min. Wash-out plateaued at 7.5 L/min.
(Less)
- author
- Jakobsson, Petter ; Lindgren, Madleine and Jakobsson, Jan G.
- publishing date
- 2017
- type
- Contribution to journal
- publication status
- published
- subject
- keywords
- End-tidal concentration, Low-flow anaesthesia, MAC, Sevoflurane, Wash-in
- in
- F1000Research
- volume
- 6
- article number
- 389
- publisher
- F1000 Research Ltd.
- external identifiers
-
- pmid:28529707
- scopus:85019147718
- ISSN
- 2046-1402
- DOI
- 10.12688/f1000research.11255.2
- language
- English
- LU publication?
- no
- id
- 03c06ca4-6977-499b-9689-1b19a6624ae7
- date added to LUP
- 2017-06-01 14:44:35
- date last changed
- 2024-09-02 01:21:39
@misc{03c06ca4-6977-499b-9689-1b19a6624ae7, abstract = {{<p>Background: Modern anaesthesia workstations are reassuringly tight and are equipped with effective gas monitoring, thus providing good opportunities for low/minimal flow anaesthesia. A prerequisite for effective low flow anaesthesia is the possibility to rapidly increase and decrease gas concentrations in the circle system, thereby controlling the depth of anaesthesia. Methods: We studied the wash-in and wash-out of sevoflurane in the circle system with fixed fresh gas flow and vaporizer setting. We compared two modern anaesthesia work stations, the Aisys (GE, Madison, WI, USA) and FLOW-i (Maquet, Solna, Sweden) in a test lung model. Results: We found fresh-gas flow to have, as expected, a major influence on wash-in, as well as wash-out of sevoflurane. The wash-in time to reach a stable circle 1 MAC (2.1%) decreased from an average of 547 ± 83 seconds with a constant fresh gas flow of 300 ml/min and vaporizer setting of 8%, to a mean of 38 ± 6 seconds at a fresh gas flow of 4 L/min. There were only minor differences between the two works-stations tested; the Aisys was slightly faster at both 300 and 4 L/min flow. Time to further increase circle end-tidal concentration from 1-1.5 MAC showed likewise significant associations to fresh gas and decreased from 330 ± 24 seconds at 300 ml/min. to less than a minute at constant 4 L/min (17 ± 11 seconds), without anaesthetic machine difference. Wash-out was also fresh gas flow dependent and plateaued at 7.5 L/min. Conclusions: Circle system wash-in and wash-out show clear fresh gas dependency and varies somewhat between the Aisys and Flow-i. The circle saturation, reaching 1 MAC end-tidal or increasing from 1-1.5 MAC can be achieved with both work-stations within 1.5 minutes at a constant fresh gas flow of 2 and 4 L/min. Wash-out plateaued at 7.5 L/min.</p>}}, author = {{Jakobsson, Petter and Lindgren, Madleine and Jakobsson, Jan G.}}, issn = {{2046-1402}}, keywords = {{End-tidal concentration; Low-flow anaesthesia; MAC; Sevoflurane; Wash-in}}, language = {{eng}}, publisher = {{F1000 Research Ltd.}}, series = {{F1000Research}}, title = {{Wash-in and wash-out of sevoflurane in a test-lung model : A comparison between Aisys and FLOW-i}}, url = {{http://dx.doi.org/10.12688/f1000research.11255.2}}, doi = {{10.12688/f1000research.11255.2}}, volume = {{6}}, year = {{2017}}, }